1
Department of Vascular Surgery, University Medical Center Utrecht,
Heidelberglaan 100, 3584 CX Utrecht, The Netherlands.
2
Department of Radiology, University Medical Center Utrecht, 3584 CX Utrecht,
The Netherlands.
OBJECTIVE. Important decisions are made on the basis of CT angiographic measurements of aneurysm size obtained after endovascular abdominal aortic aneurysm repair; however, little is known about the variability of these measurements. We evaluated the variability of CT angiographic measurements of aneurysm size obtained after endovascular abdominal aortic aneurysm repair.
MATERIALS AND METHODS. Thirty CT angiographic data sets were randomly chosen from 91 sets, including preoperative, postoperative, and follow-up CT images. All images were obtained according to a standardized acquisition protocol. On a workstation, three parameters were measured: maximum aneurysm diameter, maximum aneurysm cross-sectional area, and aneurysm volume. All data sets were measured twice by two investigators in a random order. The difference of each pair of measurements was plotted against the mean value. The mean difference and its standard deviation were calculated with a repeatability coefficient.
RESULTS. The intraobserver repeatability coefficient for observer 1 was 3.8 mm for diameter, 201.7 mm2 for cross-sectional area, and 5.6 mL for volume. For observer 2, these figures were 3.0 mm, 219.0 mm2, and 8.1 mL, respectively. The interobserver repeatability coefficients were 3.9 mm, 236.2 mm2, and 10.3 mL.
CONCLUSION. Determination of the repeatability coefficient of aneurysm size measurements obtained after endovascular abdominal aortic aneurysm repair provides a good description of precision.
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
S. W. Stavropoulos and S. R. Charagundla Imaging Techniques for Detection and Management of Endoleaks after Endovascular Aortic Aneurysm Repair Radiology, June 1, 2007; 243(3): 641 - 655. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Schoder, F. Cartes-Zumelzu, M. Grabenwoger, M. Cejna, M. Funovics, C. G. Krenn, D. Hutschala, F. Wolf, S. Thurnher, G. Kretschmer, et al. Elective Endovascular Stent-Graft Repair of Atherosclerotic Thoracic Aortic Aneurysms:Clinical Results and Midterm Follow-Up Am. J. Roentgenol., March 1, 2003; 180(3): 709 - 715. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |