AJR Women's Imaging Online
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Google Scholar
Right arrow Articles by Jaffe, T. A.
Right arrow Articles by Ravin, C. E.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Jaffe, T. A.
Right arrow Articles by Ravin, C. E.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Hotlight (NEW!)
Right arrow
What's Hotlight?
DOI:10.2214/AJR.09.2330
AJR 2009; 193:1141-1147
© American Roentgen Ray Society


Original Research

Radiation Dose for Body CT Protocols: Variability of Scanners at One Institution

Tracy A. Jaffe1, Terry T. Yoshizumi, Greta Toncheva, Colin Anderson-Evans, Carolyn Lowry, Chad M. Miller, Rendon C. Nelson and Carl E. Ravin

1 All authors: Department of Radiology, Duke University Medical Center, Erwin Rd., Box 3808, Durham, NC 27710.

OBJECTIVE. The objective of our study was to determine, using an anthropomorphic phantom, whether patients are subject to variable radiation doses based on scanner assignment for common body CT studies.

MATERIALS AND METHODS. Twenty metal oxide semiconductor field effect transistor dosimeters were placed in a medium-sized anthropomorphic phantom of a man. Pulmonary embolism and chest, abdomen, and pelvis protocols were used to scan the phantom three times with GE Healthcare scanners in four configurations and one 64-MDCT Siemens Healthcare scanner. Organ doses were averaged, and effective doses were calculated with weighting factors.

RESULTS. The mean effective doses for the pulmonary embolism protocol ranged from 9.9 to 18.5 mSv and for the chest, abdomen, and pelvis protocol from 6.7 to 18.5 mSv. For the pulmonary embolism protocol, the mean effective dose from the Siemens Healthcare 64-MDCT scanner was significantly lower than that from the 16- and 64-MDCT GE Healthcare scanners (p < 0.001). The mean effective dose from the GE 4-MDCT scanner was significantly lower than that for the GE 16-MDCT scanner (p < 0.001) but not the GE 64-MDCT scanner (p = 0.02). For the chest, abdomen, and pelvis protocol, all mean effective doses from the GE scanners were significantly different from one another (p < 0.001), the lowest mean effective dose being found with use of a single-detector CT scanner and the highest with a 4-MDCT scanner. For the chest, abdomen, and pelvis protocols, the difference between the mean effective doses from the GE Healthcare and Siemens Healthcare 64-MDCT scanners was not statistically significant (p = 0.89).

CONCLUSION. According to phantom data, patients are subject to different radiation exposures for similar body CT protocols depending on scanner assignment. In general, doses are lowest with use of 64-MDCT scanners.

Keywords: abdominal imaging • CT dose • MDCT • pulmonary angiography