1
Department of Radiology, New York University Medical Center, 530 First Ave.,
HCC Basement-MRI, New York, NY 10016.
2
Siemens Medical Engineering, MRI Division, Karl-Schall-Str., 91052 Erlangen,
Germany.
3
Present address: Magnetic Resonance Imaging, Beth Israel Deaconess Medical
Center, Harvard Medical School, 330 Brookline Ave., Boston, MA 02215.
OBJECTIVE. Real-time interactive duplex MR imaging is a new phase-contrast MR imaging technique that enables the quantification and display of flow velocities in real time without the need for cardiac gating. We investigated the feasibility and reliability of the technique to assess hemodynamic information both in vitro and in vivo in the carotid arteries and in the venous sinuses.
SUBJECTS AND METHODS. Real-time interactive duplex MR measurements (TR/TE, 53/27; flip angle, 90°; encoding velocity, 100 or 150 cm/sec) were performed in vitro with a steady-flow phantom and in 10 healthy volunteers in whom common and internal carotid artery velocities were measured. In eight volunteers, velocity measurements were also performed in the superior sagittal sinus during both normal breathing and hyperventilation. Timevelocity plots were analyzed qualitatively and quantitatively and compared with findings from conventional segmented k-space phase-contrast MR imaging and Doppler sonography.
RESULTS. Velocity determinations for real-time duplex MR and conventional phase-contrast MR imaging showed an in vitro correlation of 0.99 and an in vivo correlation of 0.83 (carotid arteries) and 0.76 (venous sinus). Velocity measurements in the carotid arteries with real-time MR imaging were significantly lower than those obtained with conventional phase-contrast MR (averaged, 7.8%; p = 0.003) or sonography (23.7%, p < 0.001), likely because of volume averaging. Small but significant velocity changes occurring in the venous sinus during hyperventilation were reliably identified with both MR techniques.
CONCLUSION. Real-time interactive duplex MR imaging can be effectively applied in neurovascular imaging to obtain hemodynamic information.
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
J. Lotz, C. Meier, A. Leppert, and M. Galanski Cardiovascular Flow Measurement with Phase-Contrast MR Imaging: Basic Facts and Implementation RadioGraphics, May 1, 2002; 22(3): 651 - 671. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |