1 From the Department of Radiology, The University of Colorado Medical Center, Denver, Colorado
Per cent depth doses were measured with LiF in water phantoms exposed to orthovoltage x rays and cobalt 60 radiation. Agreement within ±I per cent was achieved between LiF and ionization chamber measurements of cobalt 60 per cent depth dose. Unsatisfactory results were obtained with LiF measurements of per cent depth dose from orthovoltage x rays.