AJR 2005; 185:84-85
© American Roentgen Ray Society
Pediatric Return Electrodes for Radiofrequency Ablation in Children
Paul R. Morrison1,
Stephen D. Brown2 and
Eric vanSonnenberg3
1 Radiology, Brigham and Women's Hospital, Boston, MA 02115.
2 The Children's Hospital, Boston, MA.
3 Dana-Farber Cancer Institute, Boston, MA.
Received August 20, 2004;
accepted after revision November 2, 2004.
Address correspondence to P. R. Morrison
(pmorrison{at}partners.org).
Introduction
Radiofrequency energy can be delivered interstitially to destroy
tissues thermally in situ for a therapeutic effect. This process is generally
referred to as radiofrequency ablation. Although percutaneous radiofrequency
ablation has been applied extensively in adults to a range of pathologic
conditions
[1-5],
its use to treat tumors percutaneously in children has been limited primarily
to treating osteoid osteomas
[6]. Considering the increasing
interest and breadth of use in adults, percutaneous radiofrequency ablation
will likely be used for a range of applications in children as well.
Radiofrequency ablation in small children, however, raises technical concerns
because procedures typically involve the deposition of radiofrequency energy
at relatively high powers and currents (> 100 W;
1.5 A) for extended
periods (several minutes); this raises technical concerns in small children.
One such concern is the safe return of the applied current and the potential
hazard of skin burns at the return electrodes (grounding pads). In this
report, we describe the use of pediatric return electrodes for radiofrequency
ablation in the kidney of a young patient for whom standard adult pads could
not be used in a manner consistent with the manufacturer's instructions for
use.
Subjects and Methods
The procedure was performed on a 16-kg, 5-year-old girl with a 1.8-cm
central tumor (nephroblastoma) in her solitary right kidney. The
radiofrequency ablation procedure was approved by our institutional review
board under an innovative therapy mechanism.
Radiofrequency ablation was performed with a 2.0-cm-diameter array-style
electrode (LeVeen, Boston Scientific) connected to a radiofrequency generator
(RF 3000, Boston Scientific) capable of providing up to 200 W of power at a
frequency of 460 kHz. The electrode was placed percutaneously into the tumor
under CT guidance. Four rectangular, 12.1 x 9.5 cm infant return
electrodes (Infant REM PolyHesive II, ValleyLab) were applied; two pads were
placed on each thigh, one anterior and one posterior, with the long axis of
each parallel to the femur (Fig.
1). These are marketed for use in monopolar electrosurgery.

View larger version (130K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 1 Photograph of return electrodes, or grounding pads, for
radiofrequency ablation: 18.3 x 11.4 cm adult pad (left) and 12.1
x 9.5 cm infant pad (right). Smaller pad has 55% of area of larger adult
version; four such infant pads were used in case discussed.
|
|
Radiofrequency energy was delivered with the power adjusted incrementally
from an initial power setting of 10 W. The power was increased 5 W/min up to a
maximum of 60 W. Subsequently, a second phase of treatment at the same site
was applied at 40-45 W for 6.5 min and concluded the treatment. During
ablation, the grounding pads were manually assessed for excessive warmth, and
the skin around their edges was visually inspected for erythema.
Results
From a technical standpoint, the ablation was executed without incident.
After a total of 19 min in the first phase of treatment, tissue impedance rose
rapidly from an average value of approximately 70 ohms to several hundred
ohms. This increase in impedance, indicative of tissue coagulation, is the
clinical end point for the therapy using this particular ablation device. No
operational problems with the radiofrequency generator occurred from using the
pediatric return electrodes. The return electrodes did not get hot to the
touch, and no burns on the skin were observed either during the procedure or
at the end, when the pads were removed. Postprocedural imaging at 24 hr showed
evidence of complete ablation of the tumor.
Discussion
Because of the small circumference of each of the child's thighs, standard
large adult pads would not fit without considerable overlap, regardless of pad
orientation. The use of four adult pads (two on each thigh) is the standard
recommended in the instructions for use (Boston Scientific), which were
written for use in adults. Uneven or skewed alignment, uneven skin contact,
overlap, or an insufficient number of pads (and thus an inadequate surface
area for returning current) add to the possibility of an excessive
concentration of return current on the pad and therefore the risk of skin
burns [7,
8]. Furthermore, the
radiofrequency generator electronically monitors current to maintain an even
distribution throughout the pads and will interrupt transmission if excessive
current is detected. Thus, using substantially overlapping adult pads or only
two adult pads could have presented device error messages that may have been
hard to resolve, interrupting the procedure.
By using the smaller infant pads, we were able to have four well-placed,
properly applied pads with no overlap. The total surface area provided was
460.0 cm2 (55% of the 834.5 cm2 provided by four adult
pads [Adult REM PolyHesive II, ValleyLab]). In this procedure, the maximum
power applied was 60 W with a current of approximately 0.9 A, and the
treatment time was several minutes. Despite the reduced surface area and the
potential for increased heating at the return pads, the device operated as
expected in this procedure and the skin was not burned.
Ideally, the location and orientation of the pad should be in accordance
with the manufacturer's instructions. In this case, we opted for the use of an
alternative pad to get good placement on the thighs for the best electric
conduction. For future cases, if the circumference of each thigh allows, we
would recommend placing each pad with the long axis perpendicular to the femur
to maximize the leading edge of the pad for the return current
[9]. Separately, caution should
be exercised in the use of electrode arrays larger than 2 cm that may
necessitate higher applied powers or extended durations of energy deposition,
since these parameters add to the risk of untoward tissue heating. The sites
of the grounding pads should be checked manually and visually throughout the
procedure.
Although an application in a single case is not definitive, we suggest that
the judicious use of these pads is safe and advisable for radiofrequency
ablation in pediatric patients. Two technical safety concerns were addressed
successfully in this case by the use of the infant grounding pads: The skin
was not burned despite the reduced total surface area for the return current,
and the operation of the radiofrequency generator itself presented no
problems.
Acknowledgments
We acknowledge the assistance and advice of Chris Pearson and Thom
Lawson.
References
- Livraghi T, Solbiati L, Meloni F, Ierace T, Goldberg SN, Gazelle
GS. Percutaneous radiofrequency ablation of liver metastases in potential
candidates for resection: the "test of time approach."
Cancer 2003;97:3027
-3035[CrossRef][Medline]
- Curley SA, Marra P, Beaty K, et al. Early and late complications
after radiofrequency ablation of malignant liver tumors in 608 patients.
Ann Surg 2004;239:450
-458[CrossRef][Medline]
- Zagoria RJ. Percutaneous image-guided radiofrequency ablation of
renal malignancies. Radiol Clin North Am2003; 41:1067
-1075[CrossRef][Medline]
- Steinke K, Sewell PE, Dupuy D, et al. Pulmonary radiofrequency
ablation: an international study survey. Anticancer
Res 2004;24:339
-343[Abstract/Free Full Text]
- Wood BJ, Ramkaransingh JR, Fojo T, Wather MM, Libutti SK.
Percutaneous tumor ablation with radiofrequency.
Cancer 2002;94:443
-451[CrossRef][Medline]
- Torriani M, Rosenthal DI. Percutaneous radiofrequency treatment of
osteoid osteoma. Pediatr Radiol2002; 32:615
-618[CrossRef][Medline]
- de Baere T, Risse O, Kuoch V, et al. Adverse events during
radiofrequency treatment of 582 hepatic tumors. AJR2003; 181:695
-700[Abstract/Free Full Text]
- Steinke K, Gananadha S, King J, Zhao J, Morris DL. Dispersive pad
site burns with modern radiofrequency ablation equipment. Surg
Laparosc Endosc Percutan Tech2003; 13:366
-371[Medline]
- Goldberg SN, Solbiati L, Halpern EF, Gazelle GS. Variables
affecting proper system grounding for radiofrequency ablation in an animal
model. J Vasc Interv Radiol2000; 11:1069
-1075[Medline]

CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
S. D. Brown and E. vanSonnenberg
Issues in Imaging-Guided Tumor Ablation in Children Versus Adults
Am. J. Roentgenol.,
September 1, 2007;
189(3):
626 - 632.
[Abstract]
[Full Text]
[PDF]
|
 |
|