AJR ARRS Membership
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Lee, J. M.
Right arrow Articles by Choi, B. I.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Lee, J. M.
Right arrow Articles by Choi, B. I.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
AJR 2005; 184:897-905
© American Roentgen Ray Society

Comparison of Renal Ablation with Monopolar Radiofrequency and Hypertonic-Saline-Augmented Bipolar Radiofrequency: In Vitro and In Vivo Experimental Studies

Jeong Min Lee1, Joon Koo Han1, Seung Hong Choi1, Se Hyung Kim1, Jae Young Lee1, Kyung Sook Shin2, Chang Jin Han1 and Byung Ihn Choi1

1 Department of Radiology and the Institute of Radiation Medicine, Seoul National University College of Medicine, 28 Yongon-dong, Chongno-gu, Seoul 110-744, South Korea.
2 Department of Radiology, Chungnam National University College of Medicine, 6 Munhwa-dong, Daejeon 301-747, Korea.

OBJECTIVE. We sought to determine whether hypertonic-saline (HS)-augmented bipolar radiofrequency ablation has advantages over monopolar radiofrequency ablation for creating larger areas of coagulation necrosis in the kidney.

MATERIALS AND METHODS. Using a 200-W generator and bipolar perfused-cooled electrodes or a monopolar cooled-tip electrode, we performed 14 radiofrequency ablations in explanted bovine kidneys. Radiofrequency was applied in standard monopolar (n = 7) or bipolar (n = 7) modes at 100 W for 10 min. In the bipolar mode, the perfused-cooled electrodes were placed at interelectrode distances of 3 cm, and a 6% sodium chloride solution was instilled into tissue at a rate of 2 mL/min through the electrodes. For in vivo experiments, either monopolar (n = 7) or HS-augmented bipolar (n = 7) radiofrequency ablation was performed in the lower pole of canine kidneys. Three days after the procedure, contrast-enhanced CT scans were obtained to evaluate the volumes of the ablation regions, and the kidneys were harvested for gross measurements. Technical parameters such as changes in impedance and current during radiofrequency ablation and dimensions of the thermal ablation zones were compared between the two groups.

RESULTS. In ex vivo and in vivo experiments, the frequency of the pulsed radiofrequency application caused by rises in impedance was higher in the monopolar mode than in the bipolar mode during the application of radiofrequency energy. The in vivo study showed that the bipolar radiofrequency ablation allowed larger mean current flows than the monopolar radiofrequency ablation (i.e., mean ± SD, 1,654 ± 144 mA vs 967 ± 597 mA) (p < 0.05). Ex vivo studies revealed that the volumes of bipolar radiofrequency-induced ablation regions were substantially larger than those of monopolar radiofrequency-induced ablation regions (26.1 ± 10.5 cm3 vs 10.2 ± 4.2 cm3). In vivo studies showed bipolar radiofrequency ablation achieved larger coagulation necrosis than monopolar radiofrequency (3.2 ± 0.3 cm vs 2.4 ± 0.4 cm) (p < 0.05). This was confirmed by the measured volume of nonenhancing area on contrast-enhanced CT (20.4 ± 6.4 cm3 vs 13.5 ± 6.0 cm3).

CONCLUSION. HS-augmented bipolar radiofrequency ablation using perfused-cooled electrodes shows better performance in creating coagulation necrosis than monopolar radiofrequency ablation in the kidney of an animal model.


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
Br. J. Radiol.Home page
J M Lee, J K Han, J M Chang, S Y Chung, S H Kim, J Y Lee, and B I Choi