AJR Your Link to CME
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by LANE, K.
Right arrow Articles by DAVIS, L. W.
Right arrow Search for Related Content
PubMed
Right arrow Articles by LANE, K.
Right arrow Articles by DAVIS, L. W.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?

COMPUTER GENERATED ISODOSE CURVES FOR HIGH ENERGY X-RAY MACHINES

EXPERIMENTAL VERIFICATION AND TIME-SAVING STRATEGIES

KAREN LANE B.A., PETER BLOCH PH.D., and LAWRENCE W. DAVIS M.D.

Calculation of isodose patterns for high energy therapy units using Cunningham’s BEAM program requires accurate information on the beam profile in air. This may be obtained by scanning across the x-ray beam using a very small silicon diode detector. Data obtained on a particular unit may not be applicable to any other unit.

Computer calculations of isodose curves for high energy x-ray machines must use a 0.25 cm. or smaller lateral grid of dose calculation points to ensure accurate reproduction of the sharp fall-off of the dose at the beam edge. Computer calculation time to obtain the isodose curves using this size lateral grid is prohibitively long. The computing time can be appreciably reduced by: (1) changing the width of the annular sectors used in calculating the contribution of scattered radiation to the dose to 0.5 cm.; and (2) calculating the dose at 0.25 cm. lateral intervals within 1 cm. of the beam edge, but changing the spacing of the grid to 1 cm. inside the field. These programming strategies do not produce any significant change in the isodose curves but lead to a significant reduction in computer time.