For 4 months, a prototype digital subtraction system was used to obtain images of the cerebral vasculature after intraarterial contrast injections. In 12 instances, the intraarterial injections were recorded with both a digital subtraction unit and conventional direct magnification film-screen system. The digital subtraction and conventional film subtraction images were compared and graded for quality and information content by three skilled observers. In addition, quantitative measurements of contrast-detail performance and spatial resolution were obtained on both the digital system and the screen-film imaging chain. In a clinical setting, both the digital subtraction and conventional film-screen systems provided similar quality images and angiographic information. Contrast-detail curves demonstrated that digital subtraction angiography outperformed conventional film technique for low-contrast objects. Digital subtraction angiography also reduced the time required to obtain the angiogram, markedly reduced film cost, and lowered the contrast agent burden.
This article has been cited by other articles:
![]() |
W. M. Kelly Current Role of Digital Subtraction Angiography Perspectives in Vascular Surgery and Endovascular Therapy, January 1, 1989; 2(1): 105 - 119. [PDF] |
||||
![]() |
G. Andros, R. W. Harris, L. B. Dulawa, R. W. Oblath, and S. X. Salles-Cunha The Need for Arteriography in Diabetic Patients With Gangrene and Palpable Foot Pulses Arch Surg, November 1, 1984; 119(11): 1260 - 1263. [Abstract] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |