AJR 2000; 174:1345-1347
© American Roentgen Ray Society
Virtual Endoscopy of Coronary Arteries Using Contrast-Enhanced ECG-Triggered Electron Beam CT Data Sets
Tadashi Nakanishi1,
Minako Kohata1,
Kenji Miyasaka1,
Haruhito Fukuoka1,
Katsuhide Ito1 and
Michinori Imazu2
1
Department of Radiology, Hiroshima University, School of Medicine, 1-2-3
Kasumi, Minami-ku, Hiroshima 734-8551, Japan
2
Second Department of Internal Medicine, Hiroshima University, School of
Medicine, 1-2-3 Kasumi, Minami-ku, Hiroshima 734-8551, Japan.
Received July 13, 1999;
accepted after revision November 1, 1999.
Address correspondence to T. Nakanishi.
Introduction
Application of virtual endoscopy to vascular structures has been limited to
the aorta and its branch orifices
[1,2,3].
The coronary artery is a promising target for virtual endoscopic imaging
because invasive coronary angioscopy has contributed to the assessment of
atheromatous plaques and has potential application for the assessment of
ischemic heart disease [4].
Electron beam CT is capable of providing high-resolution spatial and temporal
images and ECG-triggered scaning for coronary artery imaging
[5,
6]. The present study
investigated coronary endoscopic images created from electron beam CT data
sets of both a phantom model and patients with ischemic heart disease.
Materials and Methods
A phantom and patients were evaluated with an electron beam CT scanner
(Imatron C-150; Imatron, South San Francisco, CA). A phantom for
cineangiography was scanned in an axial direction similar to the scanning of
the human heart. The phantom was constructed from a barium cast of the
coronary artery of an adult pig embedded in polyester resin. The difference in
the CT number between the cast and the resin was approximately 1400 H. In
addition, the diameter of the proximal coronary cast was 6 mm. To investigate
the effectiveness of CT data sets as a coronary artery evaluation tool, source
phantom images from four different scanning protocols were obtained as
follows: 1.5-mm thickness contiguous sections, 3.0-mm thickness sections with
1-mm overlap, 3.0-mm thickness contiguous sections, and 6.0-mm thickness
contiguous sections. These four different scanning protocols were performed
with the following parameters: field of view, 26 cm; scan time, 100 msec per
slice; number of slices, 40 covering the entire phantom.
The subjects consisted of 22 patients (17 men and five women; age range,
46-81 years; mean age, 64 years) who underwent catheter coronary angiography
showing no complete obstruction in any coronary branch. Image acquisition was
done on inspiratory breath-holds, triggered to the ECG, with one acquisition
after every QRS complex at 80% of the R-R wave interval. We used the following
parameters: acquisition time, 100 msec per section; section thickness, 1.5 mm;
field of view, 26 cm; acquisition matrix, 512 x 512. Thus, the spatial
resolution of the axial plane was approximately 0.5 mm per pixel, and the
spatial resolution in the longitudinal direction was 1.5 mm per pixel.
An 18- or 22-gauge angiographic catheter was placed in the antecubital
vein. An 80- to 100-ml injection of nonionic contrast agent (iopamidol, 370
mgI/ml) (Iopamiron 370; Nippon Shering, Osaka, Japan) was administered with a
power injector at a rate of 2-3 ml/sec, followed by a 23- to 30-sec scanning
delay.
Virtual endoscopic views were created from the patient data sets with
Navigator (General Electric Medical Systems, Milwaukee, WI). Before creating
the virtual endoscopic views, a survey of coronary calcification, showing
apparently higher attenuation than vessel lumen, was performed with
multiplanar reconstruction. The first step was to apply a lower threshold
"black in white" algorithm value setting of 70-120 H for the
patient data sets and 300 H for the phantom data sets. Interactive changing
thresholds were performed when segmentation artifacts were observed in the
proximal coronary artery branch. The second step was to add an upper threshold
individually optimized value that had been adjusted to an attenuated calcified
portion to differentiate coronary calcification from the vessel inner lumen.
The quality of the virtual endoscopic images was evaluated in terms of
visualization of coronary orifices, percentage of patients in whom coronary
arteries could be observed, and a measurement of fly-through distance for each
coronary branch.
Results
The phantom examination revealed that a 1.5-mm contiguous thickness or a
3-mm thickness with a 1-mm overlap scan produced good volumetric data sets for
coronary artery evaluation. Other scanning methods, including 3-mm contiguous
scans, produced inadequate source images with the result being poor coronary
vessel images. A fly-through within the coronary barium cast of the phantom
could be easily performed in every field from the main branch to the small
side branches in data sets involving a 1.5-mm contiguous thickness or a 3-mm
thickness with a 1-mm overlap scan. In addition, the coronary orifice was
clearly visualized in all branches (Fig.
1A,1B),
where virtual endoscopic views clearly showed a round configuration that
corresponded to the images taken from the phantom. When using the patient data
sets, a fly-through was not achieved for at least one branch in three patients
(14%) as a result of dense calcification and small coronary branch diameters.
The mean and standard deviation values for CT attenuation at the left coronary
orifice, the right coronary orifice, and at the mid left ventricular level
were 205 ± 40, 212 ± 52, and 208 ± 59 H, respectively. A
total of 21 (95%) of 22 patients showed attenuation levels of more than 130 H
at all three regions. Visualization of the coronary orifices was possible in
17 (89%) of 19 patients in the right coronary artery, 18 (95%) of 19 in the
left main coronary artery, and 18 (95%) of 19 in the left circumflex artery.
Typical coronary artery orifice images created from the patient data sets are
shown in Figure
2A,2B.
In addition, the fly-through distance was 23.5 ± 11.9 mm for the right
coronary artery, 29.4 ± 12.8 mm for the left main-left anterior
descending arteries, and 22.5 ± 11.4 mm for the left circumflex artery.
A fly-through to the lower end of the scanning field was possible in only two
vessels. Calcified plaque was successfully visualized by adjusting the upper
and lower threshold settings (Fig.
3A,3B).

