Abstract
|
|
|---|
SUBJECTS AND METHODS. Real-time interactive duplex MR measurements (TR/TE, 53/27; flip angle, 90°; encoding velocity, 100 or 150 cm/sec) were performed in vitro with a steady-flow phantom and in 10 healthy volunteers in whom common and internal carotid artery velocities were measured. In eight volunteers, velocity measurements were also performed in the superior sagittal sinus during both normal breathing and hyperventilation. Timevelocity plots were analyzed qualitatively and quantitatively and compared with findings from conventional segmented k-space phase-contrast MR imaging and Doppler sonography.
RESULTS. Velocity determinations for real-time duplex MR and conventional phase-contrast MR imaging showed an in vitro correlation of 0.99 and an in vivo correlation of 0.83 (carotid arteries) and 0.76 (venous sinus). Velocity measurements in the carotid arteries with real-time MR imaging were significantly lower than those obtained with conventional phase-contrast MR (averaged, 7.8%; p = 0.003) or sonography (23.7%, p < 0.001), likely because of volume averaging. Small but significant velocity changes occurring in the venous sinus during hyperventilation were reliably identified with both MR techniques.
CONCLUSION. Real-time interactive duplex MR imaging can be effectively applied in neurovascular imaging to obtain hemodynamic information.
|
|
|---|
Our study evaluated a new ungated real-time MR phase-contrast technique that quantifies and displays flow velocities in real times. The sequence uses a two-dimensional selective radiofrequency pulse followed by flow-sensitizing gradients with an echoplanar readout. It provides the simultaneous display in real time of both an anatomic image for positioning and the through-plane flow-velocity data. Flow velocities are displayed as timevelocity curves, closely resembling Doppler flow-velocity distribution plots. By controlling scan position and orientation interactively, one can optimize flow signal. This technique has been termed duplex MR imaging because of its similarities to sonography [12].
We have evaluated duplex MR imaging in vitro with a constant-velocity pump and then applied it in healthy volunteers to measure flow velocities and to evaluate velocity waveforms. Measurements in the common carotid artery and in the internal artery were correlated with conventional phase-contrast MR flow measurements and Doppler sonographic measurements. Flow velocities in the superior sagittal sinus were measured by real-time duplex MR imaging and phase-contrast MR imaging during both normal breathing and hyperventilation to decrease blood flow velocity.
|
|
|---|
Study Subjects
Ten healthy volunteers (eight men, two women; 25-80 years old; mean age,
31.4 years) were examined with MR imaging and Doppler sonography to measure
flow velocities in the carotid arteries within an hour. In eight subjects, MR
flow velocities of the superior sagittal sinus were determined during normal
breathing and after 3 min of hyperventilation of room air. Written informed
consent was obtained from all volunteers, and the study was sanctioned by the
institutional review board.
MR Imaging
All examinations were performed on a 1.5-T clinical imaging system
(Quantum; Siemens, Erlangen, Germany), equipped with high-performance
gradients (maximum gradient strength, 30 mT/m; rise time, 240 µsec).
MR-compatible ECG leads were placed on the subject's back before the
examination.
The imaging protocol followed by sequence details is described. For measurements of the carotid arteries, a four-element phased-array neck coil was used. Guided by localizer images, we scanned the carotid bifurcation with an axial two-dimensional time-of-flight MR angiography sequence. Conventional phase-contrast MR measurements were performed 2-3 cm below and above the carotid bifurcation in the common and internal carotid artery as seen on the time-of-flight MR source images. Acquisition was in the axial plane, which was perpendicular to the direction of blood flow. Only one coil element was activated for real-time interactive duplex MR measurements because processing and combining images from two or more elements significantly reduced temporal resolution. Duplex MR measurements were made as close as possible to the same position as the conventional phase-contrast measurements. For measurements of the superior sagittal sinus, the quadrature head coil was used. The sinus measurements were made just above the confluens sinuum, identified from conventional axial scout images, and orientated perpendicular to the blood-flow direction. Conventional phase-contrast MR and real-time interactive duplex MR measurements were first performed during normal breathing. Then volunteers were asked to hyperventilate room air for 3 min, after which conventional phase-contrast MR measurements were performed. After a break of a few minutes, hyperventilation was repeated for 3 min, and real-time interactive duplex MR measurements were performed.
Sequence parameters were as follows: axial two-dimensional time-of-flight MR angiography (TR/TE, 30/9.8; flip angle, 45°; field of view, 150 x 200; matrix, 146 x 256; matrix slice thickness, 3 mm; slices centered around the carotid bifurcation, 40) was performed with venous presaturation. With the ECG-triggered cine phase-contrast sequence (31.8/3.4; flip angle, 30°; field of view, 150 x 300 mm; matrix, 96 x 256: slice thickness, 8 mm; encoding velocity, 100 or 150 cm/sec), we measured 19-25 cardiac phases on the basis of each subject's heart rate, beginning immediately after the R wave. The real-time duplex MR imaging sequence (53/27; flip angle, 90°; slice thickness, 8 mm; field of view, 98 x 314 mm; matrix, 30 x128; encoding velocity, 100 or 150 cm/sec perpendicular to the image plane; temporal resolution, 53 msec) consisted of a two-dimensional selective radio-frequency pulse followed by flow-sensitizing gradients and an echoplanar readout (Fig. 1). The two-dimensional excitation selected a rectangular column, 8 mm in the slice-select direction by 98 mm in the phase-encoding direction. This column can be phase-encoded with good spatial resolution and without aliasing, with relatively few phase-encoding lines. Data were processed and displayed on the console in real time in a two-part display, consisting of a magnitude image and the timevelocity plot (Fig. 2A,2B). The plot showed a total of the most recent 128 acquisitions and scrolled right with each scan. The scanning plane was positioned interactively with a six-degree-of-freedom input device (SpaceMouse; Logitech, Fremont, CA). Flow information was calculated from a 2 x 2 pixel region of interest at the image center, and velocity measurements were calculated and displayed for each of the four pixels. After the real-time duplex MR sequence was started, the target vessel was moved into the center of the region of interest with the input device. When the vessel was centered and the characteristic waveform was observed consistently over a few seconds, care was taken not to alter the scan position. On average, 400-800 acquisitions were obtained for measurement of a single vessel, resulting in acquisition times of 20-40 sec for a single vessel.
|
|
|
Sonography
Immediately after the MR examination, subjects underwent bilateral carotid
sonography (Aspen; Acuson, Mountain View, CA) in the supine position, with a
broadband (5-10 MHz) linear transducer. Gray-scale and Doppler color-flow
examinations of the common carotid arteries and the internal carotid arteries
were performed in all subjects. Doppler spectral waveforms were obtained in
the distal common carotid artery and in the internal carotid artery beyond the
bulb widening, approximately 2-3 cm below and above the carotid bifurcation.
We performed the measurements using angle correction and the measured angle of
insonation was kept less than 60° for all measurements. Absolute peak
systolic velocities were recorded with the use of electronic calipers. The
radiologist performing sonography was unaware of findings on MR imaging.
Image Analysis
The radiologist who performed the duplex MR study determined peak systolic
velocities from real-time duplex MR imaging, using an electronic caliper at
the MR console (Fig.
2A,2B).
The radiologist was not aware of the results on sonography or on
phase-contrast MR imaging. We analyzed conventional phase-contrast data on a
Sparc 5 workstation (Sun Microsystems, Palo Alto, CA), using software
developed in-house in the IDL programming language (Research Systems, Boulder,
CO). The program identified the pixel with the maximum peak velocity in each
vessel and plotted velocity values over the cardiac cycle.
Statistical Analysis
Velocity determinations in the phantom and in the superior sagittal sinus
as measured by conventional phase-contrast MR imaging and real-time duplex MR
imaging and velocity measurements in the phantom by two independent
radiologists using real-time duplex MR imaging were compared with a paired
two-tailed Student's t test. We applied a paired two-tailed Student's
t test, using the Bonferroni adjustment, to evaluate differences in
the peak systolic velocities between real-time duplex MR imaging, conventional
phase-contrast MR measurements, and sonography for the carotid arteries.
The Bland-Altman method was used to describe quantitatively the agreement between the three techniques for velocity measurements. Linear regression analysis was also performed to compare velocities measured by real-time duplex MR imaging, phase-contrast MR imaging, and sonography and to compare in vitro real-time duplex MR velocity measurements made by two independent radiologists.
For the two-tailed paired Student's t test, p values of less than 0.05 were considered significant, and for the Bonferroni t method, p values of less than 0.01 were considered significant.
|
|
|---|
![]() |
Flow Evaluation in the Carotid Arteries
In all subjects, time-of-flight MR angiography and sonography showed
normal-appearing carotid arteries. The time required for real-time duplex MR
imaging was 30-40 min at the beginning of the study and dropped with
experience to about 10 min.
The locations of velocity measurements in the common carotid artery with conventional phase-contrast MR and real-time duplex MR averaged 2.31 ± 0.8 cm and 2.22 ± 0.76 cm below the carotid bifurcation, respectively, and for the internal carotid artery, 2.59 ± 1.26 and 2.05 ± 0.8 cm above the bifurcation. The averaged difference between measurement positions was 0.1 ± 1.3 cm, indicating reasonably good agreement (p = 0.79).
Velocity waveforms depicted by real-time duplex MR imaging and Doppler sonography were qualitatively similar. In the common carotid artery, a sharp systolic upstroke and little end-diastolic flow was observed. In the internal carotid artery, the waveform resembled low resistance arterial flow with substantial persistent end-diastolic flow (Fig. 2A,2B). Doppler sonographic measurements of peak systolic velocities for the common and internal carotid artery were well correlated with both real-time duplex MR imaging (r = 0.83) and phase-contrast MR imaging (r = 0.75) (Fig. 3). However, compared with sonographic measurements, significantly lower velocity determinations were obtained with both real-time duplex MR imaging (mean difference, 19.8 ± 15.1 cm/sec, representing 23.7%; p < 0.001) and with phase-contrast MR measurements (mean difference, 14.5 ± 17.5 cm/sec, representing 18.3%; p < 0.001). The two MR measurements were well correlated, but flow velocities determined with duplex MR imaging were significantly lower (r = 0.85; mean difference 5.4 ± 10.8 cm/sec representing 7.8%; p = 0.003).
|
Flow Evaluation in the Superior Sagittal Sinus
Real-time duplex MR imaging showed a slow pulsatile flow pattern in the
distal superior sagittal sinus in the craniocaudal direction
(Fig. 4). Flow velocities
measured by phase-contrast MR imaging and real-time duplex MR imaging showed a
correlation of 0.76. Conventional phase-contrast MR measurements were on
average 2.84 ± 5.1 cm/sec higher than real-time duplex MR velocity
determinations (p = 0.04). Peak velocity for real-time duplex MR
imaging determined during normal breathing in eight volunteers averaged 30.52
± 6.81 cm/sec and dropped to 19.7 ± 2.8 cm/sec after 3 min of
hyperventilation. The corresponding values for phase-contrast MR imaging were
32.0 ± 8.0 cm/sec and 24.3 ± 5.3 cm/sec. For both methods, the
drop in velocity was significant (p < 0.01).
|
|
|
|---|
Comparison with Conventional Phase-Contrast MR Imaging
Flow velocities determined by real-time duplex MR imaging showed an
excellent correlation with phase-contrast measurements in a steady-state flow
phantom and good correlation in in vivo measurements. The measurement of small
velocity differences in the superior sagittal sinus before and after
hyperventilation shows the high sensitivity of the technique in vivo. However,
on average, real-time duplex MR velocity measurements were lower than those
obtained with conventional phase-contrast MR imaging. This difference might,
in part, be explained by the larger voxels used with real-time duplex MR
imaging, leading to a greater averaging of flow velocities. This view is
supported by the observation that the difference in flow velocities was more
evident in higher velocities in which the effect of averaging between flow in
the center of a vessel and at its margin is greater than that at slow
velocities.
Our phase-contrast measurements of peak flow velocities in the common and internal carotid artery were in good accordance with findings reported by Vanninen et al. [2]. Our velocity determinations in the superior sagittal sinus were in good accordance with measurements of Mattle et al. [13], who used a bolustagging time-of-flight technique. However, Mehta et al. [10], who used a conventional phase-contrast technique, found velocities in the mid superior sagittal sinus for volunteers during normal breathing of about 15 cm/sec. These lower velocity determinations in the superior sagittal sinus might be explained because their velocity determinations were averaged over the entire vascular lumen. In our study, the phase-contrast data were postprocessed to identify the pixel with the highest velocity in the vessel area. Furthermore, our velocity determinations were performed in a more distal part of the sinus.
The real-time duplex MR sequence used in this study gives a temporal resolution of almost 20 velocity determinations per second, comparable to the temporal resolution of the conventional phase-contrast MR technique used. However, with real-time duplex MR imaging, pulsatile flow patterns can be measured without cardiac gating. The application of the technique is, therefore, simplified, and hemodynamic information can be obtained in cases in which satisfactory cardiac gating is impossible, such as in patients with arrhythmias. The real-time display of velocity has two additional advantages: first, flow velocities can be measured instantly without postprocessing, unlike those required for conventional phase-contrast MR imaging. Second, adjustments in sampling or measurement location can be made in real-time as with Doppler sonography.
On the other hand, at the beginning of this study, the overall time needed to perform flow measurement was longer for real-time MR imaging than for phase-contrast MR measurements. The primary reason was the lack of experience in controlling the MR scanner interactively with a six-degree-of-freedom device. In our study population, the straight axial plane was almost perpendicular to the course of all vessels investigated. However, improved control of the scan orientation would be desirable for reliable flow measurements in tortuous vessels and in all vessels with an anatomic oblique flow direction. To have instantaneous feedback of the current scan position and orientation, it would probably be advantageous to see the current position on a three-dimensional MR angiography data set. We believe these changes would reduce examination time to a few minutes. The current real-time duplex MR imaging technique does not evaluate volumetric flow rates, although processing modifications may allow such measurements in the future. Finally, real-time duplex MR measurements can generate a large number of images, affecting storage capacity. To reduce this problem, images with cumulative data for over 100 acquisitions (as shown in the figures) could be selectively saved.
Comparison with Sonography
Real-time duplex MR imaging has several advantages over Doppler sonography
in neurovascular applications. First, the absence of the sound propagation
problems allow its use in cases in which the presence of bone or air cavities
precludes the application of sonography. Although it is possible, for example,
to insonate the rostral part of the superior sagittal sinus through a
transtemporal window, the anterior and mid portions of the superior sagittal
sinus cannot be assessed [14,
15]. Second, the free choice
of scan-plane orientation eliminates angulation errors that are common in
sonographic measurements. Finally, flow measurements can be made in
conjunction with high-resolution anatomic MR imaging and MR angiography and do
not require a second procedure.
Real-time duplex MR imaging represents a trade-off between the flow-sampling rate and the spatial resolution. This study was performed with a lower spatial resolution than that of the phase-contrast MR scans. This difference might be the major reason that velocity measurements were lower than those with the conventional phase-contrast MR imaging technique; this observation holds true, even to a higher degree, for the comparison with duplex sonography. The lower velocity measurements with phase-contrast MR imaging compared with sonographic measurements were also described by Vanninen et al. [2]. For accurate velocity measurements in stenotic arteries or smaller vessels with real-time duplex MR imaging, improved spatial resolution would be needed without loss of temporal resolution. This improvement might be accomplished by application of faster k-space encoding schemes (e.g., elliptic k-space acquisition schemes).
In conclusion, real-time duplex MR imaging can be considered a promising tool for assessment of flow velocities in the carotid arteries and venous sinuses. From a technical standpoint, further improvements in implementation of the technique are necessary. Studies should be performed to determine the utility of the technique in neurovascular diseases such as carotid stenosis and dural venous sinus thrombosis.
|
|
|---|