Fig. 4. Graph of temperature against sonication time. Green
represents average temperature within target volume; red represents absolute
temperature at specified point. This sonication has resulted in maximum
temperature of 91.8°C. This is higher than is necessary for irreversible
damage so this information can be used to shorten sonication time or reduce
acoustic power for subsequent sonications. Figure is screen capture from
Ex-Ablate 2000 system (InSightec) during treatment.