AJR Women's Imaging Online
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Suzuki, S.
Right arrow Articles by Kaminaga, T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Suzuki, S.
Right arrow Articles by Kaminaga, T.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
AJR 2005; 184:1940-1944
© American Roentgen Ray Society

Accuracy of Automated CT Angiography Measurement of Vascular Diameter in Phantoms: Effect of Size of Display Field of View, Density of Contrast Medium, and Wall Thickness

Shigeru Suzuki, Shigeru Furui and Tatsuro Kaminaga

Department of Radiology, Teikyo University School of Medicine, 2-11-1, Kaga, Itabashi-ku, Tokyo 173-8605, Japan.

OBJECTIVE. The objective of this study was to assess the effect of wall thickness, density of intravascular contrast material, and size of the display field of view on the accuracy of measurements of vascular diameter in phantoms yielded by automated software for CT angiography.

MATERIALS AND METHODS. Vascular models with three wall thicknesses (1.0, 0.8, and 0.5 mm) and an inner diameter of approximately 4 mm were filled with contrast material of three different densities (198, 270, and 350 H) and scanned with helical CT. Three sizes of display field of view (10, 15, and 20 cm) were used. We evaluated the measurement error of the automated software, which was defined as the difference between the diameter measurement of the automated software and the true inner diameter of the vascular model. Statistical analysis involved three-way analysis of variance with repeated measures.

RESULTS. There were significant differences in the measurement errors among the three wall thicknesses of the vascular model, three densities of intravascular contrast material, and three sizes of display field of view. The overall measurement errors progressively increased with larger sizes of display field of view (p < 0.01) and with lower densities of intravascular contrast material (p < 0.001). The measurement errors tended to progressively increase as the thickness of the wall of the vascular models increased.

CONCLUSION. The accuracy of the diameter measurements by automated software for CT angiography was affected by the size of the display field of view, intravascular density of the contrast material, and wall thickness of the vessel. It is necessary to consider the effects of these factors on the diameter measurements of small arteries.


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
Am. J. Roentgenol.Home page
J. P. Villablanca, F. J. Rodriguez, T. Stockman, S. Dahliwal, M. Omura, S. Hazany, and J. Sayre
MDCT Angiography for Detection and Quantification of Small Intracranial Arteries: Comparison with Conventional Catheter Angiography
Am. J. Roentgenol., February 1, 2007; 188(2): 593 - 602.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2005 by the American Roentgen Ray Society.