DOI:10.2214/AJR.04.1353
AJR 2006; 186:416-423
© American Roentgen Ray Society
Digital Volume Tomography: Radiologic Examinations of the Temporal Bone
Carsten V. Dalchow1,
Alfred L. Weber2,
Naoaki Yanagihara3,
Siegfried Bien4 and
Jochen A. Werner1
1 Department of Otolaryngology, Head and Neck Surgery, Philipps University
Marburg, Deutschhausstr. 3, 35037 Marburg, Germany.
2 Department of Radiology, Massachusetts Eye and Ear Infirmary, Boston,
MA.
3 Department of Otolaryngology, Takanoko Hospital, Matsuyama Ehime, Japan.
4 Department of Neuroradiology, Philipps University Marburg, Marburg,
Germany.
Received August 27, 2004;
accepted after revision November 29, 2004.
Address correspondence to C. V. Dalchow
(dalchow{at}med.uni-marburg.de).
Abstract
OBJECTIVE. We evaluated the clinical applicability and the value of
digital volume tomography for visualization of the lateral skull base using
temporal bone specimens.
MATERIALS AND METHODS. Twelve temporal bone specimens were used to
evaluate digital volume tomography on the lateral skull base. Aside from the
initial examination of the temporal bones, radiologic control examinations
were performed after insertion of titanium, gold, and platinum middle-ear
implants and a cochlear implant.
RESULTS. With high-resolution and almost artifact-free visualization
of alloplastic middle-ear implants of titanium, gold, or platinum, it was
possible to define the smallest bone structures or position of the prosthesis
with high precision. Furthermore, the examination proved that digital volume
tomography is useful in assessing the normal position of a cochlear
implant.
CONCLUSION. Digital volume tomography expands the application of
diagnostic possibilities in the lateral skull base. Therefore, we believe
improved preoperative diagnosis can be achieved along with more accurate
planning of the surgical procedure. Digital volume tomography delivers a small
radiation dose and a high resolution coupled with a low purchase price for the
equipment.
Keywords: cochlear implant CT digital volume tomography middle ear middle-ear implant stapesplasty temporal bone
Introduction
In the diagnosis of diseases in the lateral skull base, detailed images
with high resolution are mandatory for visualization of small pathologic
processes. This requirement explains why CT, introduced in the 1970s, has
completely replaced conventional radiodiagnostic techniques
[1-3].
With the development of high-resolution CT in the 1980s, it became possible to
recognize pathologic processes of the temporal bone in early stages.
High-resolution CT has been the method of choice for a long time to assess
conductive hearing loss, ear malformations, and destructive processes in the
middle and inner ears [3]. It
is possible to visualize soft tissues and bone structures because of the
detailed display.
Further improvement and optimization of temporal bone imaging with higher
resolution is desirable, especially if done without an increase in the
radiation dose or additional time for data calculation. An imaging technique
with these improvements would allow integration of this new technique into
routine daily practice.
Digital volume tomography has been increasingly used in dental surgery
[4,
5]. Early in the development of
this technique, its value as a high-resolution technique was recognized in the
precise planning of dental implantation
[6]. It became evident that
precise preoperative visualization with the help of 3D reconstruction of the
operating field could be accomplished and complications avoided in cases with
insufficient bone of the alveolar part of the maxilla or mandible or in the
presence of ectopic teeth [4,
7].
Further advancement of panoramic tomography in dental surgery in 2000 led
to the introduction of the digital volume tomograph 3DX multiimage micro CT
(J. Morita Manufacturing Corp.). Using digital volume tomography, its high
resolution and minimum section distance of 0.125 mm allow small pathologic
changes to be visualized. With this method, axial, coronal, and sagittal
images can be obtained at one examination as opposed to only axial sections
with conventional CT. Additional sections in different planes can be
reconstructed from the original data. These first dental examinations
suggested that digital volume tomography in diseases occurring in the lateral
skull base may prove helpful. With limited space requirements, low equipment
cost, and a short data calculation, digital volume tomography appears helpful
with its increased spatial resolution
[8]. Because no experience with
digital volume tomography has been reported in the lateral skull base, we
evaluated this new technique on temporal bone specimens.
Previous diagnostic techniques such as high-resolution CT or
pluridirectional tomography have been evaluated on anatomic specimens
[9-11].
To answer these questions regarding clinical applicability and the value of
digital volume tomography, we also began our investigation with temporal bone
specimens. This allowed us to correlate the radiologic with microscopic
findings during preparation of the specimens and to assess the results.
Materials and Methods
The Accuitomo (J. Morita Manufacturing Corp.)
(Fig. 1) was used to perform
our examination. From 512 single images with reconstruction increments of
0.125 mm and a cylindric format, 3-cm-high and 4-cm-wide (transverse diameter)
images were reconstructed. Routine otoscopic surgical procedures with the
operating microscope were performed on temporal bone specimens after removal
of excess soft tissues (Carl Zeiss) (Table
1). Subsequently, radiologic control examinations were performed
with digital volume tomography.

View larger version (111K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 1 General view of digital volume tomograph (Accuitomo, J. Morita
Manufacturing Corp.) with patient's examination chair, radiograph device,
4-inch (10-cm) image intensifier, and control panel integrated in right
column.
|
|
After targeting the region of interest (ROI) with the help of target light
beams, the examination was performed with 60 kV (X-ray tube voltage) and 2 mA
(X-ray tube electric current). The temporal bone specimen was attached to a
holding device (Temporal Bone Holder, Model Wuerzberg, Storz Medical) and
placed on the examination chair in a way that the metallic parts of the device
were outside the radiation beam (Fig.
2). As a consequence, the specimens were not examined in typical
conventional planes. During the exposure, the X-ray tube rotated along with
the opposite sensor in 18 sec and 360° around the center of a
conical-shaped radiation beam that hits a 4-inch (10-cm) magnifying screen.
With a focal spot of 0.5 mm, the distance from the source to rotational center
was 33.5 cm and from the source to image intensifier, 63.5 cm. After
acquisition of 512 single images (frames) with a resolution of 240 x 320
pixels with a PC (Pentium 4, Intel; 2.3 GHz) and a volume composed of single
units (voxels) 0.125 mm3 in diameter, calculations of the ROI were
performed (Table 2). The images
of the temporal bone specimens were analyzed with special software (3DX
Integrated Information System version 1.52, J. Morita Manufacturing
Corp.).
The data were displayed after reconstruction on a PC monitor in three
planes with vertical orientation to each other and a minimal intersection
distance and a minimal section thickness of 0.125 mm. The section angles,
section thickness, and the intersection distance were changed at will. The
individual structures could be selectively depicted; however, in addition, the
sections in the axial, coronal, and sagittal planes were separately displayed
and the results shown on a monitor. As an alternative procedure, the collected
data were transmitted via the DICOM port and stored in variable picture
formats or printed.
We initially examined 12 temporal bone specimens with digital volume
tomography. We investigated the middle ear space, auditory canals, and the
mastoid air cells in three planes vertical to each other. Image sections were
selected to illustrate the oval window niche, stapes footplate, and cochlea
with the labyrinth (Fig. 3A).
By adapting the angle of the sections, we were able to show the ossicular
chain in two different planes. We selected a plane that would illustrate the
incus and stapes, the incudostapedial joint, and the stapes footplate together
with the oval window niche (Fig.
3B). In a second plane, we showed the head and the handle of the
malleus. In addition to the epitympanum, we delineated the facial canal
(Fig. 3C). We showed the facial
nerve canal in its entire course from the internal auditory canal to the
stylomastoid foramen (Fig.
3D).

View larger version (138K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3A Digital volume tomography images show general view of middle and
inner ear of temporal bone specimen. Co = cochlea, ICA = internal carotid
artery, JB = jugular bulb, MEC = middle ear cavity. Image shows MEC and inner
ear with labyrinth and pneumatized temporal bone with incudostapedial joint.
Vestibule and semicircular canals (yellow arrows) and facial nerve
canal (black arrow) are seen. Co = cochlea, ICA = internal carotid
artery.
|
|

View larger version (129K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3B Digital volume tomography images show general view of middle and
inner ear of temporal bone specimen. Co = cochlea, ICA = internal carotid
artery, JB = jugular bulb, MEC = middle ear cavity. Image shows vestibule
(yellow arrow) and semicircular canals, facial nerve canal (black
arrow), stapes arch (red arrow), jugular bulb (JB), and long
process of incus (white arrow).
|
|

View larger version (116K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3C Digital volume tomography images show general view of middle and
inner ear of temporal bone specimen. Co = cochlea, ICA = internal carotid
artery, JB = jugular bulb, MEC = middle ear cavity. Image shows malleus head
(green arrow), malleus handle (gray arrow), vestibule and
semicircular canals (yellow arrow), and facial nerve canal (black
arrow).
|
|

View larger version (135K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3D Digital volume tomography images show general view of middle and
inner ear of temporal bone specimen. Co = cochlea, ICA = internal carotid
artery, JB = jugular bulb, MEC = middle ear cavity. Image shows entire facial
nerve canal (black arrows) between ganglion geniculi (gg) and foramen
stylomastoideum (fs) and vestibule and semicircular canals (yellow
arrow).
|
|
For radiologic illustration of middle ear implants on digital volume
tomography, an incus stapedotomy was performed with a titanium piston
(K-Piston, 5.25 x 0.4 mm, Heinz Kurz GmbH), gold piston (Gold Piston
with band hook, 5.5 x 0.6 mm, Heinz Kurz GmbH), and a platinum-band
polytetrafluoroethylene (Teflon, DuPont) piston (Platinum-Teflon Prosthesis
Type Schuhknecht, 5.5 x 0.4 mm, Audio Technologies). For an incus
stapedotomy, the prosthesis was affixed to the loop of the long process of the
incus and placed across a created perforation in the stapes footplate to the
adjacent vestibule.
Three ossiculoplasties were performed with different implants. One
tympanoplasty type 3 was done with an autologous incus
(Fig. 4A) and with a partial
implant made of gold (Bell Prosthesis, 2.0 mm, Heinz Kurz GmbH)
(Fig. 4E). In another
ossiculoplasty, a total implant made from pure titanium (titanium Total
Implant, 7.5 mm, Spiggle & Theis) was shortened to 4.5 mm and used
(Fig. 4D). For incus
interposition, the patient's incus was removed, shaped with a diamond drill,
and interposed between the handle of the malleus and head of the stapes. The
partial prothesis was placed below the handle of the malleus and placed onto
the head of the stapes. For the preparation, the Total Implant was placed onto
the middle of the stapes footplate below the handle of the malleus. After a
macroscopic and radiologic evaluation, a prosthesis dislocation was simulated.
For this purpose, the total prosthesis made of pure titanium was repositioned
from its optimal position in the middle of the stapes footplate onto the
margin of the footplate, adjacent to bone structures next to the facial canal.
Subsequently, a radiologic control examination with digital volume tomography
was performed.

View larger version (124K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4A Comparison of otoscopic views of ossiculoplasty with operation
microscope (A, C, E) and radiologic control digital volume tomography
images (B, D, F) of temporal bone specimen. Co = cochlea, EAC =
external auditory canal, ET = eustachian tube, IAC = internal auditory canal,
ICA = internal carotid artery, TM = tympanic membrane. Images show autologous
incus interposition (green arrows), oval window niche (white
arrow, A), round window niche (black arrow, A),
and vestibule (yellow arrow, B). White arrow in B
indicates long process of incus, gray arrow indicates malleus handle atop
incus interposition.
|
|

View larger version (117K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4E Comparison of otoscopic views of ossiculoplasty with operation
microscope (A, C, E) and radiologic control digital volume tomography
images (B, D, F) of temporal bone specimen. Co = cochlea, EAC =
external auditory canal, ET = eustachian tube, IAC = internal auditory canal,
ICA = internal carotid artery, TM = tympanic membrane. Images show titanium
implant (blue arrows) (titanium Total Implant [7.5 mm], Spiggle &
Theis), malleus handle (gray arrows), round window niche (black
arrow, E) and vestibule (yellow arrow, F).
|
|

View larger version (104K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4D Comparison of otoscopic views of ossiculoplasty with operation
microscope (A, C, E) and radiologic control digital volume tomography
images (B, D, F) of temporal bone specimen. Co = cochlea, EAC =
external auditory canal, ET = eustachian tube, IAC = internal auditory canal,
ICA = internal carotid artery, TM = tympanic membrane. Images show partial
gold implant (red arrows), malleus handle (gray arrows),
round window niche (black arrow, C), and vestibule (yellow
arrow, D).
|
|
To evaluate the position of a cochlear implant, implantation was done on a
temporal bone specimen. First, a superior mastoidectomy with preservation of
the posterior wall was done. After a posterior tympanotomy with skeletonizing,
the mastoid segment of the facial canal, round window niche, the
incudostapedial joint, and the promontory were identified. A cochleostomy with
opening of the scala tympani was performed. Subsequently, an electrode was
inserted through the opening to a predetermined demarcated point in the
cochlea. After evaluation of the radiologic findings, a comparison examination
was made on the temporal bone specimen with the operating microscope.
Results
In the initial radiologic examination of the 12 temporal bones and
subsequent preparation of the specimens, the external auditory canal, middle
ear cavity with ossicular chain, cochlea, and the labyrinth and the mastoid
air cells could be identified on high-resolution images. An analysis of the
images showed no pathologic changes in the mentioned structures. Based on the
radiologic findings, the preparation of the temporal bone specimens was made,
comparing the microscopic findings with the initial digital volume tomography
images. The anatomic relationships encountered during the preparation
correlated with those obtained from the radiologic findings.
After the initial evaluation by digital volume tomography, the surgical
dissection was performed and repeat digital volume tomography was performed.
By inspection of the temporal bones with a stapes prosthesis inserted, it was
possible to identify the loop of all three prostheses that were affixed to the
long process of the incus (Figs.
5A,
5B, and
5C). The inserted titanium and
gold prostheses could be shown to the level of the vestibule. The
platinum-band Teflon piston was visualized to a level close to the stapes
footplate. At the level of the stapes footplate and vestibule, no prosthesis
was recognized (Fig. 5C).
Blurring from artifacts stemming from implanted metal did not occur.
Subsequently, the findings on digital volume tomography were compared with the
findings of the operating microscope. All three prostheses were normally
positioned in situ. The platinum wire of the platinum-wire Teflon piston
terminated above the level of the footplate while the Teflon mantle extended
further above the opening of the footplate into the vestibule. Consequently,
the macroscopic results were in agreement with the findings seen with digital
volume tomography.

View larger version (126K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5A Artifact-free digital tomography images of temporal bone specimen.
Co = cochlea, EAC = external auditory canal, IAC = internal auditory canal.
Images of stapes piston show long process of incus (white arrow,
A), stapes footplate, and tympanic segment of facial canal (black
arrows). Titanium implant (green arrow, A) (K-Piston
[5.25 x 0.4 mm], Heinz Kurz GmbH) and gold implant (red arrow,
B) (Gold Piston [5.5 x 0.4 mm], Heinz Kurz GmbH) are recognizable
reaching from incus into vestibule (yellow arrows).
|
|

View larger version (125K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5B Artifact-free digital tomography images of temporal bone specimen.
Co = cochlea, EAC = external auditory canal, IAC = internal auditory canal.
Images of stapes piston show long process of incus (white arrow,
A), stapes footplate, and tympanic segment of facial canal (black
arrows). Titanium implant (green arrow, A) (K-Piston
[5.25 x 0.4 mm], Heinz Kurz GmbH) and gold implant (red arrow,
B) (Gold Piston [5.5 x 0.4 mm], Heinz Kurz GmbH) are recognizable
reaching from incus into vestibule (yellow arrows).
|
|

View larger version (131K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5C Artifact-free digital tomography images of temporal bone specimen.
Co = cochlea, EAC = external auditory canal, IAC = internal auditory canal.
Image shows platinum-band polytetrafluoroethylene (Teflon, DuPont) piston
(blue arrow) (Platinum-Teflon Prosthesis Type Schuhknecht [5.5
x 0.6 mm], Heinz Kurz GmbH) at level close to stapes footplate.
Vestibule (yellow arrow) and tympanic segment of facial canal
(black arrow) are also visible.
|
|
Control examinations of the ossicular prosthesis on digital volume
tomography likewise revealed no artifacts. The interpolated autologous incus
was visible at the lower part of the tympanic membrane, adjacent to the handle
of the malleus, and likewise at the recognizable footplate
(Fig. 4B). The position of the
gold implant could be identified on digital volume tomography. This implant
likewise was positioned below the handle of the malleus on the stapes
footplate (Fig. 4D). The
proper position of the total implant made of pure titanium could be shown on
digital volume tomography. The whole implant was recognized and the shaft of
the prosthesis could be differentiated from the prosthesis head plate and foot
(Fig. 4F).

View larger version (136K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4B Comparison of otoscopic views of ossiculoplasty with operation
microscope (A, C, E) and radiologic control digital volume tomography
images (B, D, F) of temporal bone specimen. Co = cochlea, EAC =
external auditory canal, ET = eustachian tube, IAC = internal auditory canal,
ICA = internal carotid artery, TM = tympanic membrane. Images show autologous
incus interposition (green arrows), oval window niche (white
arrow, A), round window niche (black arrow, A),
and vestibule (yellow arrow, B). White arrow in B
indicates long process of incus, gray arrow indicates malleus handle atop
incus interposition.
|
|

View larger version (128K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4F Comparison of otoscopic views of ossiculoplasty with operation
microscope (A, C, E) and radiologic control digital volume tomography
images (B, D, F) of temporal bone specimen. Co = cochlea, EAC =
external auditory canal, ET = eustachian tube, IAC = internal auditory canal,
ICA = internal carotid artery, TM = tympanic membrane. Images show titanium
implant (blue arrows) (titanium Total Implant [7.5 mm], Spiggle &
Theis), malleus handle (gray arrows), round window niche (black
arrow, E) and vestibule (yellow arrow, F).
|
|
Simulation of a dislocation of the titanium Total Implant was diagnosed on
digital volume tomography. In this case, the titanium Total Implant was
replaced during preparation out of its ideal position in the middle of the
stapes footplate. As shifting of the implant foot onto the edge of the stapes
footplate next to the tympanic segment of the facial nerve canal would lead to
a conductive hearing loss and make an ossiculoplasty revision necessary, this
preparation was made.
Digital volume tomography visualized the minimal change in implant
positioning confirming the dislocation of the prosthesis. The electrode
position of the cochlear implant (Nucleus 24 Contour, Cochlear Ltd.) was
checked in the temporal bone specimen after insertion with the operating
microscope through the cochleostomy into the basal turn of the cochlea. The
implant was positioned with its third ring in the cochleostomy opening. The
position of the intracochlear electrode was shown on digital volume tomography
at different layers. The electrode in the cochleostomy could be seen with the
major portion of the individual electrodes within the cochlea adjacent to the
modiolus (Fig. 6). By analysis
of different imaging planes, it was possible to identify all 22 intracochlear
electrodes.

View larger version (139K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 6 Digital tomography image of temporal bone specimen shows
intracochlear position of cochlear implant (Nucleus 24 Contour, Cochlear Ltd.)
in basal turn (red arrow) of cochlea and cochleostomy (green
arrow). Eight single basal electrodes are identified with this image in
this section. MEC = middle ear cavity, EAC = external auditory canal.
|
|

View larger version (111K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4C Comparison of otoscopic views of ossiculoplasty with operation
microscope (A, C, E) and radiologic control digital volume tomography
images (B, D, F) of temporal bone specimen. Co = cochlea, EAC =
external auditory canal, ET = eustachian tube, IAC = internal auditory canal,
ICA = internal carotid artery, TM = tympanic membrane. Images show partial
gold implant (red arrows), malleus handle (gray arrows),
round window niche (black arrow, C), and vestibule (yellow
arrow, D).
|
|
Discussion
The introduction of high-resolution CT has expanded diagnostic
possibilities, especially for temporal bone surgery
[1,
3]. With this diagnostic tool,
pathologic changes can be recognized at an earlier stage compared with
conventional radiodiagnostic methods. With high-resolution CT, the surgical
approach can be planned and performed with precision
[9]. The 3D visualization
allows a detailed evaluation of important structures and simultaneously
facilitates orientation for preoperative planning. On the basis of this
examination technique, routinely performed with 1-mm sections, not all
structures are displayed in detail. When 3D reconstruction of a larger region
(obtained from data in the database) is performed, high-resolution display
with only slight loss of detail occurs
[2,
3,
9,
10]. To provide even more
sufficient information, multisectional helical CT with reconstruction
increments of 0.3 mm can be done. For highest resolution, a proportional
increase in the radiation dose must occur. A distinctly higher resolution
obtained with multisectional helical CT is possible with phase
contrast-fluorescence tomography
[12]. With this examination
technique, small changes in the inner ear and labyrinth are shown. However,
the radiation dose with this method, which is applied to modify and develop
cochlear implant electrodes, is very high. Consequently, this technique is
only used in the temporal bone laboratory.
The preoperative diagnosis before insertion of a cochlear implant electrode
is of equal value for the postoperative assessment of the proper position of
the electrode in the cochlea
[13]. An exact diagnosis can
prevent possible intraoperative complications. Furthermore, precise
postoperative localization of individual electrodes in the cochlea with
respect to their position to the modiolus contributes to the assessment of
potential hearing improvement. Our examination of the temporal bone specimens
shows that digital volume tomography will visualize even the small bone
structures of the lateral skull base. Aside from visualization of the tympanic
cavity and mastoid air cells, the cochlea and labyrinth are visualized. One
advantage is the ability to change the section angle, section thickness, and
intersectional distance after the examination to evaluate targeted individual
structures. The detailed local resolution, confirmed already in dental surgery
[8,
13-16],
helps define the ossicles precisely. On the basis of a high resolution with a
reconstruction increment of 0.125 mm, small structures such as the ossicles
can be demonstrated in a way not always possible with other alternative
methods.
Ongoing examinations have shown that the effective radiation dose of
examinations with digital volume tomography, with its limited cone beam, has a
very small value. A Rando woman phantom (UD-170A and UD-110S, Panasonic) has
been used as a subject measuring the effective dose of various organs at the
time of projection with digital volume tomography. Focusing the projection,
for example, on the upper left molar teeth, the effective dose was calculated
as follows: first, doses for the thyroid gland, lungs, esophagus, stomach,
colon, liver, urinary bladder, breasts, ovary, testis, red bone marrow, bone
surface, thymus, kidneys, small intestines, upper large intestines, skin,
brain, eyes, and salivary gland were determined. Each value was corrected with
its tissue factor and equivalent dose, and added together to obtain the
effective dose. For examinations of the upper molar teeth, temporal mandibular
joint, and the middle ear, the resultant effective doses were 6.3 µSv, 9.3
µSv, and 14.2 µSv, respectively. The obtained figures were approximately
1/300-1/100 of that of helical CT
[8,
17].
Because of the flexibility of the examination, accompanied by the same
quality, single and connected structures can be imaged in detail. It is
possible to recognize an interruption of the ossicular chain preoperatively.
On the basis of the anatomic location, the entire ossicular chain can be
visualized in two different sections on digital volume tomography. Important
details are depicted, allowing evaluation of the continuity of the ossicular
chain to include the ability to exclude ossicular discontinuity. From the
available images, section planes can be selected to illustrate the incus and
stapes, including the incudostapedial joint and the oval window niche with the
stapes footplate and malleus head and handle. In addition to the epitympanic
recess, the bone portion of the facial canal is visible. A soft-tissue mass or
erosion of the scutum can be excluded.
Visualization of the location and course of the bony canal of the facial
nerve is accomplished by identifying the location of the nerve between the
internal auditory canal and the stylomastoid foramen. This is of significance
in questionable erosion of the facial canal with exposure of the nerve,
diminishing the danger of injury to the nerve during surgery
[18].
A further indication of temporal bone digital volume tomography is the
analysis of middle ear implants. On digital volume tomography, the loops of
all three prostheses, attached to the long process of the incus, could be
recognized. The only unidentified bone part, obscured by the surrounding
metal, is located within the prosthetic loop. The titanium and gold pistons
are demonstrated up to the level of the vestibule. Artifacts impacting the in
situ prosthesis were not encountered. Even with digital volume tomography, the
proper evaluation of the position of the platinum wire Teflon piston is
difficult. The prosthesis consists of a platinum wire that is shown by
radiologic means. The Teflon mantle, enveloping the inferior aspect of the
prosthesis, is not demonstrated radiologically. This part is 0.5 mm longer
than the platinum wire, which means that with a 5.5-mm prosthesis, only 5.0 mm
of the wire is shown radiologically. As a consequence, the in situ prosthesis
in the temporal bone specimen is not completely visualized on digital volume
tomography. The recognizable platinum wire, of which the loop is fixed to the
long process of the incus, can be traced to the level of the footplate. The
wire terminates about 0.2 mm above the footplate. Realistically, the
prosthesis is 0.5 mm longer and the distance is added to the recognizable
part. Therefore, the prosthesis can be judged, with a certain degree of
certainty, as in situ positioned.
These observations, like the examinations of the gold and titanium pistons,
are important in the differential diagnosis of a conductive hearing loss after
stapes surgery. It is important to differentiate the possible dislocation of a
stapes prosthesis from fixation of the incus and malleus or an incus necrosis.
These are important findings because the subsequent operative strategy of
revision surgery, associated with possible complications, depends on it. Such
visualizations are significant not only for stapes surgery but also for
tympanoplasty.
We used three different implants for an ossiculoplasty to be evaluated and
visualized on digital volume tomography. In addition to an autologous incus
prosthesis, we applied a gold partial prosthesis with a length of 2.0 mm and a
4.5-mm titanium Total Implant. The autologous incus implant was demonstrated
to be in direct contact with the malleus handle and stapes head, without
adherence to surrounding structures, on digital volume tomography. The
radiologic results were confirmed by microscopic examination of the temporal
bone specimens. The partial gold implant, recognized between the malleus
handle and stapes head, was well illustrated. The structure of the partial
implant was visualized artifact-free on digital volume tomography as was the
titanium Total Implant, which was demonstrated between malleus handle and
stapes footplate.
With digital volume tomography, autologous and alloplastic middle ear
implants are well visualized with respect to their position and direction. By
positioning a total implant at the edge of the footplate, a slight dislocation
of a middle ear implant was simulated. Subsequently, digital volume tomography
showed the dislocation of the prosthesis with a difference of 1 mm in location
on the stapes footplate compared with its original normal position. In such a
situation, a dislocated implant with adherence to the facial nerve canal would
result in conductive hearing loss. These findings illustrate the value of
digital volume tomography for the diagnosis of a persistent conductive hearing
loss after ossiculoplasty.
Similar to the visualization of the middle ear implant position, cochlear
implants can be evaluated with digital volume tomography. By using different
planes, the position of the complete electrode within the cochlea can be
visualized despite its snail-shaped configuration. In addition to evaluation
of the intracochlear position of electrodes
[13,
19], individual electrodes and
their spatial positions in relation to the modiolus are illustrated. A more
perimodulary electrode location extends the dynamic range and reduces the
stimulation threshold of the implant as the spiral ganglion cells are located
in the canal of Rosenthal
[20].
Digital volume tomography in dental surgery has value in the diagnosis of
small bone lesions. Experience with digital volume tomography in our temporal
bone study has shown the value of visualizing and analyzing small pathologic
changes in the lateral skull base. Digital volume tomography after
ossiculoplasty with autologous and alloplastic middle ear implants showed the
position of the prosthesis. The normal position of implants was differentiated
from prosthesis dislocation, allowing the establishment of the cause of
conductive hearing loss. Furthermore, the normal position of a cochlear
implant was actually analyzed, allowing visualization of the position of
individual electrodes in relation to the modiolus.
We believe our experience will extend the previous radiologic diagnosis of
temporal bone abnormalities by digital volume tomography. The high precision
of digital volume tomography will contribute valuable information in the
preoperative planning for and potentially the prevention of intraoperative
complications. Digital volume tomography combines the advantages of a small
radiation dosage and high resolution with significant cost savings.
References
- Swartz JD, Harnsberger HR. Imaging of the temporal
bone, 3rd ed. New York, NY: Thieme, 1998
- Zinreich SJ, Mattox DE, Kennedy DW, et al. 3-D CT for cranial
facial and laryngeal surgery. Laryngoscope1988; 98:1212
-1219[Medline]
- Reisser C, Schubert O, Weidauer H. 3-Dimensional imaging of
temporal bone structures using spiral CT: initial results in normal temporal
bone anatomy [in German]. HNO 1995;43
: 596-600[Medline]
- Fuhrmann A, Schulze D, Rother U, Vesper M. Digital transversal
slice imaging in dental-maxillofacial radiology: from pantomography to digital
volume tomography. Int J Comput Dent2003; 6:129
-140[Medline]
- Ziegler CM, Woertche R, Brief J, Hassfeld S. Clinical indications
for digital volume tomography in oral and maxillofacial surgery.
Dentomaxillofac Radiol 2002;31
: 126-130[Abstract]
- Cavalcanti MG, Ruprecht A, Vannier MW. 3D volume rendering using
multislice CT for dental implants. Dentomaxillofac
Radiol 2002; 31:218
-223[Abstract]
- Kim KD, Ruprecht A, Jeon KJ, Park CS. Personal computer-based
three-dimensional computed tomographic images of the impacted teeth for
evaluating supernumerary or ectopically impacted teeth. Angle
Orthod 2003; 73:614
-621[Medline]
- Yoshinori A, Kazuya H, Kazuo I, Koji S. Practical model
"3DX" of limited cone-beam X-ray CT for dental use.
International Congress Series 2001;1230
: 712-718
- Manolidis S, Williamson B, Chan LL, Taber KH, Hayman LA. Use of
reconstructed, non-orthogonal plane, high-resolution computed tomography of
the temporal bone in the planning of temporal bone surgery. ORL J
Otorhinolaryngol Relat Spec 2003;65
: 71-75[Medline]
- Sorensen MS, Dobrzeniecki AB, Larsen P, Frisch T, Sporring J,
Darvann TA. The visible ear: a digital image library of the temporal bone.
ORL J Otorhinolaryngol Relat Spec 2002;64
: 378-381[Medline]
- Muren C, Ytterberg C. Computed tomography of temporal bone
specimens: a test of the resolution capability. Acta Radiol Diagn
(Stockh) 1986; 27:645
-651[Medline]
- Xu J, Stevenson AW, Gao D, et al. The role of radiographic
phase-contrast imaging in the development of intracochlear electrode arrays.
Otol Neurotol 2001;22
: 862-868[Medline]
- Xu J, Xu SA, Cohen LT, Clark GM. Cochlear view: postoperative
radiography for cochlear implantation. Am J Otol2000; 21:49
-56[Medline]
- Arai Y, Tammisalo E, Iwai K, Hashimoto K, Shinoda K. Development of
a compact computed tomographic apparatus for dental use.
Dentomaxillofac Radiol 1999;28
: 245-248[Abstract]
- Arai Y, Hasimoto K, Iwai K, Shinoda K. Fundamental efficiency of
limited cone-beam x-ray CT (3DX multi image micro CT) for practical use.
Jpn Dental Radiol 2000;40
: 145-154
- Honda K, Arai Y, Shinoda K. Fundamental efficiency of new-style
limited-cone-beam CT (3DX): comparison with helical CT. Jpn J
Tomogr 2001; 27:193
-198
- Iwai K, Arai Y, Hashimoto K, Nishizawa K. Estimation of effective
dose from limited cone beam X-ray CT examination. Jpn Dent
Radiol 2001; 40:251
-259
- Valavanis A, Kubik S, Schubiger O. High-resolution CT of the normal
and abnormal fallopian canal. Am J Neuroradiol1983; 4:748
-751[Abstract]
- Lawson JT, Cranley K, Toner JG. Digital imaging: a valuable
technique for the postoperative assessment of cochlear implantation.
Eur Radiol 1998;8
: 951-954[Medline]
- Shepherd RK, Hatsushika S, Clark GM. Electrical stimulation of the
auditory nerve: the effect of electrode position on neural excitation.
Hear Res 1993; 66:108
-1203[CrossRef][Medline]

CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?