|
|
||||||||
Original Report |
1
Department of Radiologic Sciences, Medical College of Pennsylvania, 3300 Henry
Ave., 5th Level, Philadelphia, PA 19129.
2
Department of Radiology, Thomas Jefferson University Hospital, 132 S. 10th
St., Ste. 1096 Main, Philadelphia, PA 19107.
Received May 18, 1998;
accepted after revision January 21, 2000.
Address correspondence to P. Chiowanich.
Abstract
|
|
|---|
CONCLUSION. Apparent stenosis of normal vessels on gadolinium-enhanced 3D MR angiography seen on the first-pass acquisition was observed in only a small proportion (approximately 2%) of our patients. The pseudostenosis was reproducible in the phantom model using rapid injection. A stenosis on first-pass images should be interpreted with caution. Confirmation of the findings on other sequences, such as the second-pass gadolinium-enhanced 3D MR angiography or 3D phase-contrast MR angiography, prevented overdiagnosis of significant stenoses.
|
|
|---|
However, we have encountered cases of pseudostenosis at or near vessel origins on gadolinium-enhanced 3D MR angiography, which has not been previously described in the radiology literature. The pseudostenosis was also reproduced in a pulsatile flow phantom experiment. Examples of this observation are presented and possible causes discussed.
|
|
|---|
All examinations were performed on Signa 1.5-T MR scanners (General Electric Medical Systems, Milwaukee, WI), software version 5.5 or 5.6, at two gradient strengths. One scanner had a high-performance gradient system (25 mT/m) and two had a standard gradient system (10 mT/m). A torso phased array coil was used in all cases except a few for which the built-in body coil was used. On the high-performance gradient system, the gadolinium-enhanced 3D MR angiography sequence parameters were enhanced fast gradient-echo 3D; TR range/TE range, 7-9/2-3 msec; flip angle, 50°; matrix, 256 x 160-192 (frequency x phase); half-Fourier acquisition (number of excitations, 0.5); receiver bandwidth, ±31.2 kHz; and fat suppression. The standard gradient systems had the following parameters: 3D fast spoiled gradient-echo; TR/TE, 13/2.3; flip angle, 60°; matrix, 256 x 160-192 (frequency x phase) or 512 x 160-192; number of excitations, one; bandwidth, ±32 kHz; and no fat or spatial saturation pulses. The field of view and the number of partitions were adjusted to accommodate each patient, yielding effective slice thickness of 1.5-3 mm and in-plane resolution of 0.94-2.37 mm. Rectangular field of view (fractional phase-encoding) of 75% was used for axial sequences, 50% for sagittal sequences, and 100% for coronal sequences. Frequency-encoding directions were superior-inferior for sagittal and coronal sequences and right-left for the axial sequence.
All examinations included an unenhanced 3D sequence and at least two successive sequences during a dynamic injection. The acquisition time for each 3D sequence was 18-26 sec (with breath-holding) for the high-performance gradient and 84-107 sec (without breath-holding) for the standard gradient systems. Injection of gadopentetate dimeglumine (Magnevist; Berlex Laboratories, Wayne, NJ) at doses of 0.1-0.2 mmol/kg and a rate of 2 mL/sec for the high-performance gradient, and 0.3 mmol/kg and rate of 0.8-2.0 mL/sec for the standard gradient system, was performed manually through a 20- to 22-gauge peripheral IV catheter. A spoiled gradient-echo sagittal test bolus timing sequence (6.2/1.5; flip angle, 90°; matrix, 256 x 128; field of view, 30 cm; number of excitations, one; slice thickness, 20 mm) was used to set the injection time so that the peak enhancement occurred during the central point of K-space. The 3D sequence on the standard gradient system used sequential ordering of K-space whereby the weakest phase-encoding steps (center of K-space) were acquired in the middle of the sequence. The 3D sequence on the high-performance gradient system also used sequential ordering, but with half-Fourier acquisition (number of excitations, 0.5). The weak phase-encoding gradients were applied near the beginning of the acquisition, rendering it closer to centric ordering. Additional two-dimensional time-of-flight MR angiography (25/1.5; flip angle, 60°; slice thickness, 2-2.5 mm) and 3D phase-contrast MR angiography (TR/TE range, 34/7-9; flip angle, 20°; encoding velocity, 30 cm/sec) sequences were obtained in several patients.
Postprocessing of 3D images was performed at an Advantage Windows station, software version 1.2 (General Electric Medical Systems) by an experienced radiologist using maximum intensity projection and multiprojection volume reconstruction. These images and the source images were reviewed on a digital workstation (Canon Medical, Los Angeles, CA). The major aortic branches were reviewed for the presence of stenosis. The degree of stenosis was categorized subjectively as normal, mild, moderate, severe, or occluded.
In vitro studies used a pulsatile flow phantom with a 2-cm main tube diameter and an 8-mm branch diameter with an orifice angulation of 110°. The total volume of circulating fluid in the phantom was 5 L. The phantom and surrounding reservoir contained a non-Newtonian polymeric solution of carboxymethylcellulose with T1 of 1415, T2 of 201 msec, and viscosity of 0.036 poise (poise = 0.1 kg·m-1·sec-1), values similar to those of blood. Pump settings included stroke volume of 60 mL; peak velocity, 100 cm/sec; rate, 70 beats per min; and systolic/diastolic flow ratio, 2:1.
Administration of gadolinium was performed using two different rates of manual injection: a rapid bolus injection and a slow injection. The experiment was performed once for each rate of injection using 20 mL of 10% gadopentetate dimeglumine solution. Gadolinium was administered into the phantom approximately 1 m from the magnet bore of the high-performance gradient system after a sagittal test bolus timing sequence identical to that previously described for patients. The enhanced fast gradient 3D sequences were obtained in a coronal plane, including the main tube and branch of the phantom, in parameters similar to those of the clinical cases, using 7/2.3; flip angle, 50°; matrix, 256 x 256 (frequency x phase); half-Fourier acquisition (number of excitations, 0.5); bandwidth, ±31.2 kHz; and fat suppression. Frequency encoding was superior-inferior. Acquisition time was 15 sec. For the rapid injection experiment, the bolus was injected at the rate of 2 mL/sec so that the gadolinium concentration varied during acquisition. The 3D sequence was first obtained before injection followed by three successive 3D sequences beginning 4 sec after the initiation of injection. Then the fluid was flushed from the system and replaced by unenhanced fluid. This was followed by the slow-injection experiment using a slow rate of injection (<1 mL/sec) to produce an approximately steady concentration of gadolinium throughout image acquisition. Three successive 3D sequences were acquired beginning approximately 10 sec after the onset of slow injection. Images were reviewed and reconstructed in the same manner as in patients.
|
|
|---|
|
|
|
|
|
|
|
|
This artifact was replicated in vitro (Fig. 3A,3B,3C,3D). A pseudostenosis was prominent during the first pass of rapid-injection experiment (Fig. 3A), was decreased during the second pass (Fig. 3B), and was absent during both the third pass (Fig. 3C) and a subsequent slow injection (Fig. 3D).
|
|
|
|
|
|
|---|
The explanation of this first-pass pseudostenosis phenomenon has not been established. We speculate that the apparent reduction in the lumen size seen only on the first contrast-enhanced sequence is a result of the rapidly changing T1 of blood during sampling of K-space soon after injection. The second-pass sequence has less variability of gadolinium concentration during its acquisition. The phantom experiment replicates the pseudostenosis seen during the early gadoliniun-enhanced period and during the rapid bolus injection, which had rapidly changing gadolinium concentration. However, the experiment does not explain why this phenomenon is most conspicuous just beyond the takeoff of the aortic branches, particularly at the branches that are approximately at a right angle to the main lumen. Variable enhancement during acquisition of the periphery of K-space, which determines edge resolution, might be necessary for this artifact to occur.
