AJR InPractice
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
 QUICK SEARCH:   [advanced]


     


This Article
Right arrow Figures Only
Right arrow Full Text
Right arrow Full Text (PDF)
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Right arrow Citation Map
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Google Scholar
Right arrow Articles by Teubner, T.
Right arrow Articles by Teubner, J.
PubMed
Right arrow PubMed Citation
Right arrow Articles by Teubner, T.
Right arrow Articles by Teubner, J.
Social Bookmarking
 Add to CiteULike   Add to Complore   Add to Connotea   Add to Del.icio.us   Add to Digg   Add to Reddit   Add to Technorati  
What's this?
Hotlight (NEW!)
Right arrow
What's Hotlight?
DOI:10.2214/AJR.09.2876
AJR 2009; 193:W563-W566
© American Roentgen Ray Society


Technical Innovation

Halfpipe Coaxial Cannula for Self-Contained Vacuum-Assisted Biopsy Systems: Feasibility in a Pig Breast Model

Tobias Teubner1,2, Matthias F. Häfner1, Max Schimmele1 and Joachim Teubner1,2

1 Diagnostic Breast Centre, St. Elisabeth Hospital, Furtwänglestr. 2/2, D-69121 Heidelberg, Germany.
2 Department of Radiology and Nuclear Medicine, University Hospital Mannheim, University of Heidelberg, Mannheim, Germany.

OBJECTIVE. The purpose of this article is to discuss the development of a dedicated halfpipe coaxial cannula for stereotactic vacuum-assisted biopsy. We evaluated the system by retrieving 18 copper targets from a pig breast model in the upright position via vertical and lateral approaches.

CONCLUSION. Sampling was successful in 15 of 18 cases. Errors occurred only in superficial lesions biopsied via the vertical approach. The halfpipe coaxial cannula shows promise for improving positioning accuracy, avoiding target dislocation, and obviating repeated needle repositioning.

Keywords: breast • coaxial systems • stereotactic biopsy • vacuum-assisted biopsy