AJR F and L Medical Products: Radiation Protection & More
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Blum, M. B.
Right arrow Articles by Funovics, M. A.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Blum, M. B.
Right arrow Articles by Funovics, M. A.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Hotlight (NEW!)
Right arrow
What's Hotlight?
DOI:10.2214/AJR.07.2501
AJR 2007; 189:1238-1242
© American Roentgen Ray Society


Original Research

Quantification and Detectability of In-Stent Stenosis with CT Angiography and MR Angiography in Arterial Stents In Vitro

Melanie B. Blum1, Maria Schmook, Rüdiger Schernthaner, Gundula Edelhauser, Stefan Puchner, Johannes Lammer and Martin A. Funovics

1 All authors: Division of Angiography and Interventional Radiology, Department of Radiology, Medical University of Vienna, Währinger Gürtel 18-20, 1090 Vienna, Austria.

OBJECTIVE. The purpose of this study was to compare CT angiography (CTA) and MR angiography (MRA) for the detectability of 75% and 95% stenoses in phantoms using six different stents.

MATERIALS AND METHODS. Six different stents (Expander, Hemobahn, SelfX, Smart, Symphony, and Wallstent) were inserted into tubes filled with contrast agent (ioversol or gadoteric acid). To mimic stenoses of 75% and 95% of the patent lumen, 8-mm-diameter nylon cylinders were bored in the central axis (2 mm and 4 mm, respectively) and placed into the stent lumen. Intensity profiles across stenoses on 2-mm coronal reformatted sections of CTA or MRA were compared, and the detectability of the residual lumen was assessed using a subjective score.

RESULTS. CTA showed relative in-stent signal attenuation for the in-stent stenoses of the tested stents ranging from 75% to 100% of the signal intensity of the control. SelfX and Symphony showed further shading of the residual lumen due to beam-hardening artifacts. Overestimation of stenosis was associated with low-grade stenoses in which the border of the lumen was closer to the stent struts. MRA showed relative in-stent signal attenuation of the in-stent stenoses ranging from 30% to 100% of the signal intensity of the control. Strut thickness tended to correlate with higher attenuation at CT.

CONCLUSION. CTA may be more suitable for differentiation between 95% stenosis and occlusion; MRA has higher sensitivity in detecting 75% stenoses. Strut thickness and mesh size did not prove to be significant predictors for signal attenuation or overall image quality.

Keywords: CT • in-stent restenosis • in vitro • MRI • stenosis