|
|
||||||||
Cardiopulmonary Imaging |
1 Department of Radiological Technology, College of Medical Sciences, Tohoku
University, 2-1 Seiryo-machi, Aoba-ku, Sendai 980-8575, Japan.
2 Department of Radiology, Tohoku University School of Medicine, 1-1
Seiryo-machi, Aoba-ku, Sendai 980-8574, Japan.
3 Department of Internal Medicine and Rehabilitation Science, Tohoku University
Graduate School of Medicine, 1-1 Seiryo-machi, Aoba-ku, Sendai 980-8574,
Japan.
Received October 16, 2003;
accepted after revision March 8, 2004.
Address correspondence to K. Chida
(chida{at}rad.cms.tohoku.ac.jp).
Abstract
|
|
|---|
MATERIALS AND METHODS. The entrance surface doses with cineangiography and fluoroscopy of acrylic phantoms were compared for 11 radiography systems at seven facilities; each performs more than 100 cardiac intervention procedures per year. The entrance surface dose for an acrylic plate (20 cm thick) was measured using a skin-dose monitor.
RESULTS. The maximum dose exceeded the minimum dose by 6.44 times for cineangiography and by 3.42 times for fluoroscopy. The entrance surface dose with acrylic plate was lower with digital-only acquisition (mean ± SD, 3.07 ± 0.84 mGy/sec) than with film recording (6.00 ± 3.04 mGy/sec). By contrast, the entrance surface frame dose, after correction for the cine frame rate, tended to be higher with digital acquisition than with film recording (0.210 ± 0.053 vs 0.179 ± 0.058 mGy/frame, respectively). CONCLUSION. The entrance surface dose was approximately 50% less with digital-only acquisition than with film recording. However, after correcting the dose for cine frame rate, filmless acquisition did not in itself reduce the exposure. For the surface dose to be reduced for cardiac interventional radiography, even with digital filmless radiography systems, a low recording speed is necessary for angiocardiography.
|
|
|---|
Recently, cardiac interventional radiography equipment has tended toward digital acquisition (filmless) archiving instead of cine film recording and pulsed fluoroscopy instead of conventional fluoroscopy. Pulsed fluoroscopy clearly reduces patient exposure [710]. By contrast, whether digital acquisition does reduce patient exposure in comparison with cine film recording is not clear, although the American College of Cardiology, Limacher et al. [11], and Aldridge et al. [12] recommend using digital-only cine acquisition to reduce radiation exposure in angiocardiography. For the exposure to patients during cardiac interventional radiography to be reduced, reducing the skin dose from both fluoroscopy and film recording is necessary. Therefore, this study simulated the skin radiation doses of patients in cardiac catheterization laboratories with various radiography systems used in cardiac interventional radiography to determine whether digital acquisition reduces patients' exposure as compared with cine film recording.
|
|
|---|
|
The entrance surface dose corrected for the cine frame rate (i.e., the frame dose) was determined as follows: frame dose (mGy / frame) = [original entrance surface dose with cineangiography (mGy / sec) / cine recording speed (frames / sec)].
The Student's t test was used to compare data for the two groups (digital-only acquisition vs film recording, pulsed fluoroscopy vs conventional fluoroscopy, and without vs with copper filtration). Statistical significance was defined as a p value of less than 0.05.
|
|
|---|
Cineangiography Dose
Digital versus film recording.The entrance surface dose
with the 20-cm-thick acrylic plate was significantly (p < 0.05)
lower with digital-only acquisition (mean ± SD, 3.07 ± 0.84
mGy/sec) than with film recording (6.00 ± 3.04 mGy/sec). By contrast,
the entrance surface frame dose, after correction for the cine frame rate,
tended to be higher with digital acquisition (0.210 ± 0.053 vs 0.179
± 0.058 mGy/frame, respectively), although the difference was not
significant (p = 0.375).
Without versus with added copper filtration.The entrance surface frame dose of the 20-cm-thick acrylic plate was not significantly lower in the systems with added copper filtration than in the systems without added beam filtration (0.172 ± 0.050 vs 0.205 ± 0.057 mGy/frame, respectively; p = 0.412).
Fluoroscopy Dose of the 11 Radiography Systems
For fluoroscopy, the average entrance surface dose of the 20-cm-thick
acrylic plate was 19.3 mGy/min (range, 8.930.4 mGy/min)
(Table 1). The maximum dose
exceeded the minimum dose by 3.42 times for the 20-cm-thick plate. The surface
doses were identical for continuous fluoroscopy (mean ± SD, 23.93
± 2.77 mGy/min) and pulsed fluoroscopy at 30 pulses/sec (25.20 ±
7.35 mGy/min).
|
|
|---|
Filmless laboratories have reduced the radiation exposure in cardiac interventional radiography. Some reports have indicated that filmless laboratories provide important further reductions in radiation exposure [11, 12]. Nevertheless, another report showed that the radiation dose was higher with a digital system than that with a conventional system in 13 of the 15 groups [15]. In our study, the entrance surface dose was approximately 50% less with digital-only acquisition than with film recording. Our results support the reports that skin radiation dose reduction is achieved in filmless laboratories.
One of the most important factors in reducing exposure is the use of a slower frame rate, even with digital-only acquisition. The use of digital acquisition in angiocardiography does not in itself reduce the surface dose, which varies markedly with the radiography system used. The maximum surface dose exceeded the minimum dose by 6.44 times for cineangiography and 3.42 times for fluoroscopy in the 11 radiography systems that we measured. By contrast, the maximum frame dose, after correction for the cine frame rate, exceeded the minimum frame dose by 2.30 times for cineangiography. The frame doses in digital-only acquisition were identical to those in cine film recording. In our study, the recording frame rate used in digital-only cine acquisition was 15 or 12.5 frames/sec and that of cine film recording was 15, 30, or 60 frames/sec. Therefore, the entrance surface dose of the acrylic plate tended to be lower for filmless acquisition than for cine film recording. For the surface dose to be reduced, the frame rate must be reduced in angiocardiographyboth in cine film recording and digital acquisition. Baim [16] stated that some laboratories record coronary images at speeds as low as 15 frames/sec. Therefore, to reduce the skin dose that patients receive from cineangiography during cardiac interventional radiography, we recommend using digital-only cine acquisition with a low recording speed.
The use of pulsed fluoroscopy can also reduce the skin exposure dose [710]. Our results showed that the entrance surface dose was significantly (p < 0.01) higher with continuous fluoroscopy and pulsed fluoroscopy at 30 pulses/sec (mean ± SD, 23.93 ± 2.77 mGy/sec) than with pulsed fluoroscopy at 15 or 7.5 pulses/sec (15.07 ± 4.13 mGy/sec). Therefore, pulsed fluoroscopy should be used at a low pulse rate, such as 15 pulses/sec, to reduce the skin dose [8].
In summary, digital acquisition and pulsed fluoroscopy do not necessarily reduce the radiation dose in cardiac interventional radiography; it is necessary to lower the recording speed with digital-only cine acquisition and to use pulsed fluoroscopy at a pulse rate lower than 30 pulses/sec.
The patient skin dose is usually reduced when added copper beam filtration is used [10, 11]. In our study, however, the entrance surface frame doses of the systems with added copper filtration were not significantly lower than those of the systems without added beam filtration, although we did not compare the same system with and without copper filtration, because we compared different systems in a clinical setting. Consequently, the effect of added copper filtration might be masked.
Physicians should also pay attention to the image intensifier field size used, degeneration of the image intensifier gain, patient body weight, and acquisition angle. A small image intensifier field (e.g., 9 inch [23 cm]) is frequently used instead of a large one (e.g., 11 inch [28 cm]). The radiation exposure with a small-field image intensifier is higher than that with a large-field image intensifier, which makes it even more important to use a lower recording speed with the filmless and pulsed fluoroscopic techniques. The image intensifier has variable magnification modes, so magnification should be used only when necessary. A much greater entrance skin dose is required in large patients and with steeply angled beam orientations.
Degeneration of the image intensifier gain increases the radiation dose. Our dose measurements for two identical pieces of equipment (B and K) differed. This disparity might have been caused by the difference in the number of years the image intensifier had been used (B, 32 months; K, 1 month) because the performance of an image intensifier deteriorates rapidly.
The maximum frame dose, after correction for the cine frame rate, exceeded the minimum frame dose by 2.30 times for cineangiography. Similarly, the maximum dose exceeded the minimum dose by 3.42 times for fluoroscopy. These differences might be caused mainly by the different tube potential (peak kilovoltage), degeneration of the image intensifier gain, insufficient quality control of the radiography unit, or a combination of these factors. Therefore, radiography units should be checked on an annual basis for optimal imaging performance, appropriate radiation dose, and so forth.
Physicians should be aware of the entrance surface dose for cineangiography and fluoroscopy with the radiography system that they use for cardiac interventional radiography. In a clinical setting, the radiation dose varies markedly with the procedure and the physician. In cardiac interventional radiography, the most prevalent factor in radiation injury is prolonged exposure of one area of skin; physicians should therefore change the X-ray and viewing angles. The use of digital-only cine acquisition with a lower recording speed and of pulsed fluoroscopy at a pulse rate lower than 30 pulses/sec should reduce the skin dose. Nevertheless, the entrance surface doses in the radiography systems tested were too large. If the surface dose of the radiography system that they use is high, physicians should determine why and make efforts to reduce patients' exposure.
This study has a limitation in that there were insufficient controls for the different radiography units examined. However, it was difficult to establish a control because the measurements were made using the clinical techniques practiced at each site under clinical conditions.
In conclusion, the entrance surface doses with cineangiography and fluoroscopy of acrylic phantoms were compared for 11 radiography systems at seven facilities; each performs more than 100 cardiac interventional radiography cases per year. The maximum surface dose exceeded the minimum dose by 6.44 times for cineangiography and by 3.42 times for fluoroscopy. The entrance surface dose of the acrylic plate tended to be lower with digital acquisition (filmless) than with cine film recording. However, after correcting the dose for cine frame rate, filmless acquisition did not in itself reduce the exposure. For a reduction in the surface dose for cardiac interventional radiography, even with digital filmless radiography systems, a low recording speed is necessary in angiocardiography and a low pulse rate in pulsed fluoroscopy. To reduce the risk of skin injury to patients, physicians should be aware of the entrance surface dose of the radiography system that they use for cardiac interventional radiography.
Acknowledgments
We thank Masatoshi Sasaki of the Tohoku University Hospital, Takashi
Kanetomo of the Sendai National Hospital, Souichiro Kamio of Sendai City
Hospital, Yoshikatu Kawamura of the Tohoku Kosai Hospital, Takashi Terui of
the Sendai Kousei Hospital, Hiroo Chiba of the Tohoku Kouseinenkin Hospital,
and Masami Sato of the Sendai Open Hospital for their invaluable
assistance.
|
|
|---|
This article has been cited by other articles:
![]() |
K. Chida, Y. Kagaya, H. Saito, Y. Takai, S. Takahashi, S. Yamada, M. Kohzuki, and M. Zuguchi Total Entrance Skin Dose: An Effective Indicator of Maximum Radiation Dose to the Skin During Percutaneous Coronary Intervention Am. J. Roentgenol., October 1, 2007; 189(4): W224 - W227. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. M. Colletti Cardiac Imaging 2006 Am. J. Roentgenol., June 1, 2006; 186(6_Supplement_2): S337 - S340. [Full Text] [PDF] |
||||
![]() |
H. T. Abada, C. Larchez, B. Daoud, A. Sigal-Cinqualbre, and J.-F. Paul MDCT of the Coronary Arteries: Feasibility of Low-Dose CT with ECG-Pulsed Tube Current Modulation to Reduce Radiation Dose Am. J. Roentgenol., June 1, 2006; 186(6_Supplement_2): S387 - S390. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. Chida, H. Saito, H. Otani, M. Kohzuki, S. Takahashi, S. Yamada, K. Shirato, and M. Zuguchi Relationship between fluoroscopic time, dose-area product, body weight, and maximum radiation skin dose in cardiac interventional procedures. Am. J. Roentgenol., March 1, 2006; 186(3): 774 - 778. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |