Department of Diagnostic Radiology, University Hospital St. Radboud, Nijmegen, the Netherlands.
Twenty-four patients with carcinoma of the urinary bladder were examined with MR imaging by using both a conventional body coil and a newly designed wraparound double surface coil. All patients had T1- and two patients had T2-weighted pulse sequences with both coils. Because of the long examination times, T2-weighted sequences in all other patients were obtained by using the double surface coil only. Nine patients underwent radical cystectomy, while the other patients had deep transurethral tumor resection and a clinical follow-up of at least 1.5 years. The results of imaging with the surface coil showed a twofold improvement of spatial resolution compared with the images obtained with the body coil. The RF field homogeneity was excellent, and the field of view was sufficiently large to achieve a complete study of the pelvis and the lower abdomen in one sequence. For the T1-weighted images, the higher spatial resolution of the double surface coil resulted in a significant (p less than .10) improvement in tumor staging. The accuracies of double-surface-coil and body-coil imaging were 79% and 54%, respectively. Double-surface-coil imaging appears to provide better MR images of the urinary bladder than body-coil imaging does, resulting in more accurate staging of tumors on T1-weighted pulse sequences.
This article has been cited by other articles:
![]() |
V. Kundra and P. M. Silverman Imaging in the Diagnosis, Staging, and Follow-Up of Cancer of the Urinary Bladder Am. J. Roentgenol., April 1, 2003; 180(4): 1045 - 1054. [Full Text] [PDF] |
||||
![]() |
T B Hall and A D MacVicar Imaging of bladder cancer Imaging, April 1, 2001; 13(1): 1 - 10. [Full Text] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |