OBJECTIVE. We evaluated the association between clinical image quality and breast cancer occurrence within 24 months of a negative mammogram.
MATERIALS AND METHODS. We identified women with breast cancer who were younger than 40 years old and older and screened from January 1, 1988, through December 31, 1993. We retrospectively assigned Breast Imaging Reporting and Data System (BI-RADS) assessments to their screening mammogram. We classified cancers (invasive or ductal in situ) as "screen-detected" when found after positive assessments (BI-RADS codes 3, 4, and 5) and "interval-detected" when found after negative assessments (BI-RADS codes 1 and 2). One reviewer evaluated mediolateral oblique and craniocaudal views for all cancer cases using a 3-point scale (failure, borderline, pass) for each measure of clinical image quality (positioning, breast compression, contrast, exposure, noise, sharpness, artifacts, overall quality). We used separate logistic regression models and evaluated the odds of interval invasive cancer or invasive plus in situ cancer as a function of each measure of quality using "pass" as the referent group.
RESULTS. We found 492 screen-detected and 164 interval-detected cancers that met study criteria. Cancer detection (sensitivity) was highest (84%) among patients with proper breast positioning, but when images failed this measure (33.4%), sensitivity fell to 66.3%. After adjustment for age, film date, and breast density, interval-detected invasive cancers were more likely after images failing positioning (odds ratio, 2.57; 95% confidence interval, 1.28-5.52%). Failures in overall quality were also associated with interval cancers when cases of ductal carcinoma in situ (p = 0.037) were included.
CONCLUSION. Invasive breast cancer detection by mammography may be improved through attention to correct positioning.
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
P. A. Carney, E. Steiner, M. E. Goodrich, A. J. Dietrich, C. J. Kasales, J. E. Weiss, and T. MacKenzie Discovery of Breast Cancers Within 1 Year of a Normal Screening Mammogram: How Are They Found? Ann. Fam. Med, November 1, 2006; 4(6): 512 - 518. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. MINIGH Mammographic Film Artifacts. Radiol. Technol., May 1, 2006; 77(5): 389M - 402M. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Hofvind, P. Skaane, B. Vitak, H. Wang, S. Thoresen, L. Eriksen, H. Bjorndal, A. Braaten, and N. Bjurstam Influence of Review Design on Percentages of Missed Interval Breast Cancers: Retrospective Study of Interval Cancers in a Population-based Screening Program Radiology, November 1, 2005; 237(2): 437 - 443. [Abstract] [Full Text] [PDF] |
||||
![]() |
E. J. Aiello, D. S.M. Buist, E. White, and P. L. Porter Association between Mammographic Breast Density and Breast Cancer Tumor Characteristics Cancer Epidemiol. Biomarkers Prev., March 1, 2005; 14(3): 662 - 668. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. E. Hendrick, G. R. Cutter, E. A. Berns, C. Nakano, J. Egger, P. A. Carney, L. Abraham, S. H. Taplin, C. J. D'Orsi, W. Barlow, et al. Community-Based Mammography Practice: Services, Charges, and Interpretation Methods Am. J. Roentgenol., February 1, 2005; 184(2): 433 - 438. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. H. Taplin, L. Ichikawa, M. U. Yood, M. M. Manos, A. M. Geiger, S. Weinmann, J. Gilbert, J. Mouchawar, W. A. Leyden, R. Altaras, et al. Reason for Late-Stage Breast Cancer: Absence of Screening or Detection, or Breakdown in Follow-up? J Natl Cancer Inst, October 20, 2004; 96(20): 1518 - 1527. [Abstract] [Full Text] [PDF] |
||||
![]() |
D. S. M. Buist, P. L. Porter, C. Lehman, S. H. Taplin, and E. White Factors Contributing to Mammography Failure in Women Aged 40-49 Years J Natl Cancer Inst, October 6, 2004; 96(19): 1432 - 1440. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. G. Elmore, P. A. Carney, L. A. Abraham, W. E. Barlow, J. R. Egger, J. S. Fosse, G. R. Cutter, R. E. Hendrick, C. J. D'Orsi, P. Paliwal, et al. The Association Between Obesity and Screening Mammography Accuracy Arch Intern Med, May 24, 2004; 164(10): 1140 - 1147. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |