OBJECTIVE. This study tests various acquisition and reconstruction protocols for MDCT of the wrist to determine the optimal protocol for obtaining diagnostic image quality with minimal radiation exposure.
MATERIALS AND METHODS. Thirty anatomic specimens were examined with an MDCT collimation of 4.0 x 1.0 mm and 2.0 x 0.5 mm (80, 120, and 140 kV; 80, 100, 130, 160, and 200 mA; rotation time, 0.5 0.75, 1.0 sec; pitch, 1.0, 1.3, 1.5, and 2.0). Coronal images were reconstructed using a slice thickness of 0.5, 1.0, and 2.0 mm with 60% overlap. Three observers evaluated all images independently for gross and fine anatomic detail. Diagnostic confidence was tested using Shrout-Fleiss intraclass correlation coefficients. Interobserver agreement was assessed by Kappa statistics and the Kruskal-Wallis test.
RESULTS. Fine anatomic detail was best presented in 0.5-mm or
1.00-mm reconstructions based on a 2.0 x 0.5 mm acquisition. A rotation
time of
0.75 sec resulted in fewer artifacts; a significant dose
reduction was achieved with 80 kV and 100 mA at the expense of somewhat
increased noise, but without significant loss of anatomic detail in bone
presentation. Artifacts were tolerable with a pitch of 1.5 or less.
CONCLUSION. MDCT at the described optimal settings allows significant dosage reduction without sacrificing image quality. An acquisition and reconstruction thickness of 0.5 mm results in the best depiction of anatomic detail. A reconstruction thickness of 1.0 mm with a reconstruction interval of 0.5 mm represents a good trade-off between noise and resolution when using low-dose protocols.
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
T. Murase, K. Oka, H. Moritomo, A. Goto, H. Yoshikawa, and K. Sugamoto Three-Dimensional Corrective Osteotomy of Malunited Fractures of the Upper Extremity with Use of a Computer Simulation System J. Bone Joint Surg. Am., November 1, 2008; 90(11): 2375 - 2389. [Abstract] [Full Text] [PDF] |
||||
![]() |
N. C. Dalrymple, S. R. Prasad, F. M. El-Merhi, and K. N. Chintapalli Price of Isotropy in Multidetector CT RadioGraphics, January 1, 2007; 27(1): 49 - 62. [Abstract] [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |