AJR 2002; 178:153-157
© American Roentgen Ray Society
Using a Sterile Disposable Protective Surgical Drape for Reduction of Radiation Exposure to Interventionalists
Jerry N. King1,2,
Anna M. Champlin3,
Charles A. Kelsey3 and
David A. Tripp4
1
Department of Radiology, Presbyterian Hospital, 1100 Central Ave. S.E.,
Albuquerque, NM 87106.
2
Radiology Associates of Albuquerque, 5890 Eubank N.E., Albuquerque, NM
87111.
3
Department of Radiology, University of New Mexico, 2211 Lomas Blvd. N.E.
Albuquerque, NM 87131.
4
Department of Radiology, University of Utah Health Sciences Center, 50 N.
Medical Dr., Salt Lake City, UT 84132.
Received March 27, 2001;
accepted after revision July 24, 2001.
Address correspondence to J. N. King.
Abstract
OBJECTIVE. The purpose of this paper is to show the effectiveness of
a new radiation protection method designed to decrease the amount of scatter
radiation received by practitioners performing procedures under fluoroscopic
guidance.
MATERIALS AND METHODS. A sterile, disposable, lead-free surgical
drape containing radiation protection material composed primarily of bismuth
was evaluated for effectiveness in reducing radiation doses to health care
personnel. Measurements of phantom scatter, patient scatter, skin entrance,
and the effects of collimation, together with comparative monthly
thermoluminescent dosimeter recordings, were taken to determine the
effectiveness of X-ray beam attenuation using the bismuth drapes.
RESULTS. Scatter radiation to physicians, as measured by
thermoluminescent dosimeters placed on each eye, the thyroid, and the wrist,
was reduced by 12-fold for the eyes, 25-fold for the thyroid, and 29-fold for
the hands when the radiation-attenuating surgical drape was used when compared
with control studies performed with a standard nonattenuating surgical drape
alone. Monthly thermoluminescent dosimeter measurements decreased fourfold in
one physician. Using the protective drape reduced exposure to the assistant in
each case to negligible levels. Skin entrance dose was not increased unless
the protective drape was placed directly in the X-ray beam. An X-ray
attenuation factor equivalent to 0.1 mm of lead with 8 x 8 cm
collimation reduced the scatter rates from five- to ninefold despite a 30-40%
increase in entrance exposure rate as the lead equivalence increased.
CONCLUSION. Depending on the procedure, the height of the
practitioner, and the positioning of the radiation-attenuating surgical drape,
use of this drape can substantially reduce the radiation dose to personnel
with minimal or no additional radiation exposure to the patient.
Introduction
The harmful effects of ionizing radiation were recognized shortly after the
discovery of the X ray by Wilhelm Conrad Roentgen in 1895
[1,2,3,4,5].
These harmful effects were particularly evident in the hands of individuals
exposed repeatedly to the X-ray beam for prolonged periods of time. Erythema,
dermatitis, and skin cancer were found to result from this exposure, and it
was initially thought that avoiding the primary beam was sufficient protection
[2,
6,7,8,9].
However, in the 1920s concerns regarding the adverse effects of radiation were
again raised with the identification of an increased rate of leukemia in
radiologists [1]. This
recognition led to the creation of organizations such as the International
Commission on Radiological Protection (1928) and the National Committee on
Radiation Protection and Measurements (1929, later the National
"Council"), which became important in making recommendations on
radiation protection. The first recommendations for tolerance doses for
radiation workers came from the National Council on Radiation Protection and
the International Commission on Radiological Protection in 1934. The
recommendations of both those organizations for tolerance doses for radiation
workers have decreased by a factor of 5-10 since 1934. This decrease is the
result of increased knowledge of the risks from radiation exposure, an
increased desire among workers to avoid the harmful side effects of radiation,
and improvements in technology
[1,
3,4,5,
8,9,10,11,12,13].
Although the recommended limit for radiation workers has not changed
greatly since about 1958, the philosophy toward radiation protection and
limits has changed dramatically. The limit is now regarded as an upper limit
of acceptability. The principle of ALARA (as low as reasonably achievable) is
used to ensure that most exposures will be well below the accepted limit.
Experience with the ALARA principle and the limit of 5 rem (50 mSv) per year
has allowed the average exposure of workers in the United Stateswith
the exception of interventional radiologists and cardiologiststo
decline steadily to about 5% of the limit. The increased use of fluoroscopy by
anesthesiologists for pain therapy and during radiation therapy procedures has
further expanded the risk to health care providers.
Although the acute effects of radiation are not commonly a problem, the
stochastic effects of radiation remain a concern. The probability of the
occurrence of stochastic effects is directly related to the radiation dose,
but the severity of these conditions is not related to the total dose
received. Stochastic effects include carcinogenesis and genetic mutation; they
are of particular concern because there is no threshold dose below which the
radiation-induced effects will not occur. The nonstochastic effects, such as
radiation-induced cataracts, do have a threshold dose, and above this
threshold the severity is directly related to the dose. Stochastic events are
considered to occur at all doses, but the less the frequency, the lower the
dosethus, the principle of ALARA
[1,
4,
5,
8,9,10].
This article focuses on the use of a new radiation protection device
intended to reduce both the unit dose and the overall level of radiation
experienced by radiation workers during interventional radiology
procedures.
Materials and Methods
This study was conducted in three phases. In the first phase, shields of
varying lead equivalency were tested for effectiveness in attenuating scatter
radiation from a standard X-ray phantom. The second phase involved testing of
a commercially available protective drape (RADPAD; Worldwide Innovations &
Technologies, Overland Park, KS) during a series of patient studies with
institutional research board approval and informed consent
(Fig. 1). The third phase was
the routine use of this protective drape, in addition to improved collimation
techniques, during interventional procedures by a radiologist who had
regularly exceeded recommended radiation exposure levels.

View larger version (98K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 1. Sterile protective surgical drape (32 x 44.5 cm)
viewed from back or side placed toward patient (see
Fig. 2 for front view). Single
arrow indicates 4.5 x 9.5 cm opening for needle and catheter placement.
Double arrows show location of adhesive tape to aid in maintaining drape
position. Arrowheads outline 11-cm channel designed to be cut if necessary for
ease of manipulation around needles and catheters. This channel has been cut
in Figure 2.
|
|

View larger version (131K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2. Interventional procedure with setup as used in this study.
Sterile, disposable, lead-free radiation-protective surgical drape
(arrows) is placed between operator and primary beam. Operator stands
in "shadow" produced by protective drape.
|
|
Phase 1
A series of measurements of shielding samples were evaluated for scatter
during routine fluoroscopy with a typical angiographic C-arm equipped
with a 38.10-cm (15-inch) image intensifier with and without tight collimation
(Philips Medical Systems, Best, The Netherlands). The secondary (scatter)
radiation exposure rates were measured using an anthropomorphic chest phantom
(Nuclear Associates, Carle Place, NY). The scatter rates at 30 cm from the
sternum, together with the automatic brightness system fluoroscopic entrance
exposure rates, were recorded with and without 8 x 8 cm collimation. In
all recordings, the image intensifier was positioned as closely as possible to
the chest with a source-to-image distance of 100 cm. The scatter rates were
measured with an MDH Model 1015 X-ray monitor using a 180-mL ion chamber
(Radcal, Monrovia, CA).
Phase 2
Twelve patient studies were performed using extensive placement of
thermoluminescent dosimeters (TLDs). Eight patient studies served as control
studies that were performed with the use of a standard nonattenuating surgical
drape; and the studies of four patients used the radiation-attenuating drape.
The drape was placed on the patient with the window of the drape over the skin
puncture site (Fig. 2). All
patient studies were performed on the same radiographic equipment (Picker,
Cleveland, OH), and all procedures were percutaneous nephrostomy placements
performed from the right side of the patient. The monitor was positioned for
viewing in the same location for all examinations. Two radiologists
participated in each study, with interpretations obtained from nine
physicians. Several physicians participated in more than one patient
evaluation. All physicians used their own type of standard radiation
protection. The 12 patient studies included TLDs placed on the patient and the
operators. A TLD was placed on the skin at the level of the entrance dose of
each of the 12 patients, and on the thyroid, left eye, right eye, and left
wrist of both the primary and the assisting physicians performing the
procedure. The TLDs were positioned outside any radiation-protective covering
such as lead aprons, thyroid shields, or leaded glasses. The TLDs were taped
to the skin immediately lateral to the eye for the eye recordings. Eight
control studies were conducted without the protective drape, and four studies
were completed under similar conditions using the protective drape, for a
total of 108 TLD recordings on the nine participating physicians and the 12
patients.
Phase 3
A single radiologist with consistently excessive TLD measurements was asked
to use the protective drapes routinely. In addition, this physician attempted
to carefully perform collimation during all procedures. TLD measurements were
recorded for 3 months: the first 2 months without the use of the protective
drape, and the last month with the protective drape used in most (but not all)
procedures.
Results
Phase 1
The scatter rates at 30 cm from the sternum, together with the automatic
brightness system fluoroscopic entrance exposure rates, are tabulated in
Table 1 for various bismuth
drape lead equivalencies with no collimation. These same results when 8
x 8 cm collimation is used are also given in
Table 1. Note the scatter rates
are reduced by five- to nine-fold despite a 30-40% increase in entrance
exposure as the lead equivalency increases. Also, the scatter (as expected)
can be substantially reduced through the use of increased collimation (see
Table 1).
View this table:
[in this window]
[in a new window]
|
TABLE 1 Fluoroscopy Scatter Rate, Skin Entrance Dose, and Percentage of
Reduction as a Function of Shielding Lead Equivalency and Collimation
|
|
Phase 2
Patient studies were performed using a drape of uniform thickness and
bismuth content. Scatter radiation to the physicians as measured by the TLDs
placed on the thyroid, each eye, and the wrist was reduced by a factor of 12
for the eyes, 25.8 for the thyroid, and 29.4 for the hands. As evidenced by
the skin entrance doses to the patients, both the control procedures and the
drape-protected procedures were of a braod range of complexity
(Table 2). Because TLDs were
positioned on the skin surface lateral to each eye, the recordings for the
right and left eyes are strikingly different, showing the effect of the skull
on scatter X-ray attenuation (Table
3). Note that with the exception of a single left wrist recording
during an extended fluoroscopic procedure, measurements for the assistant in
each procedure performed with the radiation-attenuating surgical drape were
negligible. Presumably, in this case, the assistant was asked to hold or
position a device close to the primary beam
(Table 2, radiologist 2, left
wrist).
View this table:
[in this window]
[in a new window]
|
TABLE 3 Average Thermoluminescent Dosimeter Measurements Expressed as Percentage
of Skin Entrance Dose Without and With Protective Drape
|
|
Phase 3
A single interventional radiologist who had exceptionally high TLD
measurements was given the opportunity to use the protective drapes routinely.
In addition, this physician attempted to improve the use of collimation and to
keep exposure times to a minimum. This individual's TLD measurements decreased
four-fold from the previous 2 months. These results are summarized in
Table 4.
Discussion
Radiation protection for radiologists and radiology personnel is an
accepted standard of practice in all medical facilities
[14,15,16,17,18,19].
As a result, the occupational radiation dose for most medical workers has
declined steadily during the last three or four decades
[1,
4,
5,
9,
10]. However, there are
exceptions to this decline. Areas of concern include interventional radiology,
cardiac catheterizations, and C-arm fluoroscopic procedures
[10,11,12,13,14,15,16].
