AJR Custom publishing of AJR articles and ARRS Cat. Course
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Sofka, C. M.
Right arrow Articles by Laskin, R.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Sofka, C. M.
Right arrow Articles by Laskin, R.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Hotlight (NEW!)
Right arrow
What's Hotlight?
AJR 2003; 180:1437-1441
© American Roentgen Ray Society


Sonography of Polyethylene Liners Used in Total Knee Arthroplasty

Carolyn M. Sofka1,2, Ronald S. Adler1,2 and Richard Laskin3

1 Department of Radiology and Imaging, Hospital for Special Surgery, 535 E. 70th St., New York, NY 10021.
2 Department of Radiology, Weill Medical College of Cornell University, New York, NY 10021.
3 Department of Orthopaedic Surgery, Hospital for Special Surgery, New York, NY 10021.

OBJECTIVE. We investigated the ability of sonography to reveal the polyethylene liner used in total knee arthroplasty with the hopes of establishing a possible relationship between the sonographic measurement of the actual thickness of the polyethylene liner and the estimated thickness based on conventional radiography.

MATERIALS AND METHODS. Twenty-four consecutive patients who were referred for Doppler screening for deep venous thrombosis after total knee arthroplasty were evaluated. The polyethylene liner was identified on sonography, and three measurements were obtained from four locations: anteromedial joint line (just medial to the midline incision), along the medial joint line, anterolateral joint line (just lateral to the midline incision), and along the lateral joint line. These sonographic measurements were compared with radiographic measurements of the radiolucent polyethylene liner and with the manufacturers' stated size of the polyethylene liner. Linear regression analyses were then performed.

RESULTS. The polyethylene liner is seen on sonography as a strong linear echogenic interface with posterior acoustic shadowing. Linear regression analyses showed a high correlation (r = 0.8) between the sonographic measurements and the radiographic measurements. A relatively poor correlation (r = 0.2) was noted between the manufacturers' stated size of the liner and the sonographic measurements.

CONCLUSION. We found that the polyethylene liner used in total knee arthroplasty can be clearly identified during sonographic evaluation of the knee. We also found a high correlation between the longitudinal measurement of the polyethylene liner with the thickness of the radiolucent polyethylene as measured on conventional radiographs. We propose that sonography could potentially be a useful noninvasive imaging modality to screen for subtle cases of polyethylene wear.


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
JBJSHome page
M. B. Collier, C. A. Engh Jr., K. M. Hatten, S. D. Ginn, T. M. Sheils, and G. A. Engh
Radiographic Assessment of the Thickness Lost from Polyethylene Tibial Inserts That Had Been Sterilized Differently
J. Bone Joint Surg. Am., July 1, 2008; 90(7): 1543 - 1552.
[Abstract] [Full Text] [PDF]


Home page
J Ultrasound MedHome page
R. Guillin, J.-L. Laporte, P. Sabouret, and E. Cardinal
Polyethylene Wear in Knee Arthroplasty: A New Sonographic Sign
J. Ultrasound Med., February 1, 2008; 27(2): 275 - 279.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2003 by the American Roentgen Ray Society.