AJR Women's Imaging Online
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Michaely, H. J.
Right arrow Articles by Schoenberg, S. O.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Michaely, H. J.
Right arrow Articles by Schoenberg, S. O.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
DOI:10.2214/AJR.06.0567
AJR 2007; 188:919-926
© American Roentgen Ray Society


Original Research

Semiquantitative Assessment of First-Pass Renal Perfusion at 1.5 T: Comparison of 2D Saturation Recovery Sequences With and Without Parallel Imaging

Henrik J. Michaely1, Harald Kramer1, Niels Oesingmann2, Klaus-Peter Lodemann3, Maximilian F. Reiser1 and Stefan O. Schoenberg1

1 Department of Clinical Radiology, University of Munich, Grosshadern-Campus, Marchionistrasse 15, Munich, Germany, 81377.
2 Siemens Medical Solutions, Malvern, PA.
3 Bracco-Altana Pharma, Konstanz, Germany.

OBJECTIVE. The purpose of this study was to assess the feasibility and reliability of measurements performed with true fast imaging with steady-state free precession (FISP) and turbo fast low-angle shot (FLASH) sequences with parallel imaging compared with those obtained with turbo FLASH sequences without parallel imaging in first-pass renal perfusion MRI.

SUBJECTS AND METHODS. The subjects in this prospective study were 15 healthy men who volunteered to undergo MRI for acquisition of renal perfusion measurements. Imaging was performed at 1.5 T with the following three techniques after administration of gadobenate dimeglumine at 4 mL/s: saturation recovery (SR) turbo FLASH sequences without parallel imaging, SR turbo FLASH sequences with parallel imaging, and SR true FISP sequences. The spatial resolution was 2.3 x 2.6 x 8 mm with a temporal resolution of four slices per second (turbo FLASH without parallel imaging and true FISP) or six slices per second (turbo FLASH with parallel imaging). The semiquantitative perfusion parameters mean transit time and maximal upslope were determined. Signal-to-noise ratio (SNR), delta ratio, and time to maximal signal intensity also were determined. Image quality was rated in consensus.

RESULTS. Image quality was best for turbo FLASH sequences without parallel imaging compared with true FISP and turbo FLASH sequences with parallel imaging. True FISP sequences yielded the highest baseline SNR (26.7) but the lowest delta ratio (3.2). Turbo FLASH sequences without and with parallel imaging had significantly lower SNRs (9.6 and 9.3) and significantly higher delta ratios (5.1 and 5.0). The first-pass perfusion parameters mean transit time and time to maximal signal intensity were independent of the technique used.

CONCLUSION. It seems that at 1.5 T, turbo FLASH sequences without parallel imaging are the best approach to renal first-pass perfusion imaging.

Keywords: genitourinary tract imaging • kidney • MRI technique • perfusion-weighted MRI • renal perfusion