AJR ARRS Membership
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Burton, E.
Right arrow Articles by Kereiakes, J.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Burton, E.
Right arrow Articles by Kereiakes, J.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
American Journal of Roentgenology, Vol 151, Issue 5, 999-1002
Copyright © 1988 by American Roentgen Ray Society


Articles

Evaluation of a low-dose neonatal chest radiographic system

EM Burton, DR Kirks, JL Strife, GC Henry, and JG Kereiakes

Department of Radiology, Children's Hospital Medical Center, Cincinnati, OH 45229.

A new low-dose chest radiographic system for use in the neonatal nursery was evaluated. This test system, composed of a Du Pont Kevlar fiber-front cassette, Quanta fast-detail screen, Cronex 4L film (wide latitude), and additional yttrium filtration (0.1 mm), reduced the radiation dose in neonatal chest radiography by 69% (0.9 vs 2.9 mrad [0.009 vs 0.029 mGy]) as compared with a conventional system without added yttrium filtration; the thyroid dose was reduced by 76% (0.9 vs 3.7 mrad [0.009 vs 0.037 mGy]). The cumulative dose reduction was achieved through a combination of factors, including (1) beam hardening by the added yttrium filter, (2) increased X-ray transmission through the Kevlar cassette, and (3) a fast film-screen combination. Scatter radiation at distances of 1 and 6 ft. (0.3 and 1.8 m) was negligible for both systems. Image sharpness was compared for the conventional system with and without added yttrium filtration and for the Kevlar system with yttrium. Although sharpness of bony detail was unchanged by adding yttrium filtration to the conventional system, a decrease in sharpness was noted with the Kevlar system. Because image sharpness was affected in the test system, we are not using the Kevlar-Cronex 4L system for mobile chest radiography in the neonatal intensive care unit, despite dose reductions. However, further study is recommended to determine if there is a slower film-screen combination with yttrium filtration that will not degrade image sharpness.