Fig. 3A. Aortic image noise of CT. Scatterplots show aortic image noise of CT obtained by fixed tube current–second milliampere-seconds (mAs) (A), by mAs modulation based on body weight (B), and by mAs modulation based on scout image data (C). SDs of aortic image noise were 2.03 (A), 1.61 (B), and 1.06 (C), respectively. SD of image noise was minimized with statistically significant differences when mAs was modulated on basis of scout image data (C) (p < 0.022, F test). Coefficients of determination (R2) were 0.644 (A), 0.047 (B), and 0.03 (C).