1 All authors: Department of Clinical Radiology, University of Muenster, Albert-Schweitzer-Str. 33, D-48129 Muenster, Germany.
OBJECTIVE. Three-dimensional contrast-enhanced MR angiography was performed to study MR characteristics of Hemobahn devices.
MATERIALS AND METHODS. Changes in endoluminal signal intensities and the precision of the endoluminal diameter measurement were investigated in phantom studies for different concentrations of gadopentetate dimeglumine. Before and after the Hemobahn devices had been implanted, 10 patients with peripheral arterial occlusive disease were examined on MR imaging and three-dimensional contrast-enhanced MR angiography.
RESULTS. Phantom experiments using three-dimensional MR angiography showed stent-related signal void as a dark ring in the axial image orientation, providing a precise delineation of the stent-vessel border (mean endoluminal diameter, 8.2 mm; SD, 0.6 mm). Changes in endoluminal signal intensity were evaluated quantitatively. Stent-related artifacts did not compromise diagnostic imaging quality. All Hemobahn devices were found to be patent without migration of an implanted graft. In one patient, an extensive perigraft reaction (edema and contrast-enhanced perivascular tissue) was postinterventionally detected on MR imaging and corresponded to clinically evident postimplantation symptoms.
CONCLUSION. Three-dimensional contrast-enhanced MR angiography is a suitable tool to follow up the implantation of Hemobahn devices and to detect intra- and extraluminal abnormalities.
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
O W Hamer, I Borisch, C Paetzel, W R Nitz, J Seitz, S Feuerbach, and N Zorger In vitro evaluation of stent patency and in-stent stenoses in 10 metallic stents using MR angiography Br. J. Radiol., August 1, 2006; 79(944): 636 - 643. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. W. Hamer, T. Finkenzeller, I. Borisch, C. Paetzel, N. Zorger, S. Feuerbach, and W. Nitz In Vivo Evaluation of Patency and In-Stent Stenoses After Implantation of Nitinol Stents in Iliac Arteries Using MR Angiography Am. J. Roentgenol., November 1, 2005; 185(5): 1282 - 1288. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. D. Hagspiel, D. A. Leung, K. R. Nandalur, J. F. Angle, H. S. Dulai, D. J. Spinosa, A. H. Matsumoto, J. M. Christopher, H. Ahmed, and S. S. Berr Contrast-Enhanced MR Angiography at 1.5 T After Implantation of Platinum Stents: In Vitro and In Vivo Comparison with Conventional Stent Designs Am. J. Roentgenol., January 1, 2005; 184(1): 288 - 294. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. Spuentrup, A. Ruebben, M. Stuber, R. W. Gunther, and A. Buecker Metallic Renal Artery MR Imaging Stent: Artifact-free Lumen Visualization with Projection and Standard Renal MR Angiography Radiology, June 1, 2003; 227(3): 897 - 902. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Tello, K. R. Thomson, K. U. Juergens, B. Tombach, and P. Reimer Contrast-Enhanced MR Angiography of Endovascular Stents: An Oversight in Citation? Am. J. Roentgenol., December 1, 2001; 177 (6): 1483 - 1484. [Full Text] [PDF] |
||||
![]() |
A. Buecker, E. Spuentrup, A. Ruebben, and R. W. Gunther Artifact-Free In-Stent Lumen Visualization by Standard Magnetic Resonance Angiography Using a New Metallic Magnetic Resonance Imaging Stent Circulation, April 16, 2002; 105(15): 1772 - 1775. [Abstract] [Full Text] [PDF] |
||||