DOI:10.2214/AJR.06.1146
AJR 2007; 188:1648-1650
© American Roentgen Ray Society
MOSFET Dosimetry for Radiation Dose Assessment of Bismuth Shielding of the Eye in Children
Srinivasan Mukundan, Jr.1,
Page Inman Wang,
Donald P. Frush,
Terry Yoshizumi,
Jeffrey Marcus,
Emily Kloeblen and
Meredith Moore
1 All authors: Department of Radiology, Duke University Medical Center, DUMC Box
3808, Durham, NC 27705.
Received August 27, 2006;
accepted after revision November 21, 2006.
S. Mukundan is a 20052006 American Roentgen Ray Society Scholar.
Address correspondence to S. Mukundan, Jr.
Abstract
OBJECTIVE. The purpose of our study was to measure radiation dose to
the orbit during pediatric cranial CT with and without bismuth shielding using
a novel dosimetry system. Cranial CT was performed on a pediatric
anthropomorphic phantom, with and without bismuth eye shields. A solid-state
metal oxide semiconductor field effect transistor (MOSFET) dosimeter was used
to obtain real-time dose measurements.
CONCLUSION. Bismuth shielding reduced radiation dose to the eye by
up to 42%; shield artifact fell outside the diagnostic area of interest.
Keywords: bismuth shielding CT dosimetry eye head and neck imaging pediatric imaging radiation dose safety
Introduction
The advent of MDCT has resulted in improved spatial resolution and faster
scan acquisitions. Consequently, MDCT has become a more widely used diagnostic
procedure, responsible for a greater proportion of medical radiation exposure
to patients. Although CT represents 11% of all radiographic examinations, it
accounts for 67% of medically induced radiation exposure
[1]. Ten percent of all CT
examinations are performed in children (age range, birth15 years) to
assess for disorders including trauma, congenital anomalies, metabolic
diseases, inflammatory lesions, and tumors
[1].
The eye is one of the most radiosensitive tissues, and the threshold for
inducing cataracts in adults has been documented as low as 0.52 Gy
(50200 rad) [2]. In
children this threshold is even lower, with the development of cataracts
having been documented at less than half this dose of radiation
[2]. Therefore, it is of
paramount importance to shield the pediatric orbit from any unnecessary
radiation during CT. Typically, the dose to the eye is about 50 mGy (5 rad),
depending on the instrument and protocol
[1]. Moreover, the
as-low-as-reasonably-achievable (ALARA) principle would dictate that the
radiation dose be limited.
Our objective for this study was to assess a contemporary method for
measuring radiation dose to the pediatric lens and orbit and to apply this
method to determine whether the novel use of a bismuth-impregnated latex eye
shield could offer protection from radiation without a loss in di agnostic
quality.
Materials and Methods
Metal oxide semiconductor field effect transistors (MOSFETs) were used to
measure radiation dose rather than traditional thermoluminescence dosimeters
(TLDs) because MOSFETs are sensitive to a few milligrays and have a linear
response at these doses [3].
Because MOSFETs have immediate readout and reuse, they are convenient for MDCT
use [3]. The MOSFET
high-sensitivity dosimeters (TN-1002RD, Best Medical Canada, Ltd.) were
individually calibrated at each CT energy using a National Institute of
Standards and Technology (NIST) traceable calibrated ionization chamber.
Before the study, a linearity test was performed to verify the actual dose
response to prescribed instrument mA settings for the range 80300
mA.
The 5-year-old pediatric anthropomorphic phantoms (705-D, CIRS) used in
this study were constructed of human-tissue-equivalent materials to represent
the configuration and size of a typical 5-year-old child. Before exposure in
the CT scanner, 20 MOSFET dosimeters were deployed throughout the phantom eye
lens well, retinal well (eye globe), brain, chest, and thyroid
(Fig. 1). The eye shield was
fabricated from a double layer of bismuth-impregnated latex (1.7 g
Bi/cm2, equivalent to 0.45 mg/cm3 of lead) that was
placed on top of a 1.0-cm-thick radiolucent foam step-off pad.
The anthropomorphic phantoms were scanned using a commercially available
16-MDCT scanner (LightSpeed, GE Healthcare). Both axial brain (5.0-mm-thick
slices, 140 mAs, 140 kVp) and helical craniofacial (2.5-mm-thick slices; 170
mAs; 140, 120, and 100 kVp) pediatric protocols were performed. Each
mAskVp combination was imaged three times and averaged with and without
the eye shield in position.

View larger version (60K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3A CT images of pediatric anthropomorphic phantom. Axial bone
window setting image of bismuth shield (high density, ventral to orbits) and
of step-off pad (intermediate density, ventral to orbits).
|
|

