Articles |
A method of providing in vivo measurement of entrance skin exposure for mammography using mailed thermoluminescent dosimeters has been developed. The dosimeters are calibrated to 137Cs (662 keV), and a correction is applied to account for energy response. A mailed dosimeter packet which can be used to measure the half-value layer at mammography energies is also described. The results for both systems agree well (to within 10%) with careful ion chamber determinations. These dosimeter measurements provide a simple. Inexpensive means to reliably determine exposure and to monitor machine parameters on a periodic basis.
This article has been cited by other articles:
![]() |
H M Warren-Forward and L Duggan Towards in vivo TLD dosimetry in mammography Br. J. Radiol., May 1, 2004; 77(917): 426 - 432. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |