AJR Women's Imaging Online
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Schindera, S. T.
Right arrow Articles by Yoshizumi, T. T.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Schindera, S. T.
Right arrow Articles by Yoshizumi, T. T.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Hotlight (NEW!)
Right arrow
What's Hotlight?
DOI:10.2214/AJR.07.2891
AJR 2008; 190:W100-W105
© American Roentgen Ray Society


Original Research

Effect of Patient Size on Radiation Dose for Abdominal MDCT with Automatic Tube Current Modulation: Phantom Study

Sebastian T. Schindera1,2, Rendon C. Nelson2, Thomas L. Toth3, Giao T. Nguyen4, Greta I. Toncheva4, David M. DeLong2 and Terry T. Yoshizumi2,4

1 Department of Diagnostic Radiology, University Hospital of Bern, Bern, Switzerland.
2 Department of Radiology, Duke University Medical Center, Box 3808, Erwin Rd., Durham, NC 27710.
3 GE Healthcare Inc., Waukesha, WI.
4 Division of Radiation Safety, Duke University Medical Center, Durham, NC.

OBJECTIVE. The purpose of this study was to evaluate in a phantom study the effect of patient size on radiation dose for abdominal MDCT with automatic tube current modulation.

MATERIALS AND METHODS. One or two 4-cm-thick circumferential layers of fatequivalent material were added to the abdomen of an anthropomorphic phantom to simulate patients of three sizes: small (cross-sectional dimensions, 18 x 22 cm), average size (26 x 30 cm), and oversize (34 x 38 cm). Imaging was performed with a 64-MDCT scanner with combined z-axis and xy-axis tube current modulation according to two protocols: protocol A had a noise index of 12.5 H, and protocol B, 15.0 H. Radiation doses to three abdominal organs and the skin were assessed. Image noise also was measured.

RESULTS. Despite increasing patient size, the image noise measured was similar for protocol A (range, 11.7–12.2 H) and protocol B (range, 13.9–14.8 H) (p > 0.05). With the two protocols, in comparison with the dose of the small patient, the abdominal organ doses of the average-sized patient and the oversized patient increased 161.5–190.6%and 426.9–528.1%, respectively (p < 0.001). The skin dose increased as much as 268.6% for the average-sized patient and 816.3% for the oversized patient compared with the small patient (p < 0.001).

CONCLUSION. Oversized patients undergoing abdominal MDCT with tube current modulation receive significantly higher doses than do small patients. The noise index needs to be adjusted to the body habitus to ensure dose efficiency.

Keywords: automatic tube current modulation • CT • image quality • obesity • radiation dose