AJR ARRS PQI
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Li, J.
Right arrow Articles by Kalra, M. K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Li, J.
Right arrow Articles by Kalra, M. K.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Hotlight (NEW!)
Right arrow
What's Hotlight?
DOI:10.2214/AJR.06.0370
AJR 2007; 188:547-552
© American Roentgen Ray Society


Original Research

Automatic Patient Centering for MDCT: Effect on Radiation Dose

Jianhai Li1, Unni K. Udayasankar1, Thomas L. Toth2, John Seamans2, William C. Small1 and Mannudeep K. Kalra1,3

1 Department of Radiology, Emory University School of Medicine, 1364 Clifton Rd. NE, Atlanta, GA 30322.
2 GE Healthcare, Waukesha, WI 53188.
3 Present address: Department of Radiology, Massachusetts General Hospital, 55 Fruit St., Boston, MA 02114.

OBJECTIVE. The purpose of this study was to determine with phantom and patient imaging the effect of an automatic patient-centering technique on the radiation dose associated with MDCT.

SUBJECTS AND METHODS. A 32-cm CT dose index (CTDI) phantom was scanned with 64-MDCT in three positions: gantry isocenter and 30 and 60 mm below the isocenter of the scanner gantry. In each position, surface, peripheral, and volume CTDIs were estimated with a standard 10-cm pencil ionization chamber. The institutional review board approved the study with 63 patients (36 men, 27 women; mean age, 51 years; age range, 22-83 years) undergoing chest (n = 18) or abdominal (n = 45) CT using the z-axis automatic exposure control technique. Each patient was positioned according to the region being scanned and then was centered in the gantry. Before scanning of a patient, automatic centering software was used to estimate patient off-centering and percentage of dose reduction with optimum recentering. Data were analyzed with linear correlation and the Student's t test.

RESULTS. Peripheral and surface CTDIs increased approximately 12-18% with 30-mm off-center distance and 41-49% with 60-mm off-center distance. Approximately 95% (60/63) of patients were not positioned accurately in the gantry isocenter. The mean radiation dose saving with automatic centering of all patients was 13.0% ± 0.9% (range, 2.6-29.9%). There was strong correlation between off-center distance and percentage of surface CTDI reduction with recentering of patients in the gantry isocenter (r2 = 0.85, p < 0.0001).

CONCLUSION. Surfaces doses can be reduced if radiologic technologists can better center patients within the CT gantry. Automatic centering technique can help in optimum patient centering and result in as much as 30% reduction in surface dose.

Keywords: CT • positioning • radiation dose • technology


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
Am. J. Roentgenol.Home page
K. Matsubara, K. Koshida, K. Ichikawa, M. Suzuki, T. Takata, T. Yamamoto, and O. Matsui
Misoperation of CT Automatic Tube Current Modulation Systems with Inappropriate Patient Centering: Phantom Studies
Am. J. Roentgenol., April 1, 2009; 192(4): 862 - 865.
[Abstract] [Full Text] [PDF]


Home page
Am. J. Roentgenol.Home page
J. R. Mayo and J. A. Leipsic
Radiation Dose in Cardiac CT
Am. J. Roentgenol., March 1, 2009; 192(3): 646 - 653.
[Abstract] [Full Text] [PDF]


Home page
Proc Am Thorac SocHome page
J. R. Mayo
Radiation Dose Issues in Longitudinal Studies Involving Computed Tomography
Proceedings of the ATS, December 15, 2008; 5(9): 934 - 939.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2007 by the American Roentgen Ray Society.