The following table or figure may be downloaded to PowerPoint for personal use in teaching and presentations. This feature is available to all subscribers to the journal. You MUST read and follow the guidelines at Request to Reproduce AJR Content if you are distributing or using AJR content beyond academic use (limited distribution, non-revenue producing, or educational purposes). (Downloading may take up to 30 seconds.
If the slide opens in your browser, select File -> Save As to save it.)
Fig. 1A. 50-year-old woman with nondisplaced fracture of left C6
transverse process. Both axial helical CT scans were obtained at same level
with 120 kVp, 3-mm slice thickness, and pitch of 1.5 and filmed in bone window
settings (width, 2000 H; length, 500 H). Milliampere setting was 130 mA for
A and 240 mA for B. Ionization chamber estimates indicate
protocol for A delivered 14.1 mGy to thyroid, whereas protocol for
B delivered 26.0 mGy to thyroid.