View larger version (40K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3B CT images of pediatric anthropomorphic phantom. Because
significant artifact was not present using conventional bone window setting
(A), windows were artificially widened to show artifact
(arrows). Bismuth shield with step-off pad successfully deflected
artifact away from regions of anatomic interest.
|
|
Results
To determine the dose reduction due to bismuth shielding, the following
formula was used: (NS S) / NS x 100,
in which NS represented the unshielded dose, and S
represented the shielded dose, both measured in cGy. Using the craniofacial
protocol, the average dose reduction to the eye as a result of bismuth
shielding was 39%, 42%, and 41% at 140, 120, and 100 kVp, respectively
(Fig. 2). The lens of the eye
experienced a smaller dose reduction at 36%, 25%, and 45% at 140, 120, and 100
kVp, respectively, again using the craniofacial protocol
(Fig. 2).
The axial head protocol at 140 kVp revealed a radiation dose reduction from
4.6 (unshielded) to 2.8 cGy (shielded) for the orbit and a radiation dose
reduction from 2.9 (unshielded) to 2.1 cGy (shielded) for the lens. Therefore,
the dose reduction at 140 kVp was 39% for the orbit and 28% for the lens.
Discussion
Other studies, using traditional approaches, have looked at the utility of
orbital shielding and found similar dose reduction of approximately 40%
[46].
However, one study, by McLaughlin and Mooney
[7], used an angled gantry to
exclude direct exposure to the eyes and found that bismuth shielding reduced
the radiation dose by only 18%. Another study found that when taken to an
extreme and optimized gantry angles were used, the dose reduction due to
bismuth shielding was less than 25%
[5]. Therefore, it has been
suggested that it should be routine practice to avoid the eye when possible in
head CT and align the scan with the supraorbital meatal baseline
[8].
Exclusion of the orbits as a means of reducing the dose, however, is
impractical. Yeoman et al. [8]
reported that only 32% of radiology centers surveyed routinely avoided the
orbits in head CT examinations. Even if all centers used a policy to configure
pediatric CT along the supraorbital meatal baseline, axial scanning would
still be necessary to evaluate a child for craniosynostosis or any orbital,
sinus, or mastoid problems. Therefore, excluding the orbit would not be
possible, and bismuth shielding would be helpful to reduce the dose from
radiation exposure.
Concern about artifact has limited the usage of bismuth shielding; however,
in most instances, this artifact is minimal
[47].
Using a step-off pad, the artifact from the bismuth shields was not evident on
the initial scan and only became visible after artificially widening the
window settings of the image. The step-off pads below the bismuth shields
moved the artifact to a more ventral location outside the area of diagnostic
interest (Fig. 3A,
3B).
Because all new CT scanners use automatic tube current modulation as a tool
to decrease radiation dose, it is important to evaluate the effect bismuth
shielding might have on automatic current modulation. However, to the best of
our knowledge, there has been no systematic evaluation of bismuth shields for
in-plane shielding using automatic tube current modulation in patients. The
method of tube current modulation varies depending on the manufacturer
[9]. For example, one form of
tube current modulation modulates the tube current based on regional changes
in density assessed on the topogram (scout image). In this setting, placement
of a shield could arguably offset both the dose reduction through the shield
and the benefits of tube current modulation because the increased density
through the shielded region seen on the topogram would result in an increase
in tube current. Preliminary investigation of this phenomenon has shown that
placement of the shield after the topogram has been obtained reduces this
effect (Frush DP, unpublished data).
In summary, we found that the use of a bismuth shield with a step-off pad
significantly reduces the radiation dose to the eye by 42%. This dose
reduction can be reliably detected by a MOSFET dosimeter in an MDCT scanner,
which is important because the rapid development of scanner technology has
resulted in the lack of development and validation of mathematic models for
dose estimation for all scanner types. Given the sensitivity of the pediatric
eye to radiation exposure, bismuth shields should be used in pediatric
patients when the orbits are included in the CT examination.
Acknowledgments
We thank Carolyn Lowry for technical support. In addition, this study was
performed as part of a senior-year engineering project by Jesse Riley and Jena
Jamal, students in the Pratt School of Engineering at Duke University. Giao
Nguyen and Greta Toncheva of the radiation safety office played key roles in
the data acquisition.
References
- Task Group on Control of Radiation Dose in Computed Tomography.
Managing patient dose in computed tomography: a report of the International
Commission on Radiological Protection. Ann ICRP2000; 30:7
45[Medline]
- Merriam GR Jr, Focht EF. A clinical study of radiation cataracts
and the relationship to dose. Am J Roentgenol Radium Ther Nucl
Med 1957; 77:759
785[Medline]
- Brenner DJ. Is it time to retire the CTDI for CT quality assurance
and dose optimization? Med Phys 2005;32
:3225
3226[CrossRef][Medline]
- Colombo P, Pedroli G, Nicoloso M, Re S, Valvassori L, Vanzulli A.
Evaluation of the efficacy of a bismuth shield during CT examinations.
Radiol Med (Torino) 2004;108
: 560568[Medline]
- Perisinakis K, Raissaki M, Theocharopoulos N, Damilakis J,
Gourtsoyiannis N. Reduction of eye lens radiation dose by orbital bismuth
shielding in pediatric patients undergoing CT of the head: a Monte Carlo
study. Med Phys 2005;32
:1024
1030[CrossRef][Medline]
- Hopper KD, Neuman JD, King SH, Kunselman AR. Radioprotection to the
eye during CT scanning. Am J Neuroradiol2001; 22:1194
1198[Abstract/Free Full Text]
- McLaughlin DJ, Mooney RB. Dose reduction to radiosensitive tissues
in CT: do commercially available shields meet the users' needs?
Clin Radiol 2004;59
: 446450[CrossRef][Medline]
- Yeoman LJ, Howarth L, Britten A, Cotterill A, Adam EJ. Gantry
angulation in brain CT: dosage implications, effect on posterior fossa
artifacts, and current international practice.
Radiology 1992;184
: 113116[Abstract/Free Full Text]
- McCollough CH, Bruesewitz MR, Kofler JM Jr. CT dose reduction and
dose management tools: overview of available options.
RadioGraphics 2006;26
: 503512[Abstract/Free Full Text]

CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
F. A. Mettler Jr, W. Huda, T. T. Yoshizumi, and M. Mahesh
Effective Doses in Radiology and Diagnostic Nuclear Medicine: A Catalog
Radiology,
July 1, 2008;
248(1):
254 - 263.
[Abstract]
[Full Text]
[PDF]
|
 |
|