AJR ARRS PQI
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Google Scholar
Right arrow Articles by Kim, S.
Right arrow Articles by Yoshizumi, T.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kim, S.
Right arrow Articles by Yoshizumi, T.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Hotlight (NEW!)
Right arrow
What's Hotlight?
DOI:10.2214/AJR.08.1713
AJR 2009; 192:1726-1730
© American Roentgen Ray Society


Original Research

Kerma Area Product Method for Effective Dose Estimation During Lumbar Epidural Steroid Injection Procedures: Phantom Study

Sangroh Kim1, Greta Toncheva2,3, Colin Anderson-Evans2, Billy K. Huh4, Linda Gray3 and Terry Yoshizumi2,3

1 Medical Physics Graduate Program, Duke University Medical Center, Durham, NC.
2 Division of Radiation Safety, Duke University Medical Center, Box 3155, Durham, NC 27710.
3 Department of Radiology, Duke University Medical Center, Durham, NC.
4 Department of Anesthesiology, Duke University Medical Center, Durham, NC.

OBJECTIVE. The purpose of this study was to derive from the kerma area product the dose conversion coefficient for estimating the effective dose for lumbar epidural steroid injection procedures.

MATERIALS AND METHODS. A mobile fluoroscopy system was used for fluoroscopic imaging guidance of lumbar epidural steroid injection procedures. For acquisition of organ dose measurements, 20 diagnostic metal oxide semiconductor field effect transistor detectors were placed at each organ in an anthropomorphic phantom of a man, and these detectors were attached to four mobile metal oxide semiconductor field effect transistor wireless bias supplies to obtain the organ dose readings. The kerma area product was recorded from the system console and independently validated with an ion chamber and therapeutic x-ray film. Fluoroscopy was performed on the phantom for 10 minutes for acquisition of the dose rate for each organ, and the average clinical procedure time was multiplied by each organ dose rate for acquisition of individual organ doses. The effective dose was computed by summing the product of each organ dose and the corresponding tissue weighting factor from International Commission on Radiologic Protection publication 60.

RESULTS. The effective dose was computed as 0.93 mSv for an average lumbar epidural steroid injection procedure (fluoroscopic time, 40.7 seconds). The corresponding kerma area product was 2.80 Gy·cm2. The dose conversion coefficient was derived as 0.33 mSv/(Gy·cm2).

CONCLUSION. The effective dose for lumbar epidural steroid injection can be easily estimated by multiplying the derived dose conversion coefficient by the console-displayed kerma area product.

Keywords: anthropomorphic phantom • effective dose • kerma area product • lumbar epidural steroid injection • MOSFET