1 Department of Radiology, Osaka University Graduate School of Medicine, 2-2
Yamadaoka, Suita, Osaka, 565-0871, Japan.
2 Department of Radiology, Osaka University Hospital, Osaka, 565-0871,
Japan.
3 Department of Radiology, Vancouver Hospital and Health Sciences Centre and
University of British Columbia, 855 W. 12th Ave., Vancouver, B. C. V5Z 1M9,
Canada.
OBJECTIVE. The purpose of this study was to compare the quality of coronal multiplanar reconstructions with the quality of direct coronal thin-section multidetector CT (MDCT) scans.
MATERIALS AND METHODS. Axial multidetector CT (MDCT) scans were obtained through the entire lung in 10 normal autopsy lung specimens using an MDCT scanner. Four protocols were used: 0.5-mm collimation with a 0.5-mm reconstruction interval; 0.5-mm collimation with a 0.3-mm reconstruction interval; 1-mm collimation with a 0.5-mm reconstruction interval; and 2-mm collimation with a 1-mm reconstruction interval. Multiplanar reconstruction images with 0.5-mm slice thickness were obtained from the four types of data sets. Direct coronal thin-section CT of the same 10 autopsy lung specimens was performed using 0.5-mm scan collimation, a 0.3-mm reconstruction interval, a 25.6-cm field of view, and a 512 x 512 matrix. Two independent observers compared the image quality of each of the four coronal multiplanar reconstruction sets with that of direct coronal thin-section CT scans. The observers analyzed visualization of anatomic features and artifacts.
RESULTS. The total image quality of the multiplanar reconstructions obtained from 0.5-mm collimation data with or without 0.3-mm overlapping reconstruction was equal to that of direct coronal thin-section CT scans in all 20 interpretations. The image quality of multiplanar reconstruction images from 0.5-mm collimation data either with or without overlapping reconstruction was superior to multiplanar reconstruction images obtained from 1- or 2-mm collimation scans (p < 0.01, Fisher's exact test). Stairstep artifacts in multiplanar reconstructions using 0.5-mm collimation without overlapping reconstruction were equal to those with overlapping reconstruction and were fewer than those on 1- or 2-mm collimation (p < 0.01, Mann-Whitney U test).
CONCLUSION. The image quality of coronal multiplanar reconstructions from isotropic voxel data obtained using 0.5-mm collimation, with or without overlapping reconstruction, is similar to that of direct coronal thin-section CT scans.
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
E. Y. Lee, P. M. Boiselle, and R. H. Cleveland Multidetector CT Evaluation of Congenital Lung Anomalies Radiology, June 1, 2008; 247(3): 632 - 648. [Abstract] [Full Text] [PDF] |
||||
![]() |
E J CHOI, Y-W OH, S Y HAM, K Y LEE, and E-Y KANG Comparison between coronal reformatted images and direct coronal CT images of the swine lung specimen: assessment of image quality with 64-detector row CT Br. J. Radiol., June 1, 2008; 81(966): 463 - 467. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. La Fata, N. McLean, S.K. Wise, J.M. DelGaudio, and P.A. Hudgins CSF Leaks: Correlation of High-Resolution CT and Multiplanar Reformations with Intraoperative Endoscopic Findings AJNR Am. J. Neuroradiol., March 1, 2008; 29(3): 536 - 541. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Jaffe, L. C. Martin, C. M. Miller, K. M. Franklin, E. M. Merkle, W. M. Thompson, R. C. Nelson, D. M. DeLong, and E. K. Paulson Abdominal Pain: Coronal Reformations from Isotropic Voxels with 16-Section CT--Reader Lesion Detection and Interpretation Time Radiology, January 1, 2007; 242(1): 175 - 181. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Dodd, C. A. Souza, and N. L. Muller High-resolution MDCT of pulmonary septic embolism: evaluation of the feeding vessel sign. Am. J. Roentgenol., September 1, 2006; 187(3): 623 - 629. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Takahashi, B. Thompson, W. Stanford, Y. Sato, K. Nagasawa, H. Sato, M. Kubota, A. Kashiba, and H. Sugimori Visualization of normal pulmonary fissures on sagittal multiplanar reconstruction MDCT. Am. J. Roentgenol., August 1, 2006; 187(2): 389 - 397. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. D. Dodd, C. A. Souza, and N. L. Muller Conventional high-resolution CT versus helical high-resolution MDCT in the detection of bronchiectasis. Am. J. Roentgenol., August 1, 2006; 187(2): 414 - 420. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Jaffe, R. C. Nelson, G. A. Johnson, E. R. Lee, T. T. Yoshizumi, C. R. Lowry, A. B. Bullard, D. M. DeLong, and E. K. Paulson Optimization of Multiplanar Reformations from Isotropic Data Sets Acquired with 16-Detector Row Helical CT Scanner Radiology, January 1, 2006; 238(1): 292 - 299. [Abstract] [Full Text] [PDF] |
||||
![]() |
T. A. Jaffe, L. C. Martin, J. Thomas, A. R. Adamson, D. M. DeLong, and E. K. Paulson Small-Bowel Obstruction: Coronal Reformations from Isotropic Voxels at 16-Section Multi-Detector Row CT Radiology, December 1, 2005; 238(1): 135 - 142. [Abstract] [Full Text] [PDF] |
||||
![]() |
S. Matsuoka, Y. Kurihara, Y. Nakajima, H. Niimi, H. Ashida, and K. Kaneoya Serial Change in Airway Lumen and Wall Thickness at Thin-Section CT in Asymptomatic Subjects Radiology, February 1, 2005; 234(2): 595 - 603. [Abstract] [Full Text] [PDF] |
||||
![]() |
W K Chooi and S K Morcos High resolution volume imaging of airways and lung parenchyma with multislice CT Br. J. Radiol., December 1, 2004; 77(suppl_1): S98 - S105. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |