OBJECTIVE. Computer-aided detection (CAD) systems have been used successfully to detect malignant calcifications on mammography, with sensitivities ranging from 86% to 99%. Amorphous calcifications are a subset of small indistinct calcifications of intermediate concern that have a 20% likelihood of being malignant and that are frequently overlooked on mammography. The purpose of our study was to determine the sensitivity of one commercially available CAD system for detecting amorphous calcifications.
MATERIALS AND METHODS. A commercially available CAD system evaluated mammograms of 82 patients with 85 mammographically detected and histologically sampled groups of amorphous calcifications (21 malignant, 14 high risk, and 50 benign). The sensitivity of the system for detecting the calcifications on at least one image of the two-view mammographic examination (case sensitivity) and on each individual mammographic image (image sensitivity) was determined. Findings were correlated with results from large core needle biopsy or surgical excision in each case.
RESULTS. The CAD system detected amorphous calcifications in 43 of 85 cases (case sensitivity, 51%) and in 59 of 146 mammographic images (image sensitivity, 40%). The case sensitivities by histologic outcome were 57% for malignant calcifications, 29% for high-risk calcifications, and 54% for benign calcifications. An average of 2.0 false-positive marks were displayed per case.
CONCLUSION. The CAD sensitivity for malignant amorphous calcifications is markedly lower than previously reported for all malignant calcifications. Breast imaging radiologists who use CAD systems should continue to search diligently for these difficult-to-detect lesions.
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
R. L. Birdwell The Preponderance of Evidence Supports Computer-aided Detection for Screening Mammography Radiology, October 1, 2009; 253(1): 9 - 16. [Full Text] [PDF] |
||||
![]() |
L. E. Philpotts Can Computer-aided Detection Be Detrimental to Mammographic Interpretation? Radiology, October 1, 2009; 253(1): 17 - 22. [Full Text] [PDF] |
||||
![]() |
D. Georgian-Smith, R. H. Moore, E. Halpern, E. D. Yeh, E. A. Rafferty, H. A. D'Alessandro, M. Staffa, D. A. Hall, K. A. McCarthy, and D. B. Kopans Blinded Comparison of Computer-Aided Detection with Human Second Reading in Screening Mammography Am. J. Roentgenol., November 1, 2007; 189(5): 1135 - 1141. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. L. Ellis, A. A. Meade, M. A. Mathiason, K. M. Willison, and W. Logan-Young Evaluation of Computer-aided Detection Systems in the Detection of Small Invasive Breast Carcinoma Radiology, October 1, 2007; 245(1): 88 - 94. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. M. Ko, M. J. Nicholas, J. B. Mendel, and P. J. Slanetz Prospective assessment of computer-aided detection in interpretation of screening mammography. Am. J. Roentgenol., December 1, 2006; 187(6): 1483 - 1491. [Abstract] [Full Text] [PDF] |
||||
![]() |
F. M. Hall Computer-Aided Detection (CAD) of Amorphous Calcifications Am. J. Roentgenol., March 1, 2006; 186(3): 902 - 902. [Full Text] [PDF] |
||||
![]() |
M. S. Soo, S. Ghate, J. A. Baker, and E. Rosen Reply Am. J. Roentgenol., March 1, 2006; 186(3): 902 - 903. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |