OBJECTIVE. The purpose of this study was to evaluate the effect of mean heart rate and heart rate variability on the image quality of dual-source CT coronary angiography.
SUBJECTS AND METHODS. Eighty patients underwent dual-source CT coronary angiography. Thirteen data sets were reconstructed in 5% steps from 20-80% of the R-R interval. Heart rate variability was calculated as SD of mean heart rate. Two independent blinded reviewers assessed the image quality of each segment.
RESULTS. Mean heart rate was 65.3 ± 13.9 (SD) beats per minute (bpm) (range, 35-99 bpm) with a variability of 3.4 ± 4.1 bpm (range, 0.4-17.5 bpm). Image quality was sufficient for diagnosis for 97.8% (1,043/1,066) of arterial segments. No significant correlation was found between mean heart rate and image quality in any segment or any coronary artery. No significant correlation was found between heart rate variability and image quality in any segment, the right coronary artery, or the left anterior descending artery, but there was a significant (p < 0.05) correlation in the left circumflex artery. A significant correlation (p < 0.05) between overall image quality was found for mean and variability of heart rate as shared predictors, the latter having a greater contribution.
CONCLUSION. The overall image quality of dual-source CT coronary angiography is sufficient for diagnosis within a wide range of mean heart rates and variability of heart rates. Only heart rates that are both high and variable significantly deteriorate image quality, but the quality remains adequate for diagnosis.
Keywords: coronary angiography dual-source CT heart rate image quality
![]()
CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:
![]() |
W. P. Shuman, K. R. Branch, J. M. May, L. M. Mitsumori, J. N. Strote, B. H. Warren, T. J. Dubinsky, D. W. Lockhart, and J. H. Caldwell Whole-Chest 64-MDCT of Emergency Department Patients with Nonspecific Chest Pain: Radiation Dose and Coronary Artery Image Quality with Prospective ECG Triggering Versus Retrospective ECG Gating Am. J. Roentgenol., June 1, 2009; 192(6): 1662 - 1667. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Donnino, J. E. Jacobs, J. V. Doshi, E. M. Hecht, D. C. Kim, J. S. Babb, and M. B. Srichai Dual-Source Versus Single-Source Cardiac CT Angiography: Comparison of Diagnostic Image Quality Am. J. Roentgenol., April 1, 2009; 192(4): 1051 - 1056. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. Sundaram, S. Patel, N. Bogot, and E. A. Kazerooni Anatomy and Terminology for the Interpretation and Reporting of Cardiac MDCT: Part 1, Structured Report, Coronary Calcium Screening, and Coronary Artery Anatomy Am. J. Roentgenol., March 1, 2009; 192(3): 574 - 583. [Abstract] [Full Text] [PDF] |
||||
![]() |
B. A. Herzog, L. Husmann, N. Burkhard, O. Gaemperli, I. Valenta, F. Tatsugami, C. A. Wyss, U. Landmesser, and P. A. Kaufmann Accuracy of low-dose computed tomography coronary angiography using prospective electrocardiogram-triggering: first clinical experience Eur. Heart J., December 2, 2008; 29(24): 3037 - 3042. [Abstract] [Full Text] [PDF] |
||||
![]() |
H. Matsumoto, T. Kondo, S. Watanabe, R. Kikumoto, T. Shimada, Y. Hiraoka, and K. Ueda ECG-Edited Middiastolic Phase Reconstruction Improves Image Quality at 64-MDCT Coronary Angiography of Patients with Atrial Fibrillation Am. J. Roentgenol., December 1, 2008; 191(6): 1659 - 1666. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stolzmann, H. Scheffel, S. Leschka, A. Plass, S. Baumuller, B. Marincek, and H. Alkadhi Influence of Calcifications on Diagnostic Accuracy of Coronary CT Angiography Using Prospective ECG Triggering Am. J. Roentgenol., December 1, 2008; 191(6): 1684 - 1689. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stolzmann, S. Leschka, H. Scheffel, T. Krauss, L. Desbiolles, A. Plass, M. Genoni, T. G. Flohr, S. Wildermuth, B. Marincek, et al. Dual-Source CT in Step-and-Shoot Mode: Noninvasive Coronary Angiography with Low Radiation Dose Radiology, October 1, 2008; 249(1): 71 - 80. [Abstract] [Full Text] [PDF] |
||||
![]() |
G. Bastarrika, C. N. De Cecco, M. Arraiza, S. Mastrobuoni, J. C. Pueyo, M. Ubilla, and G. Rabago Dual-Source CT for Visualization of the Coronary Arteries in Heart Transplant Patients with High Heart Rates Am. J. Roentgenol., August 1, 2008; 191(2): 448 - 454. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. P. Shuman, K. R. Branch, J. M. May, L. M. Mitsumori, D. W. Lockhart, T. J. Dubinsky, B. H. Warren, and J. H. Caldwell Prospective versus Retrospective ECG Gating for 64-Detector CT of the Coronary Arteries: Comparison of Image Quality and Patient Radiation Dose Radiology, August 1, 2008; 248(2): 431 - 437. [Abstract] [Full Text] [PDF] |
||||
![]() |
L. Husmann, I. Valenta, O. Gaemperli, O. Adda, V. Treyer, C. A. Wyss, P. Veit-Haibach, F. Tatsugami, G. K. von Schulthess, and P. A. Kaufmann Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating Eur. Heart J., January 2, 2008; 29(2): 191 - 197. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. Colletti Coronary CT Angiography Without {beta}-Blockers Am. J. Roentgenol., December 1, 2007; 189(6): 1324 - 1325. [Full Text] [PDF] |
||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |