DOI:10.2214/AJR.06.1616
AJR 2007; 188:1447-1474
© American Roentgen Ray Society
ACR Guidance Document for Safe MR Practices: 2007
Emanuel Kanal1,
A. James Barkovich2,
Charlotte Bell3,
James P. Borgstede4,
William G. Bradley, Jr.5,
Jerry W. Froelich6,
Tobias Gilk7,
J. Rod Gimbel8,
John Gosbee9,
Ellisa Kuhni-Kaminski1,
James W. Lester, Jr.10,
John Nyenhuis11,
Yoav Parag1,
Daniel J. Schaefer12,
Elizabeth A. Sebek-Scoumis1,
Jeffrey Weinreb13,
Loren A. Zaremba14,
Pamela Wilcox15,
Leonard Lucey15,
Nancy Sass15 the ACR Blue Ribbon Panel on MR Safety
1 Department of Radiology, University of Pittsburgh Medical Center, Pittsburgh,
PA.
2 Neuroradiology Section, University of California San Francisco, San Francisco,
CA.
3 American Society of Anesthesiologists and Department of Anesthesiology, New
York University School of Medicine, New York, NY.
4 Colorado Springs Radiologists, Colorado Springs, CO.
5 Professor and Chairman, Department of Radiology, University of California San
Diego, San Diego, CA.
6 Department of Radiology, University of Minnesota, Minneapolis, MN.
7 MRI-Planning, Kansas City, MO.
8 East Tennessee Heart Consultants, Lenoir City, TN.
9 University of Michigan Health System and Red Forest Consulting, Ann Arbor,
MI.
10 Chapel Hill, NC.
11 Department of Electrical and Computer Engineering, Purdue University, West
Lafayette, IN.
12 MR Systems Engineering, GE Healthcare, Milwaukee, WI.
13 Yale University School of Medicine, New Haven, CT.
14 U.S. Food and Drug Administration, Rockville, MD.
15 American College of Radiology, 1891 Preston White Dr., Reston, VA 20191.
Received December 8, 2006;
accepted after revision December 18, 2006.
E. Kanal is a consultant for, is a member of the speakers bureau of, and
provides research support for Bracco Diagnostics and GE Healthcare; is a
member of the speakers bureau of and provides research support for Siemens
Medical Solutions; and provides research support for Berlex and Medtronic.
T. Gilk is a consultant for Mednovus, Inc.
J. R. Gimbel provides research support for St. Jude Medical, Medtronic, and
Biotronik.
J. Nyenhuis is a consultant for and provides research support to
Medtronic.
J. Weinreb is a consultant and member of the speakers bureau for GE
Healthcare.
Address correspondence to N. Sass.
FOR YOUR INFORMATION
See the accompanying CME/SAM article in this month's issue of AJR
Integrative Imaging included with your AJR.
FOR YOUR INFORMATION
The reader's attention is directed to the commentary on this article, which
appears on page 1446.
Keywords: MR contrast agents MRI safety
Introduction
There are potential risks in the MR environment, not only for the patient
[1,
2] but also for the
accompanying family members, attending health care professionals, and others
who find themselves only occasionally or rarely in the magnetic fields of MR
scanners, such as security or housekeeping personnel, firefighters, police,
etc.
[36].
There have been reports in the medical literature and print media detailing
magnetic resonance imaging (MRI) adverse incidents involving patients,
equipment, and personnel that spotlighted the need for a safety review by an
expert panel. To this end, the American College of Radiology (ACR) originally
formed the Blue Ribbon Panel on MR Safety. First constituted in 2001, the
panel was charged with reviewing existing MR safe practices and guidelines
[59]
and issuing new ones as appropriate for MR examinations. Published initially
in 2002 [3], the ACR MR Safe
Practice Guidelines established de facto industry standards for safe and
responsible practices in clinical and research MR environments. These were
subsequently reviewed and updated in May 2004
[4]. After reviewing
substantial feedback from the field and installed bases, as well as changes
that had transpired throughout the MR industry since the publication of the
2004 version of this document, the panel extensively reviewed, modified, and
updated the entire document in 20062007.
The present panel consists of the following members: A. James Barkovich,
MD; Charlotte Bell, MD (American Society of Anesthesiologists); James P.
Borgstede, MD, FACR; William G. Bradley, MD, PhD, FACR; Jerry W. Froelich, MD;
Tobias Gilk, architect; J. Rod Gimbel, MD, FACC, cardiologist; John Gosbee,
MD, MS; Ellisa Kuhni-Kaminski, RT (R)(MR); Emanuel Kanal, MD, FACR, FISMRM
(chair); James W. Lester, MD; John Nyenhuis, PhD; Yoav Parag, MD; Daniel Joe
Schaefer, PhD, engineer; Elizabeth A. Sebek-Scoumis, RN, BSN, CRN; Jeffrey
Weinreb, MD; Loren A. Zaremba, PhD, FDA; Pamela Wilcox, RN, MBA (ACR staff);
Leonard Lucey, JD, LLM (ACR staff); and Nancy Sass, RT (R)(MR)(CT) (ACR
staff). The following represents the most recently modified and updated
version of the combined prior two reports
[3,
4] issued by the American
College of Radiology Blue Ribbon Panel on MR Safety, chaired by Emanuel Kanal,
MD, FACR. It is important to note that nothing that appears herein is the
result of a "majority vote" of the members of this panel. As with
each prior publication of these ACR MR Safe Practice Guidelines, the entire
document, from introduction to the markedly expanded appendices, represents
the unanimous consensus of each and every member of this Safety Committee and
the various areas of expertise that they represent. This includes
representation from fields and backgrounds as diverse as MR physicists,
research/academic radiologists, private practice radiologists, MR safety
experts, patient safety experts/researchers, MR technologists, MR nursing,
National Electrical Manufacturers Association, the U.S. Food and Drug
Administration (FDA), the American Society of Anesthesiologists, legal
counsel, and others. Lay personnel, physicians, PhDs, department chairs and
house-staff/residents, government employees and private practitioners,
doctors, nurses, technologists, radiologists, anesthesiologists,
cardiologists, attorneysthese are all represented on this Committee. It
was felt that achieving unanimity for these guidelines was critical in order
to demonstrate to all that these guidelines are not only appropriate from a
scientific point of view, but are reasonably applicable in the real world in
which we all must live, with all its patient care, financial, and throughput
pressures and considerations.
The following MR safe practice guidelines document is intended to be used
as a template for MR facilities to follow in the development of an MR safety
program. These guidelines were developed to help guide MR practitioners
regarding these issues and to provide a basis for them to develop and
implement their own MR policies and practices. It is intended that these MR
safe practice guidelines (and the policies and procedures to which they give
rise) be reviewed and updated on a regular basis as the field of MR safety
continues to evolve.
The principles behind these MR safe practice guidelines are specifically
intended to apply not only to diagnostic settings but also to patient,
research subject, and health care personnel safety for all MRI settings,
including those designed for clinical diagnostic imaging, research,
interventional, and intraoperative MR applications.
With the increasing advent and use of 3.0-Tesla and higher strength
magnets, users need to recognize that one should never assume MR compatibility
or safety information about a device if it is not clearly documented in
writing. Decisions based on published MR safety and compatibility claims
should recognize that all such claims apply only to specifically tested
conditions, such as static magnetic field strengths, static gradient magnetic
field strengths and spatial distributions, and the strengths and rates of
change of gradient and radiofrequency (RF) magnetic fields.
Finally, there are many issues that impact MR safety that should be
considered during site planning for a given MR installation. These have
historically not been dealt with in the prior versions of the ACR MR Safe
Practice Guidelines. For the first time, we include in this article, as
separate appendices, sections that address such issues as well, including
cryogen emergency vent locations and pathways, 5-gauss lines, siting
considerations, patient access pathways, etc. Yet despite their appearance
herein, these issues, and many others, should be reviewed with those
experienced in MR site planning and familiar with the patient safety and
patient flow considerations prior to committing to construction of a specific
site design. In this regard, enlisting the assistance of an architectural firm
experienced in this area, and doing so early in the design stages of the
planning process, may prove most valuable.
It remains the intent of the ACR that these MR Safe Practice Guidelines
will prove helpful as the field of MRI continues to evolve and mature,
providing MR services that are among the most powerful, yet safest, of all
diagnostic procedures to be developed in the history of modern medicine.
ACR Guidance Document for Safe MR Practices: 2007
A. Establish, Implement, and Maintain Current MR Safety Policies and Procedures
- All clinical and research MR sites, irrespective of magnet format or field
strength, including installations for diagnostic, research, interventional,
and/or surgical applications, should maintain MR safety policies.
- These policies and procedures should also be reviewed concurrently with the
introduction of any significant changes in safety parameters of the MR
environment of the site (e.g., adding faster or stronger gradient capabilities
or higher RF duty cycle studies) and updated as needed. In this review
process, national and international standards and recommendations should be
taken into consideration prior to establishing local guidelines, policies, and
procedures.
- Each site will name an MR medical director whose responsibilities will
include ensuring that MR safe practice guidelines are established and
maintained as current and appropriate for the site. It is the responsibility
of the site's administration to ensure that the policies and procedures that
result from these MR safe practice guidelines are implemented and adhered to
at all times by all of the site's personnel.
- Procedures should be in place to ensure that any and all adverse events, MR
safety incidents, or "near incidents" that occur in the MR site
are reported to the medical director in a timely fashion (e.g., within 24
hours or 1 business day of their occurrence) and used in continuous quality
improvement efforts. It should be stressed that the Food and Drug
Administration states that it is incumbent upon the sites to also report
adverse events and incidents to them via their MedWatch program. The ACR
supports this requirement and feels that it is in the ultimate best interest
of all MR practitioners to create and maintain this consolidated database of
such events to help us all learn about them and how to better avoid them in
the future [10,
11].
B. Static Magnetic Field Issues: Site Access Restriction
1. Zoning
The MR site is conceptually divided into four Zones (see
Figure 1 and Appendix 1):

View larger version (20K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 1 Idealized sample floor plan illustrates site access
restriction considerations. Other MR potential safety issues, such as magnet
site planning related to fringe magnetic field considerations, are not meant
to be include herein. See Appendix 1 for personnel and zone definitions.
NoteIn any zone of the facility, there should be compliance with Health
Insurance Portability and Accountability Act (HIPAA) regulations in regard to
privacy of patient information. However, in Zone III, there should be a
privacy barrier so that unauthorized persons cannot view control panels.
|
|
- Zone I: This region includes all areas that are freely accessible to the
general public. This area is typically outside the MR environment itself and
is the area through which patients, health care personnel, and other employees
of the MR site access the MR environment.
- Zone II: This area is the interface between the publicly accessible,
uncontrolled Zone I and the strictly controlled Zones III and IV. Typically,
patients are greeted in Zone II and are not free to move throughout Zone II at
will, but are rather under the supervision of MR personnel (see section B.2.b,
below). It is in Zone II that the answers to MR screening questions, patient
histories, medical insurance questions, etc. are typically obtained.
- Zone III: This area is the region in which free access by unscreened non-MR
personnel or ferromagnetic objects or equipment can result in serious injury
or death as a result of interactions between the individuals or equipment and
the MR scanner's particular environment. These interactions include, but are
not limited to, those involving the MR scanner's static and time-varying
magnetic fields. All access to Zone III is to be strictly restricted, with
access to regions within it (including Zone IV, see below) controlled by, and
entirely under the supervision of, MR personnel (see section B.2.b, below).
Specifically identified MR personnel (typically, but not necessarily only, the
MR technologists) are to be charged with ensuring that this MR safe practice
guideline is strictly adhered to for the safety of the patients and other
non-MR personnel, the health care personnel, and the equipment itself. This
function of the MR personnel is directly under the authority and
responsibility of the MR medical director or the level 2 MR
personneldesignated (see section B.2.b, below) physician of the day for
the MR site.
Zone III regions should be physically restricted from general public access
by, for example, key locks, passkey locking systems, or any other reliable,
physically restricting method that can differentiate between MR personnel and
non-MR personnel. The use of combination locks is discouraged as combinations
often become more widely distributed than initially intended, resulting in
site restriction violations being more likely with these devices. Only MR
personnel shall be provided free access, such as the access keys or passkeys,
to Zone III.
There should be no exceptions to this guideline. Specifically,
this includes hospital or site administration, physician, security, and other
non-MR personnel (see section B.2.c, below). Non-MR personnel are not to be
provided with independent Zone III access until such time as they undergo the
proper education and training to become MR personnel themselves. Zone III, or
at the very least the area within it wherein the static magnetic field's
strength exceeds 5 gauss, should be demarcated and clearly marked as being
potentially hazardous.
Because magnetic fields are three-dimensional volumes, Zone III controlled
access areas may project through floors and ceilings of MRI suites, imposing
magnetic field hazards on persons on floors other than that of the MR scanner.
Zones of magnetic field hazard should be clearly delineated, even in typically
nonoccupied areas such as rooftops or storage rooms, and access to these Zone
III areas should be similarly restricted from non-MR personnel as they would
be inside any other Zone III region associated with the MRI suite. For this
reason, magnetic field strength plots for all MRI systems should be analyzed
in vertical section as well as in horizontal plan, identifying areas above or
below, in addition to areas on the same level, where persons may be at risk of
interactions with the magnetic field.
- Zone IV: This area is synonymous with the MR scanner magnet room itself,
that is, the physical confines of the room within which the MR scanner is
located. Zone IV, by definition, will always be located within Zone III, as it
is the MR magnet and its associated magnetic field that generates the
existence of Zone III. Zone IV should also be demarcated and clearly marked as
being potentially hazardous due to the presence of very strong magnetic
fields. As part of the Zone IV site restriction, all MR installations should
provide for direct visual observation by level 2 personnel to access pathways
into Zone IV. By means of illustration only, the MR technologists would be
able to directly observe and control, via line of sight or via video monitors,
the entrances or access corridors to Zone IV from their normal positions when
stationed at their desks in the scan control room.
Zone IV should be clearly marked with a red light and lighted sign stating,
"The Magnet is On." Except for resistive systems, this light and
sign should be illuminated at all times and should be provided with a backup
energy source to continue to remain illuminated for at least 24 hours in the
event of a loss of power to the site.
In case of cardiac or respiratory arrest or other medical emergency within
Zone IV for which emergent medical intervention or resuscitation is required,
appropriately trained and certified MR personnel should immediately initiate
basic life support or CPR as required by the situation while the
patient is being emergently removed from Zone IV to a predetermined,
magnetically safe location. All priorities should be focused on
stabilizing (e.g., basic life support with cardiac compressions and manual
ventilation) and then evacuating the patient as rapidly and safely as possible
from the magnetic environment that might restrict safe resuscitative
efforts.
Further, for logistical safety reasons, the patient should always be moved
from Zone IV to the prospectively identified location where full resuscitative
efforts are to continue. (See Appendix 2.)
Quenching the magnet (for superconducting systems only) is not routinely
advised for cardiac or respiratory arrest or other medical emergency, since
quenching the magnet and having the magnetic field dissipate could easily take
more than a minute. Furthermore, as quenching a magnet can theoretically be
hazardous, ideally one should evacuate the magnet room, when possible, for an
intentional quench. One should rather use that time wisely to initiate life
support measures while removing the patient from Zone IV to a location where
the strength of the magnetic field is insufficient to be a medical concern.
Zones III and IV site access restriction must be maintained during
resuscitation and other emergent situations for the protection of all
involved.
2. MR personnel and non-MR personnel
- All individuals working within at least Zone III of the MR environment
should be documented as having successfully completed at least one of the MR
safety live lectures or prerecorded presentations approved by the MR medical
director. Attendance should be repeated at least annually, and appropriate
documentation should be provided to confirm these ongoing educational efforts.
These individuals shall be referred to henceforth as MR personnel.
- There are two levels of MR personnel:
- Level 1 MR personnel: Those who have passed minimal safety educational
efforts to ensure their own safety as they work within Zone III will be
referred to henceforth as level 1 MR personnel.
- Level 2 MR personnel: Those who have been more extensively trained and
educated in the broader aspects of MR safety issues, including, for example,
issues related to the potential for thermal loading or burns and direct
neuromuscular excitation from rapidly changing gradients, will be referred to
henceforth as level 2 MR personnel. It is the responsibility of the MR medical
director not only to identify the necessary training, but also to identify
those individuals who qualify as level 2 MR personnel. It is understood that
the medical director will have the necessary education and experience in MR
safety to qualify as level 2 MR personnel. (See Appendix 1.)
- All those not having successfully complied with this MR safety instruction
guideline shall be referred to henceforth as non-MR personnel. Specifically,
non-MR personnel will be the terminology used to refer to any individual or
group who has not within the previous 12 months undergone the designated
formal training in MR safety issues defined by the MR safety director of that
installation.
3. Patient and non-MR personnel screening
- All non-MR personnel wishing to enter Zone III must first pass an
MR safety screening process. Only MR personnel are authorized to perform an MR
safety screen before permitting non-MR personnel into Zone III.
- The screening process and screening forms for patients, non-MR personnel,
and MR personnel should be essentially identical. Specifically, one should
assume that non-MR personnel, health care practitioners, or MR personnel may
enter the bore of the MR imager during the MR imaging process.
Examples of this might include when a pediatric patient cries for his
mother, who then leans into the bore, or when the anesthetist leans into the
bore to manually ventilate a patient in the event of a problem.
- Metal detectors
The usage in MR environments of conventional metal detectors which do not
differentiate between ferrous and nonferromagnetic materials is not
recommended. Reasons for this recommendation against conventional metal
detector usage include, among others:
- They have variedand variablesensitivity settings.
- The skills of the operators can vary.
- Today's conventional metal detectors cannot detect, for example, a 2
x 3 mm, potentially dangerous ferromagnetic metal fragment in the orbit
or near the spinal cord or heart.
- Today's conventional metal detectors do not differentiate between
ferromagnetic and nonferromagnetic metallic objects, implants, or foreign
bodies.
- Metal detectors should not be necessary for the detection of large metallic
objects, such as oxygen tanks on the gurney with the patients. These objects
are fully expected to be detectedand physically excludedduring
the routine patient screening process.
However, ferromagnetic detection systems are currently available that are
simple to operate, capable of detecting even very small ferromagnetic objects
external to the patient, and now, for the first time, differentiating between
ferromagnetic and nonferromagnetic materials. While the use of conventional
metal detectors is not recommended, the use of ferromagnetic detection
systems is recommended as an adjunct to thorough and conscientious
screening of persons and devices approaching Zone IV. It should be reiterated
that their use is in no way meant to replace a thorough screening practice,
which rather should be supplemented by their usage.
- Non-MR personnel should be accompanied by, or under the immediate
supervision of and in visual or verbal contact with, one specifically
identified level 2 MR person for the entirety of their duration within Zone
III or Zone IV restricted regions. However, it is acceptable to have non-MR
personnel in a changing room or restroom in Zone III without visual contact as
long as the personnel and the patient can communicate verbally with each
other.
Level 1 MR personnel are permitted unaccompanied access throughout Zones
III and IV. Level 1 MR personnel are also explicitly permitted to be
responsible for accompanying non-MR personnel into and throughout Zone III,
excluding Zone IV. However, level 1 MR personnel are not permitted to
directly admit, or be designated responsible for, non-MR personnel in Zone
IV.
In the event of a shift change, lunch break, etc., no level 2 MR personnel
shall relinquish their responsibility to supervise non-MR personnel still
within Zone III or Zone IV until such supervision has been formally
transferred to another of the site's level 2 MR personnel.
- Nonemergent patients should be MR safetyscreened on site by a
minimum of 2 separate individuals. At least one of these individuals should be
level 2 MR personnel. At least one of these 2 screenings should be performed
verbally or interactively.
Emergent patients and their accompanying non-MR personnel may be screened
only once, providing the screening individual is level 2 MR personnel. There
should be no exceptions to this.
- Any individual undergoing an MR procedure must remove all readily removable
metallic personal belongings and devices on or in them (e.g., watches,
jewelry, pagers, cell phones, body piercings [if removable], contraceptive
diaphragms, metallic drug delivery patches [see section I, below], cosmetics
containing metallic particles [such as eye make-up], and clothing items that
may contain metallic fasteners, hooks, zippers, loose metallic components, or
metallic threads). It is therefore advisable to require that the patients or
research subjects wear a site-supplied gown with no metal fasteners when
feasible.
- All patients and non-MR personnel with a history of potential ferromagnetic
foreign object penetration must undergo further investigation prior to being
permitted entrance to Zone III. Examples of acceptable methods of screening
include patient history, plain X-ray films, prior CT or MR studies of the
questioned anatomic area, or access to written documentation as to the type of
implant or foreign object that might be present. Once positive identification
has been made as to the type of implant or foreign object that is within a
patient, best-effort assessments should be made to identify the MR
compatibility or MR safety of the implant or object. Efforts at identification
might include written records of the results of formal testing of the implant
prior to implantation (preferred), product labeling regarding the implant or
object, and review of peer-reviewed publications regarding MR compatibility
and MR safety testing of the make, model, and type of the object. MR safety
testing would be of value only if the object or device had not been altered
since such testing results had been published.
All patients who have a history of orbit trauma by a potential
ferromagnetic foreign body for which they sought medical attention
are to have their orbits cleared either by plain X-ray orbit films (2 views)
[12,
13] or by a radiologist's
review and assessment of contiguous cut prior CT or MR images (obtained since
the suspected traumatic event), if available.
- Conscious, nonemergent patients and research and volunteer subjects are to
complete written MR safety screening questionnaires prior to their
introduction to Zone III. Family or guardians of nonresponsive patients or of
patients who cannot reliably provide their own medical histories are to
complete a written MR safety screening questionnaire prior to their
introduction to Zone III. These completed questionnaires are then to be
reviewed orally with the patient, guardian, or research subject in their
entirety prior to permitting the patient or research subject to be cleared
into Zone III.
The patient, guardian, or research subject as well as the screening MR
staff member must both sign the completed form. This form should then become
part of the patient's medical record. No empty responses will be
acceptedeach question must be answered with a
"yes" or "no" or specific further information must be
provided as requested. A sample pre-MR screening form is provided (see
Appendix 3). This is the minimum information to be obtained; more may be added
if the site so desires.
- Screening of the patient or non-MR personnel with, or suspected of having,
an intracranial aneurysm clip should be performed as per the separate MR safe
practice guideline addressing this particular topic (see section M,
below).
- Screening of patients for whom an MR examination is deemed clinically
indicated or necessary, but who are unconscious or unresponsive, who cannot
provide their own reliable histories regarding prior possible exposures to
surgery, trauma, or metallic foreign objects, and for whom such histories
cannot be reliably obtained from others:
- If no reliable patient metal exposure history can be obtained, and if the
requested MR examination cannot reasonably wait until a reliable history might
be obtained, it is recommended that such patients be physically examined by
level 2 MR personnel. All areas of scars or deformities that might be
anatomically indicative of an implant, such as on the chest or spine region,
and whose origins are unknown and which may have been caused by ferromagnetic
foreign bodies, implants, etc., should be subject to plain-film radiography
(if recently obtained plain films or CT or MR studies of such areas are not
already available). The investigation described above should be made to ensure
there are no potentially harmful embedded or implanted metallic foreign
objects or devices. All such patients should also undergo plain film imaging
of the skull or orbits and chest to exclude metallic foreign objects (if
recently obtained plain films or CT or MR studies of such areas are not
already available).
- Monitoring of patients in the MR scanner is sometimes necessary. The
potential for thermal injury from excessive RF power deposition exists.
Sedated, anesthetized, or unconscious patients may not be able to express
symptoms of such injury. This potential for injury is greater on especially
higher-field whole-body scanners (e.g., 1 Tesla and above). Distortion of the
electrocardiogram within the magnetic field makes interpretation of the ECG
complex unreliable, even with filtering used by contemporary monitoring
systems. However, routine monitoring of heart rate and rhythm may be
accomplished using pulse oximetry, which also eliminates the risks of thermal
injury from electrocardiography. Patients who require ECG monitoring and who
are unconscious, sedated, or anesthetized should be examined after each
imaging sequence, with potential repositioning of the ECG leads and any other
electrically conductive material with which the patient is in contact.
Alternatively, cold compresses or ice packs could be placed upon all necessary
electrically conductive material that touches the patient during scanning.
- Final determination of whether or not to scan any given patient with any
given implant, foreign body, etc., is to be made by the level 2 MR
personneldesignated attending MR radiologist, the MR medical director,
or specifically designated level 2 MR personnel following criteria for
acceptability predetermined by the medical director.
For implants that are strongly ferromagnetic, an obvious concern is that of
magnetic translational and rotational forces upon the implant which might move
or dislodge the device from its implanted position. If an implant has
demonstrated weak ferromagnetic forces on formal testing, it might be prudent
to wait several weeks for fibrous scarring to set in, as this may help anchor
the implant in position and help it resist such weakly attractive magnetic
forces that might arise in MR environments.
For all implants that have been demonstrated to be nonferrous in nature,
however, the risk of implant motion is essentially reduced to those resulting
from Lenz's forces alone. These tend to be quite trivial for typical metallic
implant sizes of a few centimeters or less. Thus, a waiting period for fibrous
scarring to set in is far less important, and the advisability for such a
waiting period may well be easily outweighed by the potential clinical
benefits of undergoing an MR examination at that time. As always, clinical
assessment of the riskbenefit ratio for the particular clinical
situation and patient at hand are paramount for appropriate medical decision
making in these scenarios.
It is possible that during the course of an MRI examination an
unanticipated ferromagnetic implant or foreign body is discovered within a
patient or research subject undergoing the examination. This is typically
suspected or detected by means of a sizable field-distorting artifact seen on
spin-echo imaging techniques that grows more obvious on longer TE studies and
expands markedly on typical moderate or long TE gradient-echo imaging
sequences. In such cases, it is imperative that the medical director, safety
officer, and/or physician in charge be immediately notified of the suspected
findings. This individual should then assess the situation, review the imaging
information obtained, and decide what the best course of action might be.
It should be noted that there are numerous potentially acceptable courses
that might be recommended which in turn depend upon many factors, including
the status of the patient, the location of the suspected ferromagnetic
implant/foreign body relative to local anatomic structures, the mass of the
implant, etc. Appropriate courses of action might include proceeding with the
scan under way, immobilizing the patient and the immediate removal of the
patient from the scanner, or other intermediate steps. Regardless of the
course of action selected, it is important to note that the forces on the
implant will change, and may actually increase, during the attempt to remove
the patient from the scanner bore. Further, the greater the rate of motion of
the patient/device through the magnetic fields of the scanner bore, the
greater the forces acting upon that device will likely be. Thus, it is prudent
to ensure that, if at all possible, immobilization of the device during
patient extraction from the bore, and the slow, cautious, deliberate rate of
extricating the patient from the bore, will likely result in weaker and
potentially less harmful forces on the device as it traverses the various
static magnetic field gradients associated with the MR imager.
It is also worthy of note that the magnetic fields associated with the MR
scanner are distributed throughout space three-dimensionally. Thus, especially
for superconducting systems, one should avoid the temptation to have the
patient sit up as soon as he or she is physically out of the bore. Doing so
may expose the ferrous object to still-significant torque- and
translation-related forces despite the patient's being physically outside the
scanner bore. It is therefore advisable to continue to extract the patient
along a straight line course parallel to the center of the magnet while the
patient remains immobilized until they are as far as physically possible from
the MR imager itself, before any other patient/object motion vector is
attempted or permitted.
- All non-MR personnel (e.g., patients, volunteers, varied site employees,
and professionals) with implanted cardiac pacemakers, autodefibrillators,
diaphragmatic pacemakers, or other electromechanically activated devices upon
which the non-MR personnel is dependent should be precluded from Zone IV and
physically restrained from the 5-gauss line unless specifically cleared in
writing by a level 2 MR personneldesignated attending radiologist or
the medical director of the MR site. In such circumstances, a specific
defending riskbenefit rationale should be provided in writing and
signed by the authorizing radiologist.
Should it be determined that non-MR personnel wishing to accompany a
patient into an MR scan room require their orbits to be cleared by plain-film
radiography, a radiologist must first discuss with the non-MR personnel that
plain X-ray films of their orbits are required prior to permitting them access
to the MR scan room. Should they still wish to proceed with access to Zone IV
or within the 5-gauss line, and should the attending radiologist deem it
medically advisable that they do so (e.g., for the care of their child about
to undergo an MR study), written informed consent should be provided by these
accompanying non-MR personnel prior to their undergoing X-ray examination of
their orbits.
- MR scanning of patients, prisoners, or parolees with metallic
prisoner-restraining devices or RF ID or tracking bracelets could lead to
theoretic adverse events, including: (1) ferromagnetic attractive effects and
resultant patient injury, (2) possible ferromagnetic attractive effects and
potential damage to the device or its battery pack, (3) RF interference with
the MRI study and secondary image artifact, (4) RF interference with the
functionality of the device, (5) RF power deposition and heating of the
bracelet or tagging device or its circuitry and secondary patient injury (if
the bracelet were in the anatomic volume of the RF transmitter coil being used
for imaging). Therefore, when requested to scan a patient, prisoner, or
parolee wearing RF bracelets or metallic handcuffs or ankle cuffs, request
that the patient be accompanied by the appropriate authorities who can and
will remove the restraining device prior to the MR study and be charged with
its replacement following the examination.
- Firefighter, police, and security safety considerations: For the safety of
firefighters and other emergent services responding to an emergent call at the
MR site, it is recommended that all fire alarms, cardiac arrests, or other
emergent service response calls originating from or located in the MR site
should be forwarded simultaneously to a specifically designated individual
from among the site's MR personnel. This individual should, if possible, be on
site prior to the arrival of the firefighters or emergent responders to ensure
that they do not have free access to Zone III or Zone IV. The site might
consider assigning appropriately trained security personnel, who have been
trained and designated as MR personnel, to respond to such calls.
In any case, all MR sites should arrange to prospectively educate their
local fire marshals, firefighters' associations, and police or security
personnel about the potential hazards of responding to emergencies in the MR
suite.
It should be stressed that even in the presence of a true fire (or other
emergency) in Zone III or Zone IV, the magnetic fields may be present and
fully operational. Therefore, free access to Zone III or Zone IV by
firefighters or other non-MR personnel with air tanks, axes, crowbars, other
firefighting equipment, guns, etc., might prove catastrophic or even lethal to
those responding or to others in the vicinity.
As part of the Zone III and Zone IV restrictions, all MR sites must have
clearly marked, readily accessible MR-conditional or MR-safe fire
extinguishing equipment physically stored in Zone III or Zone IV. All
conventional fire extinguishers and other firefighting equipment not tested
and verified safe in the MR environment should be restricted from Zone
III.
For superconducting magnets, the helium (and the nitrogen as well, in older
MR magnets) is not flammable and does not pose a fire hazard directly.
However, the liquid oxygen that can result from the supercooled air in the
vicinity of the released gases might well increase the fire hazard in this
area. If there are appropriately trained and knowledgeable MR personnel
available during an emergency to ensure that emergency response personnel are
kept out of the MR scanner or magnet room and away from the 5-gauss line,
quenching the magnet during a response to an emergency or fire should not be a
requirement.
However, if the fire is in such a location where Zone III or Zone
IV needs to be entered for whatever reason by firefighting or emergency
response personnel and their firefighting and emergent equipment, such as air
tanks, crowbars, axes, and defibrillators, a decision to quench a
superconducting magnet should be very seriously considered to protect
the health and lives of the emergent responding personnel. Should a quench be
performed, appropriately designated MR personnel still need to ensure that
all non-MR personnel (including and especially emergent response
personnel) continue to be restricted from Zones III and IV until the
designated MR personnel has personally verified that the static field is
either no longer detectable or at least sufficiently attenuated as to no
longer present a potential hazard to one moving by it with, for example, large
ferromagnetic objects such as air tanks or axes.
For resistive systems, the magnetic field of the MR scanner should be shut
down as completely as possible and verified as such prior to permitting the
emergency response personnel access to Zone IV. For permanent, resistive, or
hybrid systems whose magnetic fields cannot be completely shut down, MR
personnel should ideally be available to warn the emergency response personnel
that a very powerful magnetic field is still operational in the magnet
room.
4. MR personnel screening
All MR personnel are to undergo an MR screening process as part of their
employment interview process to ensure their safety in the MR environment. For
their own protection and for the protection of the non-MR personnel under
their supervision, all MR personnel must immediately report to the MR medical
director any trauma, procedure, or surgery they experience or undergo in which
a ferromagnetic metallic object or device may have become introduced within or
on them. This will permit appropriate screening to be performed on the
employee to determine the safety of permitting that employee into Zone
III.
5. Device and object screening
Ferrous objects, including those brought by patients, visitors,
contractors, etc., should be restricted from entering Zone III, whenever
practical.
As part of the Zone III site restriction and equipment testing and clearing
responsibilities, all sites should have ready access to a strong handheld
magnet (
1000 gauss). This will enable the site to test external, and even
some superficial internal, devices or implants for the presence of grossly
detectable ferromagnetic attractive forces.
- All portable metallic or partially metallic devices that are on or external
to the patient (e.g., oxygen cylinders) are to be positively identified in
writing as ferromagnetic or, alternatively, nonferromagnetic and safe or
conditionally safe in the MR environment prior to permitting them into Zone
III. For all device or object screening, verification and positive
identification should be in writing. Examples of devices that need to be
positively identified include fire extinguishers, oxygen tanks, and aneurysm
clips.
- External devices or objects demonstrated to be ferromagnetic and MR unsafe
or incompatible in the MR environment may still, under specific circumstances,
be brought into Zone III if, for example, they are deemed by MR personnel to
be necessary and appropriate for patient care. They should only be brought
into Zone III if they are under the direct supervision of specifically
designated level 1 or level 2 MR personnel who are thoroughly familiar with
the device, its function, and the reason supporting its introduction to Zone
III. The safe utilization of these devices while they are present in Zone III
will be the responsibility of specifically named level 1 or 2 MR personnel.
These devices must be appropriately physically secured or restricted at all
times during which they are in Zone III to ensure that they do not
inadvertently come too close to the MR scanner and accidentally become exposed
to static magnetic fields or gradients that might result in their becoming
either hazardous projectiles or no longer accurately functional.
- Never assume MR compatibility or safety information about the device if it
is not clearly documented in writing. All unknown external objects or devices
being considered for introduction beyond Zone II should be tested with a
strong handheld magnet (
1000 gauss) for ferromagnetic properties before
permitting them entry to Zone III. The results of such testing, as well as the
date, time, and name of the tester, and methodology used for that particular
device, should be documented in writing. If a device has not been tested, or
if its MR compatibility or safety status is unknown, it should not be
permitted unrestricted access to Zone III.
- All portable metallic or partially metallic objects that are to be brought
into Zone IV must be properly identified and appropriately labeled utilizing
the current FDA labeling criteria developed by ASTM (American Society for
Testing and Materials) International
(http://www.astm.org)
(see Fig. 2). Those items which
are wholly nonmetallic should be identified with a square green "MR
safe" label. Items which are clearly ferromagnetic should be identified
as "not MR safe" and labeled appropriately with the corresponding
round red label with a slash through it. Objects with an "MR
conditional" rating should be affixed with a triangular yellow MR
conditional label prior to being taken into the scan room/Zone IV.
As noted in the introduction to this section B.5, above, if MR safety data
are not prospectively available for a given device, initial testing for the
purpose of this labeling is to be accomplished by the site's MR personnel by
exposing the metallic object to a handheld magnet (
1000 gauss). If
grossly detectable attractive forces are observed between the object being
tested or any of its components and the handheld magnet, it is to be labeled
with a circular red "not MR safe" label. If no or negligible
attractive forces are observed, a triangular yellow "MR
conditional" label is to be attached to the object. It is only when the
composition of an object and its components are known to be nonmetallic that
the green "MR safe" label is to be affixed to a device or
object.
Particularly with regard to nonclinical and incidental equipment, current
products marketed with ill-defined terminology such as
"non-magnetic," or outdated classifications such as
"MR-compatible," should not be presumed to conform to a particular
current ASTM classification. Similarly, any product marketed as "MR
safe" but with metallic construction or components should be treated
with suspicion. Objects intended for use in Zone IV, including nonclinical
incidental products such as stepping stools or ladders, which are not provided
with manufacturer or third-party MR safety test results under the new ASTM
criteria, should be site tested as described above.
- Decisions based on published MR compatibility or safety claims should
recognize that all such claims apply to specifically tested static field and
static gradient field strengthsfor example, "MR conditional,
having been tested to be safe up to 3.0 Tesla at gradient strengths of 400
G/cm," or "MR conditional, having been tested to be safe up to 1.5
Tesla up to maximum static gradient fields experienced in an unshielded
1.5-Tesla [manufacturer's name] wholebody MR scanner tested 1.5 feet (roughly
45 cm) within the bore."
- It should be noted that alterations performed by the site on MR safe, MR
unsafe, and MR conditional equipment or devices may alter the MR safety or
compatibility properties of the device. For example, tying a ferromagnetic
metallic twisting binder onto a sign labeling the device as MR conditional or
MR safe might result in artifact inductionor worseif introduced
into the MR scanner.
Lenz's Forces:
Faraday's law states that a moving or changing magnetic field will induce a
voltage in a perpendicularly oriented electrical conductor. Lenz's law builds
upon this and states that the induced voltage will itself be such that it will
secondarily generate its own magnetic field whose orientation and magnitude
will oppose those of the initial time-varying magnetic field that created it
in the first place. For example, if an electrical conductor is moved
perpendicularly toward the magnetic field, B0, of an MR scanner,
even if this conductor is not grossly ferromagnetic, the motion itself will
result in the generation of voltages in this conductor whose magnitude is
directly proportional to the rate of motion as well as the spatial gradient of
the magnetic field, B0, through which it is being moved. Conducting
objects turning in the static field will also experience a torque due to the
induced eddy currents. Lenz's law states that this induced current will in
turn create a magnetic field whose orientation will oppose the B0
magnetic field that created this current.
Thus, moving a large metallic but nonferromagnetic electrical conductor
toward the magnet bore will result in the induction of a voltage and
associated magnetic field which will orient in such a manner and at such a
strength as to oppose the motion of the metallic object into the bore of the
MR scanner. If, for example, one tries to move a nonferrous oxygen tank into
the bore of an MR scanner, as the scanner bore is approached Lenz's forces
will be sufficiently strong to virtually stop forward progress of the tank.
Further, the faster one moves the tank into the bore, the greater the opposing
force that is created to stop this motion.
This also has potential consequences for large implanted metallic devices
such as certain metallic nonferrous infusion pumps. Although they may not pose
a projectile hazard, rapid motion of the patient/implant perpendicular to the
magnetic field of the MR imager can be expected to result in forces on the
implant that would oppose this motion and may likely be detected by the
patient. If the patient were to complain of experiencing forces tugging or
pulling on the implant, this might bring the patient or health care personnel
to erroneously conclude that there were ferrous components to the device,
which might lead to cancellation of the examination. Slowly moving such large
metallic devices into and out of the bore is a key factor in decreasing any
Lenz's forces that might be induced and in decreasing the likelihood of a
misunderstanding or an unnecessary study cancellation.

View larger version (16K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2 U.S. Food and Drug Administration labeling criteria
(developed by ASTM [American Society for Testing and Materials] International)
for portable objects taken into Zone IV. Square green "MR safe"
label is for wholly nonmetallic objects, triangular yellow label is for
objects with "MR conditional" rating, and round red label is for
"not MR safe" objects.
|
|
C. MR Technologists
- MR technologists should be ARRT (American Registry of Radiologic
Technologists)registered technologists (RTs). Furthermore, all MR
technologists must be trained as level 2 MR personnel during their orientation
prior to being permitted free access to Zone III.
- All MR technologists will maintain current certification in American Heart
Association basic life support at the health care provider level.
- Except for emergent coverage, there will be a minimum of 2 MR technologists
or one MR technologist and one other individual with the designation of MR
personnel in the immediate Zone II through Zone IV MR environment. For
emergent coverage, the MR technologist can scan with no other individuals in
their Zone II through Zone IV environment as long as there is in-house, ready
emergent coverage by designated department of radiology MR personnel (e.g.,
radiology house staff or attending radiologist).
D. Pregnancy-Related Issues
1. Health care practitioner pregnancies
Pregnant health care practitioners are permitted to work in and around the
MR environment throughout all stages of their pregnancy
[14]. Acceptable activities
include, but are not limited to, positioning patients, scanning, archiving,
injecting contrast material, and entering the MR scan room in response to an
emergency. Although permitted to work in and around the MR environment,
pregnant health care practitioners are requested not to remain within the MR
scanner bore or Zone IV during actual data acquisition or scanning.
2. Patient pregnancies
Present data have not conclusively documented any deleterious effects of MR
imaging exposure on the developing fetus. Therefore, no special consideration
is recommended for the first, versus any other, trimester in pregnancy.
Nevertheless, as with all interventions during pregnancy, it is prudent to
screen women of reproductive age for pregnancy prior to permitting them access
to MR imaging environments. If pregnancy is established, consideration should
be given to reassessing the potential risks versus benefits of the pending
study in determining whether performance of the requested MR examination could
safely wait until the end of the pregnancy.
- Pregnant patients can be accepted to undergo MR scans at any stage of
pregnancy if, in the determination of a level 2 MR personneldesignated
attending radiologist, the riskbenefit ratio to the patient warrants
that the study be performed. The radiologist should confer with the referring
physician and document the following in the radiology report or the patient's
medical record:
- The information requested from the MR study cannot be acquired via
nonionizing means (e.g., ultrasonography).
- The data are needed to potentially affect the care of the patient or fetus
during the pregnancy.
- The referring physician does not feel it is prudent to wait until the
patient is no longer pregnant to obtain these data.
- MR contrast agents should not be routinely provided to pregnant
patients. This decision, too, is one that must be made on a case-by-case basis
by the covering level 2 MR personneldesignated attending radiologist
who will assess the riskbenefit ratio for that particular patient.
The decision to administer a gadolinium-based MR contrast agent to pregnant
patients should be accompanied by a well-documented and thoughtful
riskbenefit analysis. This analysis should be able to defend a decision
to administer the contrast agent based on overwhelming potential benefit to
the patient or fetus outweighing the theoretic but potentially real risks of
long-term exposure of the developing fetus to free gadolinium ions.
Studies have demonstrated that gadolinium-based MR contrast agents pass
through the placental barrier and enter the fetal circulation. From there,
they are filtered in the fetal kidneys and then excreted into the amniotic
fluid. In this location the gadolinium-chelate molecules are in a relatively
protected space and may remain in this amniotic fluid for an indeterminate
amount of time before finally being reabsorbed and eliminated. As with any
equilibrium situation involving any dissociation constant, the longer the
chelate molecule remains in this space, the greater the potential for
dissociation of the potentially toxic gadolinium ion from its chelate
molecule. It is unclear what impact such free gadolinium ions might have if
they were to be released in any quantity in the amniotic fluid. Certainly,
deposition into the developing fetus would raise concerns of possible
secondary adverse effects.
The risk to the fetus with administration of gadolinium-based MR contrast
agents remains unknown and may be harmful.
- It is recommended that pregnant patients undergoing an MR examination
provide written informed consent documenting that they understand the
potential risks and benefits of the MR procedure to be performed, are aware of
the alternative diagnostic options available to them (if any), and wish to
proceed.
E. Pediatric MR Safety Concerns
1. Sedation and monitoring issues
Children form the largest group requiring sedation for MRI, largely because
of their inability to remain motionless during scans. Sedation protocols may
vary from institution to institution according to the procedures performed
(diagnostic vs interventional), the complexity of the patient population
(healthy preschoolers vs premature infants), the method of sedation (mild
sedation vs general anesthesia), and the qualifications of the sedation
provider.
Adherence to standards of care mandates following the sedation guidelines
developed by the American Academy of Pediatrics
[15,
16], the American Society of
Anesthesiologists [17], and
the Joint Commission on Accreditation of Healthcare Organizations (JCAHO)
[18]. In addition, sedation
providers must comply with protocols established by the individual state and
the practicing institution. These guidelines require the following
provisions:
- Preprocedural medical history and examination for each patient
- Fasting guidelines appropriate for age
- Uniform training and credentialing for sedation providers
- Intraprocedural and postprocedural monitors with adaptors appropriately
sized for children (compatible with the magnetic field)
- Method of patient observation (window, camera)
- Resuscitation equipment, including oxygen delivery and suction
- Uniform system of record keeping and charting (with continuous assessment
and recording of vital signs)
- Location and protocol for recovery and discharge
- Quality assurance program that tracks complications and morbidity.
For the neonatal and the young pediatric population, special attention is
needed in monitoring body temperature for both hypo- and hyperthermia in
addition to other vital signs
[19]. Temperature-monitoring
equipment that is approved for use in the MR suite is becoming more readily
available. Commercially available, MR-approved neonatal isolation transport
units and other warming devices are also available for use during MR
scans.
2. Pediatric screening issues
Children may not be reliable historians and, especially in cases of older
children and teenagers, should be questioned both in the presence of parents
or guardians and separately to maximize the possibility that all potential
dangers are disclosed. Therefore, it is recommended that children be gowned
before entering Zone IV to help ensure that no metallic objects, toys, etc.
inadvertently find their way into Zone IV. Pillows, stuffed animals, and other
comfort items brought from home represent real risks and should be discouraged
from entering Zone IV. If unavoidable, each such item should be carefully
checked with the powerful handheld magnet and perhaps again in the MR scanner
prior to permitting the patient to enter Zone IV with the object in order to
ensure that it does not contain any objectionable metallic components.
3. MR safety of accompanying family or personnel
Although any age patient might request that others accompany them for their
MR examination, this is far more common in the pediatric population. Those
accompanying or remaining with the patient should be screened using the same
criteria as anyone else entering Zone IV.
In general, it would be prudent to limit accompanying adults to a single
individual. Only a qualified, responsible MR physician should make screening
criteria exceptions.
Hearing protection and MR safe/MR conditional seating are recommended for
accompanying family members within the MR scan room.
F. Time-Varying Gradient Magnetic FieldRelated Issues: Induced Voltages
Types of patients needing extra caution:
Patients with implanted or retained wires in anatomically or functionally
sensitive areas (e.g., myocardium or epicardium, implanted electrodes in the
brain) should be considered to be at higher risk, especially from faster MRI
sequences, such as echo-planar imaging (which may be used in such sequences as
diffusion-weighted imaging, functional imaging, perfusion-weighted imaging, MR
angiographic imaging, etc.). The decision to limit the dB/dt (rate of magnetic
field change) and maximum strength of the magnetic field of the gradient
subsystems during imaging of such patients should be reviewed by the level 2
MR personneldesignated attending radiologist supervising the case or
patient.
G. Time-Varying Gradient Magnetic FieldRelated Issues: Auditory Considerations
- All patients and volunteers should be offered and encouraged to use hearing
protection prior to undergoing any imaging in the MR scanners.
- All patients or volunteers in whom research sequences are to be performed
(i.e., MR scan sequences that have not yet been approved by the Food and Drug
Administration) are to have hearing protective devices in place prior
to the initiation of any MR sequences. Without hearing protection in place,
MRI sequences that are not FDA-approved should not be performed on patients or
volunteers.
H. Time-Varying Radiofrequency Magnetic FieldRelated Issues: Thermal
- All unnecessary or unused electrically conductive materials should be
removed from the MR system before the onset of imaging. It is not sufficient
to merely to "unplug" or disconnect unused, unnecessary
electrically conductive material and leave it within the MR scanner with the
patient during imaging. All electrical connections, such as on surface coil
leads or monitoring devices, must be visually checked by the scanning MR
technologist prior to each use to ensure the integrity of the thermal and
electrical insulation.
- Electrical voltages and currents can be induced in electrically conductive
materials that are within the bore of the MR imager during the MR imaging
process. This might result in the heating of this material by resistive
losses. This heat might be of a caliber sufficient to cause injury to human
tissue. Among the variables that determine the amount of induced voltage or
current is the consideration that the larger the diameter of the conductive
loops, the greater the potential induced voltages or currents, and thus the
greater the potential for resultant thermal injury to adjacent or contiguous
patient tissue.
Therefore, when electrically conductive material (wires, leads, implants,
etc.) are required to remain within the bore of the MR scanner with the
patient during imaging, care should be taken to ensure that no large-caliber
electrically conducting loops (including patient tissue; see section H.5,
below) are formed within the MR scanner during imaging. Furthermore, it is
possible, with the appropriate configuration, lead length, static magnetic
field strength, and other settings, to introduce resonant circuitry between
the transmitted RF power and the lead. This could result in very rapid and
clinically significant lead heating, especially at the lead tips, in a matter
of seconds to a magnitude sufficient to result in tissue thermal injury or
burns. This can also theoretically occur with implanted leads or wires, even
when they are not connected to any other device at either end. For
illustration, the FDA has noted several reports of serious injury, including
coma and permanent neurologic impairment, in patients with implanted
neurologic stimulators who underwent MR imaging examinations. The injuries in
these instances resulted from heating of the electrode tips
[20,
21].
Further, it is entirely possible for a lead or wire to demonstrate no
significant heating while undergoing MR imaging examinations at 1.5 Tesla, yet
demonstrate clinically significant and potentially harmful degrees of heating
within seconds at, for example, 3 Tesla. It has also been demonstrated that
leads may show no significant heating at 3 Tesla yet may rapidly heat to
hazardous levels when undergoing MR imaging at, for example, 1.5 Tesla
(personal observation, MR safety testing, E. Kanal, MD, University of
Pittsburgh Medical Center MR Research Center, 8/28/05). Thus, at no time
should a label of "MR conditionally safe for thermal issues at [a given
field strength]" be applied to any field strength, higher or lower,
other than the specific one at which safety was demonstrated.
Thus, exposure of electrically conductive leads or wires to the RF
transmitted power during MR scanning should only be performed with caution and
with appropriate steps taken to ensure significant lead or tissue heating does
not result (see section H.9, below).
- When electrically conductive materials are required to be within the bore
of the MR scanner with the patient during imaging, care should be taken to
place thermal insulation (including air, pads, etc.) between the patient and
the electrically conductive material, while simultaneously attempting (as much
as feasible) to keep the electrical conductor from directly contacting the
patient during imaging. It is also appropriate to try to position the leads or
wires as far as possible from the inner walls of the MR scanner if the body
coil is being used for RF transmission. When it is necessary that electrically
conductive leads directly contact the patient during imaging, consideration
should be given to prophylactic application of cold compresses or ice packs to
such areas.
- Depending on specific magnet designs, care may be needed to ensure that the
patient's tissue(s) do not directly come into contact with the inner bore of
the MR imager during the MRI process. This is especially important for several
higher-field MR scanners. The manufacturers of these devices provide pads and
other such insulating devices for this purpose, and manufacturer's guidelines
should be strictly adhered to for these units.
- It is important to ensure the patient's tissues do not form large
conductive loops. Therefore, care should be taken to ensure that the patient's
arms or legs are not positioned in such a way as to form a large-caliber loop
within the bore of the MR imager during the imaging process. For this reason,
it is preferable that patients be instructed not to cross their arms or legs
in the MR scanner. We are also aware of unpublished reports of thermal
injuries that seem to have been associated with skin folds, such as in the
region of the inner thighs. While the cause of this is not yet fully
understood, it might be prudent to consider ensuring that skin folds and other
such examples of tissue-to-tissue contact are minimized or eliminated in the
region undergoing radiofrequency energy irradiation.
- Skin staples and superficial metallic sutures: Patients requested to
undergo MR studies in whom there are skin staples or superficial metallic
sutures (SMS) may be permitted to undergo the MR examination if the skin
staples or SMS are not ferromagnetic and are not in the anatomic volume of RF
power deposition for the study to be performed. If the nonferromagnetic skin
staples or SMS are within the volume to be RF-irradiated for the requested MR
study, several precautions are recommended.
- Warn the patient and make sure that they are especially aware of the
possibility that they may experience warmth or even burning along the skin
staple or SMS distribution. The patient should be instructed to report
immediately if they experience warmth or burning sensations during the study
(and not, for example, wait until the "end of the knocking
noise").
- It is recommended that a cold compress or ice pack be placed along the skin
staples or SMS if this can be safely clinically accomplished during the MRI
examination. This will help to serve as a heat sink for any focal power
deposition that may occur, thus decreasing the likelihood of a clinically
significant thermal injury or burn to adjacent tissue.
- For patients with extensive or dark tattoos, including tattooed eyeliner,
in order to decrease the potential for RF heating of the tattooed tissue, it
is recommended that cold compresses or ice packs be placed on the tattooed
areas and kept in place throughout the MRI process if these tattoos are in the
volume in which the body coil is being used for RF transmission. This approach
is especially appropriate if fast spin-echo (or other high RF duty cycle) MRI
sequences are anticipated in the study. If another coil is being used for RF
transmission, a decision must be made if high RF transmitted power is to be
anticipated by the study protocol design. If so, then the above precautions
should be followed. Additionally, patients with tattoos that had been placed
within 48 hours prior to the pending MR examination should be advised of the
potential for smearing or smudging of the edges of the freshly placed
tattoo.
- In the unconscious or unresponsive patient, all attached leads that will be
in or partly in the volume undergoing RF irradiation should be covered with a
cold compress or ice pack at the lead attachment site for the duration of the
MR study.
- As noted above, it has been demonstrated that resonant circuitry can be
established during MRI between the RF energies being transmitted and specific
lengths of long electrically conductive wires or leads, which can thus act as
efficient antennae. This can result in heating of the tips of these wires or
leads to temperatures in excess of 90°C in a few seconds. Therefore,
patients in whom there are long electrically conductive leads, such as
Swan-Ganz thermodilution cardiac outputcapable catheters or Foley
catheters with electrically conductive leads, should be considered at risk for
MR studies if the body coil is to be used for RF transmission over the region
of the electrically conductive lead. This is especially true for higher-field
systems and for imaging protocols utilizing fast spin-echo or other high RF
duty cycle MRI sequences. Each such patient should be reviewed and cleared by
an attending level 2 radiologist and a riskbenefit ratio assessment
performed prior to permitting them access to the MR scanner.
- The potential to establish substantial heating is itself dependent on
multiple factors, including, among others, the static magnetic field strength
of the MR scanner (as this determines the transmitted radiofrequencies [RF] at
which the device operates) and the length, orientation, and inductance of the
electrical conductor in the RF-irradiated volume being studied. Virtually
any lead lengths can produce substantial heating. Innumerable factors can
affect the potential for tissue heating for any given lead. It is therefore
critical to recognize that of all electrically conductive implants, it is
specifically wires, or leads, that pose the greatest potential hazard for
establishing substantial power deposition/heating considerations.
Another important consideration is that as a direct result of the above, it
has already been demonstrated in vitro that heating of certain implants or
wires may be clinically insignificant at, for example, 1.5 Tesla but quite
significant at 3.0 Tesla. However, it has also been demonstrated that specific
implants might show no significant thermal issues or heating at 3.0 Tesla,
but may heat to clinically significant or very significant levels in
seconds at, for example, 1.5 Tesla. Thus, it is important to follow
established product MR safety guidelines carefully and precisely, applying
them to, and only to, the static magnetic field strengths at which they had
been tested. MR scanning at either stronger and/or weaker magnetic field
strengths than those tested may result in significant heating where none had
been observed at the tested field strength(s).
I. Drug Delivery Patches and Pads
Some drug delivery patches contain metallic foil. Scanning the region of
the metallic foil may result in thermal injury
[22]. Since removal or
repositioning can result in altering of patient dose, consultation with the
patient's prescribing physician would be indicated in assessing how to best
manage the patient. If the metallic foil of the patch delivery system is
positioned on the patient so that it is in the volume of excitation of the
transmitting RF coil, the case should be specifically reviewed with the
radiologist or physician covering the patient. Alternative options may include
placing an ice pack directly on the patch. This solution may still
substantially alter the rate of delivery or absorption of the medication to
the patient (and be less comfortable to the patient, as well). This
ramification should therefore not be treated lightly, and a decision to
proceed in this manner should be made by a knowledgeable radiologist attending
the patient and with the concurrence of the referring physician as well.
If the patch is removed, a specific staff member should be given
responsibility for ensuring that it is replaced or repositioned.
J. Cryogen-Related Issues
- For superconducting systems, in the event of a system quench, it is
imperative that all personnel and patients be evacuated from the MR scan room
as quickly as safely feasible and that the site access be immediately
restricted to all individuals until the arrival of MR equipment service
personnel. This is especially so if cryogenic gases are observed to have
vented partially or completely into the scan room, as evidenced in part by the
sudden appearance of white "clouds" or "fog" around or
above the MR scanner. As noted in section B.3.n above, it is especially
important to ensure that all police and fire response personnel are restricted
from entering the MR scan room with their equipment (axes, air tanks, guns,
etc.) until it can be confirmed that the magnetic field has been successfully
dissipated, because there may still be a considerable static magnetic field
present despite a quench or partial quench of the magnet
[23].
- It should be pointed out that room oxygen monitoring was discussed by the
MR Blue Ribbon Panel and rejected at this time because the present oxygen
monitoring technology was considered by industry experts not to be
sufficiently reliable to allow continued operation during situations of power
outages, etc.
K. Claustrophobia, Anxiety, Sedation, Analgesia, and Anesthesia
Adult and pediatric patient anxiolysis, sedation, analgesia, and anesthesia
for any reason should follow established ACR
[24,
25], American Society of
Anesthesiologists (ASA)
[2629],
and JCAHO standards [29].
L. Contrast Agent Safety
1. Contrast agent administration issues
No patient is to be administered prescription MR contrast agents without
orders from a duly licensed physician. Intravenous injectionqualified
MR technologists may start and attend to peripheral IV access lines if they
have undergone the requisite site-specified training in peripheral IV access
and have demonstrated and documented appropriate proficiency in this area.
IV-qualified MR technologists may administer FDA-approved gadolinium-based MR
contrast agents via peripheral IV routes as a bolus or as a slow or continuous
injection as directed by the orders of a duly licensed site physician.
Administration of these agents is to be performed according to the ACR
policy. The ACR approves of the injection of contrast material and diagnostic
levels of radiopharmaceuticals by certified and/or licensed radiologic
technologists and radiologic nurses under the direction of a radiologist or
his or her physician designee who is personally and immediately available, if
the practice is in compliance with institutional and state regulations. There
must also be prior written approval by the medical director of the radiology
department or service of such individuals. Such approval process must follow
established policies and procedures, and the radiologic technologists and
nurses who have been so approved must maintain documentation of continuing
medical education related to materials injected and to the procedures being
performed [30].
2. Prior contrast agent reaction issues
- According to the ACR Manual on Contrast Media
[31], adverse events after
intravenous injection of gadolinium seem to be more common in patients who had
previous reactions to an MR contrast agent. In one study, 16 (21%) of 75
patients who had previous adverse reactions to MR contrast agents reacted to
subsequent injections of gadolinium
[31]. Patients with asthma
also seem to be more likely to have an adverse reaction to the administration
of a gadolinium-based MR contrast agent. Patients with allergies also seemed
to be at increased risk (
2.03.7 times, compared with patients
without allergies). Patients who have had adverse reactions to iodinated
contrast media are more than twice as likely to have an adverse reaction to
gadolinium (6.3% of 857 patients)
[31].
- At present, there are no well-defined policies for patients who are
considered to be at increased risk for having an adverse reaction to MR
contrast agents. However, the following recommendations are suggested:
Patients who have previously reacted to one MR contrast agent can be injected
with another agent if they are restudied, and at-risk patients can be
premedicated with corticosteroids and, occasionally, antihistamines
[31].
- All patients with asthma, a history of allergic respiratory disorders,
prior iodinated or gadolinium-based contrast reactions, etc., should be
followed m