View larger version (154K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 1B. Virtual endoscopic images created from phantom data sets.
Conversely, viewing from aortic root shows virtually two orifices represented
as left anterior descending artery and left circumflex artery.
|
|

View larger version (125K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3A. 64-year-old man with ischemic heart disease. Multiplanar
reconstruction image created from contrast-enhanced ECG-triggered 1.5-mm
contiguous thickness sections clearly shows focal calcification
(arrow).
|
|
Discussion
Vascular disorders have been successfully evaluated with virtual endoscopy
since additional information on projectional images of the vessel orifices
could be obtained
[1,2,3].
To our knowledge, the use of endoscopic views for coronary arteries has not
been reported.
Recent reports have shown the feasibility of electron beam CT data sets for
coronary artery stenosis with several postprocessing methods including
multiplanar reconstruction, three-dimensional surface display, and maximum
intensity projection [5,
6]. Previous coronary artery
evaluation studies involving electron beam CT reportedly used 1.5-mm
collimation gapless [5] or 3-mm
with 1-mm overlap scanning [6].
To reduce any partial volume averaging effects, a thinner section is desirable
because the diameter of the coronary artery is small and differentiation of
the vessel inner lumen and coronary artery wall is essential. It is especially
important to render calcification as a wall structure.
Direct visualization of vessel orifices by CT angiography is fundamentally
difficult because the projectional nature of these images simulates the
appearance of conventional angiograms. Virtual endoscopic images could show
the coronary artery orifices.
One advantage associated with virtual endoscopy is that it enables an
evaluation of the inner surface of the coronary artery images. Visualization
of coronary artery plaque can be established by viewing from the inside,
especially when coronary artery calcification is so severe that it precludes
viewing from the outside [5]. A
characterization of the atheromatous plaque is essential to properly determine
treatment. Virtual endoscopic imaging might have potential applications for
showing disorders of the coronary artery surface such as ulceration, intimal
flap, and irregularity [4].
Practical disadvantages include a relatively short fly-through distance
covering the proximal coronary artery tree. However, this difference depends
largely on source image quality because the phantom data sets enable a long
fly-through to the periphery. Because an apparent difference exists between
the phantom and patient data sets, the fly-through distance depends on the
image quality of a given data set. Vessel contrast determined by CT number is
considered to be a major contributing factor to the quality of the endoscopic
images. Although endoscopy does not provide a quantitative method for
evaluating vessel stenosis, the images obtained provide unique information
about the vessel lumen. Further research is needed to improve source data sets
and postprocessing methods.
References
-
Davis CP, Ladd ME, Romanowski BJ, et al. Human aorta: preliminary
results with virtual endoscopy based on three-dimensional MR imaging data
sets. Radiology 1996;199
: 37-40[Abstract/Free Full Text]
-
Kimura F, Shen Y, Date S, Azemoto S, Mochizuki T. Thoracic aortic
aneurysm and aortic dissection: new endoscopic mode for three-dimensional CT
display of aorta. Radiology 1996;198
: 573-578[Abstract/Free Full Text]
-
Marro B, Galanaud D, Valery CA, et al. Intracranial aneurysm: inner
view and neck identification with CT angiography virtual endoscopy.
J Comput Assist Tomogr 1997;21
: 587-589[Medline]
-
Mizuno K, Miyamoto A, Satomura K, et al. Angioscopic coronary
macromorphology in patients with acute coronary disorders.
Lancet 1991;337
: 809-812[Medline]
-
Nakanishi T, Ito K, Imazu M, Yamakido M. Evaluation of coronary
artery stenoses using electronbeam CT and multiplanar reformation.
J Comput Assist Tomogr 1997;21
: 121-127[Medline]
-
Moshage WE, Achenbach S, Seese B, Bachmann K, Kirchgeorg M.
Coronary artery stenoses: three-dimensional imaging with
electrocardiographically triggered, contrast agentenhanced,
electron-beam CT, Radiology 1995;196
: 707-714[Abstract/Free Full Text]

CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
D. B. Orbach, B. K. Pramanik, J. Lee, T. S. Maldonado, T. Riles, and R. I. Grossman
Carotid Artery Stent Implantation: Evaluation with Multi-Detector Row CT Angiography and Virtual Angioscopy--Initial Experience
Radiology,
January 1, 2006;
238(1):
309 - 320.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. Nakanishi, Y. Kayashima, R. Inoue, K. Sumii, and Y. Gomyo
Pitfalls in 16-Detector Row CT of the Coronary Arteries
RadioGraphics,
March 1, 2005;
25(2):
425 - 438.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
N Rilinger, H Seifarth, R Sokiranski, S Kramer, F Liewald, J Goerich, A Tomzcak, and C E E Nierhoff
Virtual intra-arterial angioscopy (VIA) of the carotid artery based on helical CT data
Br. J. Radiol.,
November 1, 2003;
76(911):
792 - 797.
[Abstract]
[Full Text]
[PDF]
|
 |
|