The following table or figure may be downloaded to PowerPoint for personal use in teaching and presentations. This feature is available to all subscribers to the journal. You MUST read and follow the guidelines at Request to Reproduce AJR Content if you are distributing or using AJR content beyond academic use (limited distribution, non-revenue producing, or educational purposes). (Downloading may take up to 30 seconds.
If the slide opens in your browser, select File -> Save As to save it.)
Click on image to view larger version.
Fig. 4A —Phantom with simulated vessels with 50% stenosis. Results
show low attenuation decreased contrast-to-noise ratio and caused
overestimation of stenosis, especially in small vessels. High attenuation
enhanced contrast-to-noise ratio but also caused severe partial volume effect.
Thus larger stenosis in larger vessel was clearly depicted but stenosis was
underestimated in smaller vessels. CT attenuation of 350 H is optimal for
accurate quantification of stenosis in larger and smaller vessels. CT scan
shows 3- and 5-mm coronary arteries under low-attenuation (200 H)
protocol.