|
|
||||||||
Perspective |
1 All authors: Department of Radiology, Division of Abdominal Imaging and Interventional Radiology, Harvard Medical School and Massachusetts General Hospital, 55 Fruit St., White 270, Boston, MA 02114.
Received March 24, 2006; accepted after revision June 28, 2006.
Address correspondence to R. N. Uppot
(ruppot{at}partners.org).
OBJECTIVE. The purpose of this article is to discuss the impact of obesity on medical imaging and provide some solutions that are currently available to tackle the challenges of imaging obese patients.
CONCLUSION. Increasingly, radiologists are asked to image morbidly obese patients. The challenges facing radiology departments include difficulties in transporting patients to the department, inability to accommodate large patients on currently designed imaging equipment, and difficulties in acquiring desired image quality.
Keywords: abdominal imaging body imaging interventional radiology obesity radiology practice technologists
The prevalence of obesity is increasing in the United States. According to the Centers for Disease Control and Prevention (CDC), approximately 64% of Americans are overweight, obese, or morbidly obese [1]. Obesity directly impacts health, with an increased incidence of diabetes, heart disease, and certain types of cancer [2].
Increasingly, however, hospitals are faced with the indirect impact of obesity. As the prevalence of obesity increases in the United States, there is a growing need for larger hospital beds, larger wheelchairs, and larger operating room tables [3].
Radiology departments are also facing the impact of obesity [4]. The increasing prevalence of obesity in the United States and the growing popularity of bariatric surgery have led to increased use of medical imaging of obese patients. Currently designed imaging technology has limited ability to accommodate and provide the desired quality of images in obese patients.
Obesity is defined by body mass index (BMI), which takes into account both height and weight (Table 1). Despite the importance of BMI in the clinical classification of obesity, for radiologists two more appropriate measurements are weight and body diameter.
|
In the appropriate management of obese patients, radiology departments often have to address several issues: What is the best technique to image the obese patient for a given indication? How should the obese patient be scheduled and transported to the radiology department? How will the patient fit on the imaging equipment available? How should equipment settings be modified to optimize image quality for the obese patient? This review will describe the difficulties in imaging obese patients and discuss imaging strategies to overcome these difficulties.
Imaging Techniques for Obese Patients
More than the clinical indications, a patient's weight and body diameter are important factors in deciding which imaging technique to use. Although a brain MRI may be the most appropriate technique for evaluating an acute stroke, if the patient cannot fit on the MR scanner, it is of no practical use. Therefore, before scheduling an obese patient for a diagnostic imaging procedure, it is important to know the patient's weight and diameter to assess whether the patient can fit on the imaging equipment. Table weight limits and aperture diameters for fluoroscopy differ from those for CT and for MRI.
Currently, industry standards exist for table weight limits and aperture diameters for each of the imaging techniques (Table 2). In increasing order of cross-sectional diameter, according to current industry standards, the imaging techniques are fluoroscopy, CT, cylindric bore MRI, and vertical bore open MRI.
|
|
Aperture Diameter
Patients may meet the weight limit of a table but may exceed the gantry or
bore diameter because of their girth. Typically, the industry-standard
aperture in fluoroscopy is 45 cm; the gantry diameter in CT, including MDCT,
is 70 cm; and the bore aperture in MRI is 60 cm. Although the aperture
diameters are accurate in the horizontal plane on CT and MRI, they do not
account for the table thickness entering the gantry or bore and therefore
overestimate the vertical distance (anteroposterior distance in a supine
patient). Typically, in the vertical plane, 15-18 cm must be subtracted from
the gantry or bore diameter to account for the table thickness
(Fig. 1).
Radiologists and technologists should be aware of weight and aperture limits of all imaging equipment at their diagnostic imaging facility. Weight and aperture limit data should be posted and be made easily available within the department. Despite these industry standard limits, some imaging vendors are now recognizing the issue of obesity and have increased the table weights and aperture dimensions of their newest imaging equipment (Table 3).
|
Transportation to the Radiology Department
Once the imaging technique is selected, the next step is appropriate scheduling and transporting of the patient. Obese patients who exceed the weight limits of the transportation equipment can complicate transportation within a hospital. Typically, in most institutions, the resources (larger wheelchairs and stretchers) suitable for transporting larger patients are limited. Delay in transporting scheduled patients to the department increases the downtime of imaging equipment, hampers throughput, and increases the wait time for other patients.
Proposed solutions to address this issue include coordinating with the transport department to schedule obese patients only when the larger wheelchairs and stretchers are available, purchasing more wheelchairs or stretchers that can accommodate larger patients if the institution is seeing a growing percentage of obese patients, and acquiring portable radiography or sonography equipment for patients who cannot be transported.
Can the Patient Fit?
In addition to the discussed limitations of table weight and aperture diameter for fluoroscopy, CT, and MRI, other techniques have constraint. For radiography, obese patients' surface areas may be too large to fit on a 14 x 17 inch (35.56 x 43.18 cm) cassette. Magnification may also cone off the area of interest even if it overlies the cassette. On sonography, patient obesity may limit the ability of technologists to appropriately position patients for quality images.
|
Diagnostic Quality Images
After choosing the appropriate technique, transporting the patient, and fitting the patient, the final hurdle in imaging obese patients is the ability to obtain diagnostic quality images. The difficulties and solutions for imaging obese patients are specific for each imaging technique.
Radiography
Radiographs are limited by X-ray beam attenuation that results in lower
image contrast. Also, the increased body thickness through which the X-ray
beam must travel results in increased exposure time and introduces motion
artifact. The typical setting to obtain a chest radiograph is a kVp of 90-95
and mAs of 2-2.5. However, in obese patients these settings can result in
inadequate penetration of the X-ray beams through the patient's body, along
with more background scatter (Fig.
3A).
|
|
|
|
Solutions to improving image quality in obese patients include using the lowest frequency transducer available (2 MHz), positioning the transducer to image the organ of interest within the range of the focal length of the transducer, and examining the patient's previous imaging (CT or MRI) to determine the thickness of subcutaneous fat (Fig. 5A, 5B).
|
|
Increased noiseIncreased noise is a result of inadequate beam penetration (Fig. 6A). Solutions to decrease noise involve increasing the kVp to 140 and increasing the effective mAs. Two ways that this may be accomplished include decreasing the gantry rotation speed from one rotation in 0.5 second up to one rotation in 1 second to increase the effective mAs and, on MDCT scanners with automated tube current modulation, changing the scanner settings from "fixed mAs" to "automatic mAs," which will allow the scanner to determine the amount of mAs to deliver per body section. Although both of these solutions improve the image quality (Fig. 6B), they increase the radiation dose to the patient.
|
|
Limited field of viewStandard 70-cm gantry CT instruments have a field of view of 55-65 cm. Increases in gantry diameter on the newer, larger gantry CT scanners also permit increases in the field of view to accommodate larger patients. In areas where the patient's body is larger than the field of view but smaller then the gantry diameter, beam-hardening artifact can limit evaluation of internal organs. One solution is to recognize the artifact during image acquisition and adjust the patient position so that the area of interest does not exceed the field of view.
Dangers of croppingAfter a CT study is acquired, technologists may crop the images to focus on internal organ structures at the expense of subcutaneous tissues. Cropping subcutaneous fat can result in the loss of valuable information, especially in patients undergoing evaluation for malignancies or nonspecific abdominal pain for which the abnormality may be within the subcutaneous tissues (Fig. 7A, 7B, 7C).
|
|
|
CT benefits of fatObesity, however, is not always detrimental to CT image quality. Patients who have predominantly intraperitoneal or retroperitoneal fat have improved visualization of internal organ structures compared with patients with a paucity of intraperitoneal fat because of the better delineation of internal organ structures by the fat (Fig. 8A, 8B).
|
|
1.5 T) cannot accommodate patients weighing 350 lb (159 kg) or more or
who are greater than 60 cm in diameter. On the other hand, typical vertical
field open MRI systems can accommodate patients up to 550 lb (250 kg) and
offer a range of vertical apertures from 40 to 55 cm but have a lower SNR and
weaker gradients [5]. In addition to table weight limits and bore diameter limits, another design factor in MRI that is relevant in imaging obese patients is the bore length. Long bore lengths in cylindric MR scanners, typically 170 cm, may be uncomfortable for obese patients who can become claustrophobic if the entire torso is in the bore. Shorter bore length magnets, 125 cm, on the other hand, may be better tolerated by obese patients because they allow the head and neck area to be outside the bore during scanning. Manufacturers of newer MR scanners recently marketed for the obese population try to address this issue by offering larger bore diameters, shorter bore lengths, and higher table weight limits in a 1.5-T system.
In imaging obese patients, several technical limitations need to be accounted for in MR scanning, including radiofrequency penetration and gradient strengths, limited field of view, scanning time, and radiofrequency energy deposition of the skin.
Radiofrequency penetration and gradient strengthsThe disadvantage of low-gradientstrength MR scanners is the low SNR. Increasing gradient strength will not allow improved radiofrequency depth penetration in obese patients but can increase SNR and spectral resolution. Larger body habitus also introduces noise and will therefore decrease the contrast- to-noise ratio (CNR). For receiver coils, the increased distance of inner organs from the coils in obese patients also affects the SNR.
Limited field of viewLarger patients occupy a larger field of view. The maximum field of view for 1.5-T cylindric bore MR scanners is around 40-50 cm. The maximum field of view for vertical field open systems ranges from 35 to 40 cm. Although a larger field of view is advantageous in imaging obese patients, there is an inverse relationship between field of view and image resolution. Larger fields of view can decrease image resolution. Therefore, in imaging obese patients, the smallest possible field of view is used to image the organ of interest without inducing wraparound artifacts (Fig. 9). Some manufacturers have a "no phase wrap" option to prevent wraparound artifact.
|
Scanning timeObesity can affect MRI scanning times. Scanning times are increased because of the larger cross-sectional area and longer craniocaudal dimensions, which require more slices. Increased scanning times can also lead to patient motion, with associated motion artifacts.
Other factorsOther techniques that can aid in improving image quality include using a body coil rather than a phased-array multicoil and using saturation bands to decrease noise from subcutaneous fat. Another factor to consider in MRI of obese patients is the deposition of radiofrequency energy on the skin where it abuts the gantry. At our institution, technologists pad areas of the patient that abut the bore to prevent minor burns.
Nuclear Medicine
In nuclear medicine, obesity degrades image quality by the scatter of
photons within the soft tissues, decreasing the SNR
(Fig. 10). Also, because the
administration of radioisotopes is based on weight, obese patients may exceed
the maximum allowable dose and may not be able to receive the proportionate
dose of radionuclide for their body weight. Solutions in nuclear medicine
imaging of obese patients include using the maximum allowable dose and imaging
for a longer time to maximize counts.
|
Interventional Radiology
Obese patients can present scheduling, positioning, and technical challenges to the interventional radiologist. Obese patients may also require high doses of weight-based sedative medications, which may put them at risk for respiratory depression.
Individuals responsible for scheduling interventional procedures should routinely ask for the patient's weight. From our experience, all patients greater than 250 lb (159 kg) should be flagged. In this group of patients, consultation with the referring clinical service and review of prior images by the interventional radiologist before scheduling the procedure can ensure the use of the most appropriate imaging technique and availability of equipment appropriate for the procedure.
Proper patient positioning can also be a problem in obese patients. Several ancillary staff members need to always be available to help move the patient from the stretcher to the procedure table and to help position the patient if necessary. The use of pillows and sandbags is important to secure the patient position before the start of the procedure.
Meticulous planning is important before starting the procedure. Prior imaging can help determine the depth of fat tissue and the most direct approach to the organ of interest. Instruments of appropriate length must be chosen before starting the procedure. Technically, obese patients are more challenging. The accuracy in targeting lesions decreases the deeper the lesion that is to be biopsied or drained.
Sedation of the obese patient can also present challenges. If patients do not tolerate the administered doses of sedatives, the use of a mix of different active sedatives or the assistance of an anesthesiologist may be necessary. Also, if a patient's airway has the potential to be compromised or difficult to access because of body habitus, the patient may not be a candidate for conscious sedation, making general anesthesia mandatory. A few obese patients may require a surgical approach for diagnosis or treatment if suitable imaging guidance cannot be provided.
Summary
The prevalence of obesity is increasing in the United States. Standard, widely deployed imaging equipment is approaching its limits in the ability to image obese patients. Radiologists and technologists need to be aware of the limitations of imaging equipment and the equipment adjustments that can be made to improve image quality in obese patients. Manufacturers of imaging equipment are making design changes to accommodate larger patients and will begin marketing such equipment over the next several years.
References
This article has been cited by other articles:
![]() |
B. Siewert, J. Sosna, A. McNamara, V. Raptopoulos, and J. B. Kruskal Quality Initiatives: Missed Lesions at Abdominal Oncologic CT: Lessons Learned from Quality Assurance RadioGraphics, May 1, 2008; 28(3): 623 - 638. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |