The following table or figure may be downloaded to PowerPoint for personal use in teaching and presentations. This feature is available to all subscribers to the journal. You MUST read and follow the guidelines at Request to Reproduce AJR Content if you are distributing or using AJR content beyond academic use (limited distribution, non-revenue producing, or educational purposes). (Downloading may take up to 30 seconds.
If the slide opens in your browser, select File -> Save As to save it.)

Fig. 4. Average score for image quality of specified image against
quality of reference image obtained at 160 mAs. Error bars correspond to
standard error at each setting for five interpreters. These data clearly show
that increasing radiographic tube output (and patient dose) beyond 120 mAs is
not justified by any corresponding benefit in terms of CT image quality.