Fig. 3B All images show right common and external iliac artery after
implantation of self-expandable nitinol stent (diameter, 10 mm; length, 44 mm)
in 66-year-old man. Maximum intensity projection reconstruction of MR
angiography image acquired immediately after procedure (MR angiography 1).
Bandlike artifacts are present at both ends of stent (stent ends indicated by
arrows). Artifact at proximal stent end is more pronounced and was
misinterpreted by all three observers as stenosis. Signal in stent is
inhomogeneous and weakened in comparison with adjacent vessel segments.