Personnel in each of these areas receive radiation exposures that approach or
even exceed the recommended dose limits
[10,
20,21,22,23,24].
As many as 50% of these personnel have been found to rarely or never wear
their assigned dosimeters. Reasons for this lack of compliance probably
include skepticism about the risk as estimated by the dosimeter, fear of being
asked to limit fluoroscopy time, and the excessive administrative requirements
of investigating high TLD measurements even if they are within the recommended
guidelines [12,
13]. Interventional
radiologists continue to be exposed to radiation that will produce substantial
side effects such as radiation-induced cataracts and radiation dermatitis
[2,
3]. Routine angiography
procedures, particularly cerebral angiography, result in relatively low levels
of radiation exposure unless advanced interventional procedures are required
[12,
20,
22,
24,
25]. However, abdominal
angiography may result in substantially greater doses, especially to the hand,
for which the dose may be nearly 10 times as great as for cerebral
angiography. The hand dose during cardiac catheterization may be as much as
20-50 times the dose in cerebral procedures with similar procedure times.
Similar differences are also seen for exposure to the eyes and the thyroid
gland. When the procedures are complex, the doses may be as much as 50%
greater than has typically been reported
[12,
20,
24].
As interventional radiology procedures increase in complexity, the length
of these procedures often increases proportionately, and the physicians and
staff are required to be near the patient and the X-ray tube for prolonged
periods of fluoroscopic time. It is increasingly common for the fluoroscopy
beam to be on for as long as 60 min or more for a given case, and dose rates
may be as much as 5 R (1290 µC/kg) per minute. Workers in this environment
are increasingly exposed to greater levels of radiation, and the trend is
toward more rather than less radiation risk
[2,
8,
10,11,12,13,14,15,16,17,18,19,20,21,22,23,24,25,26,27,28].
A number of methods have been devised to decrease the radiation dose to
medical workers, including radiation-protective aprons, thyroid shields,
leaded glasses, and several types of ceiling- or floor-mounted shields
[2,
6,
7,
14,15,16,17,18,19,
25,26,27,28].
In addition, radiation-attenuating surgical gloves have been advocated because
hand doses are generally greater than doses to other parts of the body
[2,
6,
7,
25,
28] (Tables
2 and
3). The combination of
forward-scattered X rays, backscattered X rays, and secondary electrons
released from the glove material reduces the effectiveness of these
radiation-attenuating surgical gloves
[7]. Some publications caution
that radiation-attenuating gloves should not be relied on as a primary means
of radiation protection, and that the first line of defense is to keep the
examiner's hands out of the primary beam
[2,
6]. Our study supports this
point by showing that limiting exposure to the primary beam and decreasing the
amount of scatter will markedly decrease the radiation dose not only to the
hands (as much as 30-fold) but also to the thyroid (25-fold) and the eyes
(12-fold). The radiation exposure to support staff or an assistant may be
negligible, even for prolonged and complicated procedures, when the method
reported here is used. Given the demand for complex procedures that involve
prolonged radiation exposure, reducing the dose to physicians and staff is
increasingly important.
The concept of using a radiation protection shield placed on the patient to
attenuate the scatter radiation that may reach an operator is not new. Results
similar to those presented here have been shown for similar shields produced
from lead
[15,16,17,18,19].
However, because of the need for universal precautions and the toxic nature of
lead, the use of nondisposable lead devices has not proven practical
[29,30,31,32,33,34,35].
The use of materials other than lead for radiation shielding has also been
thoroughly investigated. The device investigated here is a lead-free,
disposable, radiation-protective, and sterile surgical drape that not only
allows a sterile surgical barrier but also provides substantial radiation
protection to the primary operator (Tables
2,3,4).
Because it is lead-free, the device does not raise additional environmental
concerns and can be disposed of in the same way as any surgical drape
[29,30,31,32].
The results of our study are equally important for support personnel and
assistants, who may receive negligible amounts of radiation even during long
and complex fluoroscopic procedures when this radiation-protective surgical
drape is used (Table 2).
At the time of this writing, the price is approximately $34 per drape,
which adds relatively little to the cost of the procedure. Opening the package
and positioning the drape on the patient usually takes less than 1 min.
Because the drape is radiopaque, it may need to be repositioned from time to
time during complex procedures, or it may be removed completely and later
replaced if necessary. Because the drape material is flexible, these
adjustments can usually be done in a few seconds. However, if the largest area
of the drape is initially placed on the side of the patient nearest the
operator, the need for repositioning is usually minimal. We believe that the
large reduction in radiation dose to the operator is well worth the small
amount of time and the relatively little added cost required to use the
drape.
References
-
National Council on Radiation Protection and Measurements.
Guidance on radiation received in space activities,
NCRP Report No. 98. Bethesda, MD: National Council on Radiation Protection and
Measurements, 1992
-
Marx MV, Ellis JH. Radiation protection of the hand in
interventional radiology: should it fit like a glove?
Radiology
1996;200:24
-25[Free Full Text]
-
Wagner LK. CT fluoroscopy: another advancement with additional
challenges in radiation management. Radiology
2000;216:9
-10[Free Full Text]
-
Matanoski GM, Sartwell P, Elliott E, et al. Cancer risks in
radiologists and radiation workers. In: Boice JD Jr, Fraumeni JF Jr, eds.
Radiation carcinogenesis: epidemiology and biological
significance. New York: Raven 1984:83
-96
-
Kumazawa S, Nelson DR, Richardson AC. Occupational
exposure to ionizing radiation in the United States: a comprehensive review
for the year 1980 and a summary of trends for the years
1960-1985. Environmental Protection Agency, EPA 520/1-84-005.
Springfield, VA: National Technical Information Service,
1984
-
Kelsey CA, Mettler FA. Flexible protective gloves: the emperor's
new clothes? Radiology
1990;174:275
-276[Abstract/Free Full Text]
-
Wagner LK, Mullhern OR. Radiation-attenuating-surgical gloves:
effects of scatter and secondary electron production.
Radiology
1996;200:45
-48[Abstract/Free Full Text]
-
Broadbent MV, Hubbard LB. Science and perception of radiation risk.
RadioGraphics
1992;12:381
-392[Abstract]
-
National Council on Radiation Protection and Measurements.
Limitation of exposure to ionizing radiation, NCRP
Report No. 116. Bethesda, MD: National Council on Radiation Protection and
Measurements, 1993
-
Hendee WR. Estimation of radiation risks: BEIR V and its
significance for medicine. JAMA
1992;268:620
-624[Abstract/Free Full Text]
-
Tryhus M, Mettler FA Jr, Kelsey C. The radiologist and angiographic
procedures: absorbed radiation dose. Invest Radiol
1987;22:747
-750[Medline]
-
Niklason LT, Marx V, Chan HP. Interventional radiologists:
occupational radiation doses and risks. Radiology
1993;187:729
-733[Abstract/Free Full Text]
-
Quinn AD, Taylor CG, Sabharwal T, et al. Pictorial review:
radiation protection awareness in non-radiologists. Br J
Radiol 1997;70:102
-106[Abstract]
-
Miotto D, Feltrin G, Calamosca M. A radiation protection device for
use during percutaneous transhepatic examinations.
Radiology
1984;151:799[Abstract/Free Full Text]
-
Young AT, Morin RL, Hunter DW, et al. Surface shield: device to
reduce personnel radiation exposure. Radiology
1986;159:801
-803[Abstract/Free Full Text]
-
Miller DL, Vucich JJ, Cope C. A flexible shield to protect
personnel during interventional procedures. Radiology
1985;155:825[Abstract/Free Full Text]
-
Yaffe MY, Mawdsley GE, Lilley M, Servant R, Reh G. Composite
materials for X-ray protection. Health Phys
1991;60:661
-664[Medline]
-
Webster EW. Addendum to `Composite materials for X-ray protection.'
(letter) Health Phys
1991;61:917
-918
-
Webster EW. Experiments with medium-Z materials for shielding
against low-energy X-rays. Radiology
1966;86:146
-
Kelsey CA, King JN, Orrison WW Jr, et al. Scattered radiation
levels from a portable fluoroscopic/angiographic unit. Health
Phys 1989;57:817
-818[Medline]
-
Williams JR. Scatter dose estimation based on dose-area product and
the specification of radiation barriers. Br J Radiol
1996;69:1032
-1037[Abstract/Free Full Text]
-
Zweers D, Geleijns J, Aarts NJM, et al. Patient and staff radiation
dose in fluoroscopy-guided TIPS procedures and dose reduction, using dedicated
fluoroscopy exposure settings. (commentary) Br J
Radiol 1998;71:672
-676[Abstract]
-
McParland BJ. A study of patient radiation doses in interventional
radiological procedures. Br J Radiol
1998;71:175
-185[Abstract]
-
Vehmas T. Radiation exposure during standard and complex
interventional procedures. Br J Radiol
1997;70:296
-298[Abstract]
-
Anderson NE, King SH, Miller KL. Variations in dose to the
extremities of vascular/interventional radiologists. Health
Phys 1999;76[suppl 2]:S39
-S40[Medline]
-
Thomson KR, Brammall J, Wilson BC. A "flagpole"
lead-glass screen for radiographic procedures.
Radiology
1982;143:557
-558[Free Full Text]
-
Allison JD, Teeslink CR. A special procedures screen.
Radiology
1980;136:233
-234[Abstract/Free Full Text]
-
Gilula LA, Barbeir J, Totty WG, Eichling J. Radiation shielding
device for fluoroscopy. Radiology
1983;147:882
-883[Free Full Text]
-
Wall SD, Howe JM, Sawhney R. Human immunodeficiency virus infection
and hepatitis: biosafety in radiology. Radiology
1997;205:619
-628[Abstract/Free Full Text]
-
Mitchell DM. Aspects of HIV infection: current infection control
policies for HIV. Monaldi Arch Chest Dis
1997;52:371
-375[Medline]
-
Walker PP, Reynolds MT. Prophylaxis after occupational exposure to
HIV: universal precautions should be used during all surgical procedures.
(letter) BMJ
1998;316:702[Free Full Text]
-
Beltrami EM, Williams IT, Shapiro CN, Chamberland ME. Risk and
management of blood-borne infections in health care. Clin Microbiol
Rev 2000;13:385
-407[Abstract/Free Full Text]
-
Mortimer PP. Unusual transmissions of blood-borne viruses, with
special reference to HIV. J Hosp Infect
1999;43[suppl]:S145
-S148
-
Finkelstein Y, Markowitz ME, Rosen JF. Low-level lead-induced
neurotoxicity in children: an update on central nervous system effects.
Brain Res Rev
1998;27:168
-176[Medline]
-
Staudinger KC, Roth VS. Occupational lead poisoning. Am
Fam Physician 1998;57:719
-726[Medline]

CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
A.O. Ortiz, V. Natarajan, D.R. Gregorius, and S. Pollack
Significantly Reduced Radiation Exposure to Operators during Kyphoplasty and Vertebroplasty Procedures: Methods and Techniques.
AJNR Am. J. Neuroradiol.,
May 1, 2006;
27(5):
989 - 994.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
G. Singer
Occupational Radiation Exposure to the Surgeon
J. Am. Acad. Ortho. Surg.,
January 1, 2005;
13(1):
69 - 76.
[Abstract]
[Full Text]
[PDF]
|
 |
|