1 All authors: Department of Diagnostic Radiology, Ruprecht-Karls-University of Heidelberg, Radiological University Hospital, Im Neuenheimer Feld 110, 69120 Heidelberg, Germany.
OBJECTIVE. To evaluate image quality of a large-area direct-readout flat-panel detector system in chest radiography, we conducted an observer preference study. A clinical comparative study was conducted of the flat-panel system versus the storage phosphor and standard film-screen systems.
MATERIALS AND METHODS. Routine chest radiographs (posteroanterior) of 30 patients that were obtained using flat-panel, storage phosphor, and film screen systems were compared. The visibility of 10 anatomic regions and the overall image quality criteria were rated independently by three radiologists using a 5-point scale. The significance of the differences in diagnostic performance was tested with a Wilcoxon's signed rank test. Dose measurements for the three modalities were performed.
RESULTS. The flat-panel radiography system showed an improved visibility in most anatomic structures when compared with a state-of-the-art conventional film-screen system and an equal visibility when compared with a storage phosphor system. The flat-panel system showed the greatest enhancement in the depiction of small detailed structures (p < 0.05) and achieved this with a reduction in overall radiation dose of more than 50%.
CONCLUSION. The visibility of anatomic structures provided by this flat-panel detector system is as good as if not better than that provided by conventional or storage phosphor systems while emitting a reduced radiation dose.
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
L J M Kroft, W J H Veldkamp, B J A Mertens, J-P A van Delft, and J Geleijns Dose reduction in digital chest radiography and perceived image quality Br. J. Radiol., December 1, 2007; 80(960): 984 - 988. [Abstract] [Full Text] [PDF] |
||||
![]() |
C S Moore, G P Liney, A W Beavis, and J R Saunderson A method to optimize the processing algorithm of a computed radiography system for chest radiography Br. J. Radiol., September 1, 2007; 80(957): 724 - 730. [Abstract] [Full Text] [PDF] |
||||
![]() |
M. Korner, C. H. Weber, S. Wirth, K.-J. Pfeifer, M. F. Reiser, and M. Treitl Advances in Digital Radiography: Physical Principles and System Overview RadioGraphics, May 1, 2007; 27(3): 675 - 686. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. M. Kroft, W. J. H. Veldkamp, B. J. A. Mertens, J. P. A. van Delft, and J. Geleijns Detection of simulated nodules on clinical radiographs: dose reduction at digital posteroanterior chest radiography. Radiology, November 1, 2006; 241(2): 392 - 398. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Bacher, P. Smeets, L. Vereecken, A. De Hauwere, P. Duyck, R. De Man, K. Verstraete, and H. Thierens Image quality and radiation dose on digital chest imaging: comparison of amorphous silicon and amorphous selenium flat-panel systems. Am. J. Roentgenol., September 1, 2006; 187(3): 630 - 637. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. Neofotistou, V. Tsapaki, S. Kottou, A. Schreiner-Karoussou, and E. Vano Does digital imaging decrease patient dose? A pilot study and review of the literature Radiat Prot Dosimetry, December 1, 2005; 117(1-3): 204 - 210. [Abstract] [Full Text] [PDF] |
||||
![]() |
O. W. Hamer, C. B. Sirlin, M. Strotzer, I. Borisch, N. Zorger, S. Feuerbach, and M. Volk Chest Radiography with a Flat-Panel Detector: Image Quality with Dose Reduction after Copper Filtration Radiology, November 1, 2005; 237(2): 691 - 700. [Abstract] [Full Text] [PDF] |
||||
![]() |
A. BRINDHABAN and K. AL KHALIFAH Radiation Dose in Pelvic Imaging Radiol. Technol., September 1, 2005; 77(1): 32 - 40. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. J. M. Kroft, W. J. H. Veldkamp, B. J. A. Mertens, M. V. Boot, and J. Geleijns Comparison of Eight Different Digital Chest Radiography Systems: Variation in Detection of Simulated Chest Disease Am. J. Roentgenol., August 1, 2005; 185(2): 339 - 346. [Abstract] [Full Text] [PDF] |
||||
![]() |
I D Honey, A Mackenzie, and D S Evans Investigation of optimum energies for chest imaging using film-screen and computed radiography Br. J. Radiol., May 1, 2005; 78(929): 422 - 427. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |