AJR ARRS: Your Link to CME
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Takahashi, N.
Right arrow Articles by McCollough, C. H.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Takahashi, N.
Right arrow Articles by McCollough, C. H.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Hotlight (NEW!)
Right arrow
What's Hotlight?
DOI:10.2214/AJR.07.3154
AJR 2008; 190:1169-1173
© American Roentgen Ray Society


Original Research

Dual-Energy CT Iodine-Subtraction Virtual Unenhanced Technique to Detect Urinary Stones in an Iodine-Filled Collecting System: A Phantom Study

Naoki Takahashi1, Robert P. Hartman1, Terri J. Vrtiska1, Akira Kawashima1, Andrew N. Primak1, Oleksandr P. Dzyubak1, Jayawant N. Mandrekar2, Joel G. Fletcher1 and Cynthia H. McCollough1

1 Department of Radiology, Mayo Clinic, 200 First St. SW, Rochester, MN 55905.
2 Division of Biostatistics, Mayo Clinic, Rochester, MN.

OBJECTIVE. The objective of our study was to evaluate the feasibility of virtual unenhanced images reconstructed from a dual-energy CT scan to depict urinary stones in an iodine solution in a phantom study.

MATERIALS AND METHODS. Twenty urinary stones of different sizes (1.4–4.2 mm in short-axis diameter) were placed in plastic containers. The containers were consecutively filled with different concentrations of iodine solution (21, 43, 64, 85, and 107 mg/dL; CT attenuation value range, 510–2,310 H at 120 kVp). Dual-energy CT was repeated with 80–140 and 100–140 kVp pairs, two collimation–slice thickness combinations, and the presence or absence of a 4-cm-thick oil gel around the phantom. The iodine-subtraction virtual unenhanced images were reconstructed using commercial software. The images were evaluated by three radiologists in consensus for the visibility of the stones and the presence of residual nonsubtracted iodine. Stone visibility rates were compared between the 80–140 and 100–140 kVp pairs and the five different iodine concentrations.

RESULTS. Stone visibility rates with the 80–140 kVp pair were 99%, 93%, 96%, 94%, and 3% and those with the 100–140 kVp pair were 98%, 95%, 99%, 94%, and 99% for an iodine concentration of 21, 43, 64, 85, and 107 mg/dL, respectively. The poor visibility rate with 80–140 kVp and 107 mg/dL iodine concentration was due to the failure of iodine subtraction.

CONCLUSION. Dual-energy CT iodine-subtraction virtual unenhanced technique is capable of depicting urinary stones in iodine solutions of a diverse range of concentrations in a phantom study.

Keywords: dual-energy CT • genitourinary imaging • iodine-subtraction imaging technique • reconstructed images • urinary stones


Add to CiteULike CiteULike   Add to Complore Complore   Add to Connotea Connotea   Add to Del.icio.us Del.icio.us   Add to Digg Digg   Add to Reddit Reddit   Add to Technorati Technorati    What's this?


This article has been cited by other articles:


Home page
Am. J. Roentgenol.Home page
B. M. Yeh, J. A. Shepherd, Z. J. Wang, H. Seong Teh, R. P. Hartman, and S. Prevrhal
Dual-Energy and Low-kVp CT in the Abdomen
Am. J. Roentgenol., July 1, 2009; 193(1): 47 - 54.
[Abstract] [Full Text] [PDF]


Home page
RadiologyHome page
L.-J. Zhang, Y.-E Zhao, S.-Y. Wu, B. M. Yeh, C.-S. Zhou, X.-B. Hu, Q.-J. Hu, and G.-M. Lu
Pulmonary Embolism Detection with Dual-Energy CT: Experimental Study of Dual-Source CT in Rabbits
Radiology, July 1, 2009; 252(1): 61 - 70.
[Abstract] [Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 2008 by the American Roentgen Ray Society.