Alternative explanations include incomplete mixing at the branch origin as a result of complex flow. The degree of spin dephasing from turbulent flow is not likely to produce a loss of signal as a result of the nature of the short TE of the 3D gradient-echo technique. Additionally, this artifact was not seen on the second-pass images, which should be equally susceptible to spin dephasing. Furthermore, the pseudostenosis was absent on 3D phase-contrast MR angiography, which was far more susceptible to spin dephasing [7].
The edge-ringing or vessel-broadening artifacts described in the in vitro study by Maki et al. [8] are attributed to the filling of the center of the K-space during the rapid rising of contrast concentration. These artifacts may lead to over-estimation of vessel caliber, in contrast to the first-pass pseudostenosis artifact described in our study. The effect of phase- and frequency-encoding directions relative to the vessel orientation does not account for the fact that we observed this pseudostenosis in all situations.
A few limitations of this study should be noted. Because of the retrospective nature of our study, we were unable to determine the exact rate and the timing of injection for each case, which may have contributed to the variability of rate of change and onset of T1 shortening effect relative to the central point of K-space. The power injectors were not available during the study period. Despite the routine use of the sagittal test bolus timing sequence, individual differences in injection methods may have occurred.
Although only one patient underwent catheter angiography as a gold standard, we believe that the routinely performed second-pass sequences are valuable. Both first-pass and second-pass 3D sequences should be obtained. Although during the first pass the vessel-to-background contrast is superior (avoidance of venous signal and, consequently, superior reconstructed MR angiogram), vessel origins are usually visible on the second pass. Therefore, we suggest that second-pass images should be reviewed when the diagnosis of a significant stenosis is suspected because potential inconsistency in the timing of bolus injection does occur in routine clinical practice. It is possible that using other methods of K-space ordering or an improved consistency and precision in the bolus injection technique may reduce the frequency of this first-pass pseudostenosis artifact [9].
Our observation of a pseudostenosis on gadolinium-enhanced 3D MR angiography should alert the radiologist to a potential error in diagnosing significant stenoses. Analysis of other secondary findings, such as renal morphology and enhancement pattern, and of other MR angiographic techniques, particularly 3D phase-contrast MR angiography, is also helpful [7]. To increase the specificity and accuracy of this technique, interpretation should include a careful review of other images, particularly the second-pass images from the same injection.
|
|
|---|
This article has been cited by other articles:
![]() |
M. A. Schmidt and R. Morgan Renal Contrast-enhanced MR Angiography: Timing Errors and Accurate Depiction of Renal Artery Origins Radiology, October 1, 2008; 249(1): 178 - 186. [Abstract] [Full Text] [PDF] |
||||
![]() |
V. L. Wright, W. Olan, B. Dick, H. Yu, N. Alberts-Grill, L. L. Latour, and A. E. Baird Assessment of CE-MRA for the rapid detection of supra-aortic vascular disease Neurology, July 12, 2005; 65(1): 27 - 32. [Abstract] [Full Text] [PDF] |
||||
![]() |
K.-M. Han, L. E. M. Duijm, G. R. P. Thelissen, P. W. M. Cuypers, P. Douwes-Draaijer, A. V. Tielbeek, J. H. M. Wondergem, and H. C. M. van den Bosch Failing Hemodialysis Access Grafts: Evaluation of Complete Vascular Tree with 3D Contrast-enhanced MR Angiography with High Spatial Resolution: Initial Results in 10 Patients Radiology, May 1, 2003; 227(2): 601 - 605. [Abstract] [Full Text] [PDF] |
||||
![]() |
R. Wyttenbach, S. Gianella, M. Alerci, A. Braghetti, L. Cozzi, and A. Gallino Prospective Blinded Evaluation of Gd-DOTA- versus Gd-BOPTA-enhanced Peripheral MR Angiography, as Compared with Digital Subtraction Angiography Radiology, April 1, 2003; 227(1): 261 - 269. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |