DOI:10.2214/AJR.07.2069
AJR 2007; 189:197-203
© American Roentgen Ray Society
Sixty-Four Slice CT Evaluation of Aortic Stenosis Using Planimetry of the Aortic Valve Area
Gudrun M. Feuchtner1,
Silvana Müller2,
Johannes Bonatti3,
Thomas Schachner3,
Corinna Velik-Salchner4,
Otmar Pachinger2 and
Wolfgang Dichtl2
1 Clinical Department of Radiology II, Innsbruck Medical University,
Anichstrasse 35, A-6020 Innsbruck, Austria.
2 Clinical Department of Cardiology, Innsbruck Medical University, Innsbruck,
Austria.
3 Clinical Department of Cardiac Surgery, Innsbruck Medical University,
Innsbruck, Austria.
4 Clinical Department of Anesthesia and Intensive Care Medicine, Innsbruck
Medical University, Innsbruck, Austria.
Received October 24, 2006;
accepted after revision March 26, 2007.
Address correspondence to G. M. Feuchtner
(gudrun.feuchtner{at}i-med.ac.at).
Abstract
OBJECTIVE. The purpose of our study was to evaluate planimetry of
the aortic valve area with 64-slice CT in comparison with transthoracic
echocardiography (TTE) and transesophageal echocardiography (TEE) in patients
with aortic stenosis.
MATERIALS AND METHODS. Thirty-six patients with aortic valve disease
referred for coronary 64-slice CT angiography were examined. Planimetry of the
aortic valve area with 64-slice CT was compared with TTE using the Doppler
continuity equation for calculation of the aortic valve area and with
planimetric measurement of the aortic valve area using TEE.
RESULTS. Planimetry of the aortic valve area with CT (1.11 ±
0.42 cm2) showed a good correlation with TTE (1.05 ± 0.42
cm2) (r = 0.88, p < 0.001) in 32 patients and
a good correlation with TEE (1.41 ± 1.61 cm2) (r =
0.99, p < 0.0001) in 10 patients. The mean and maximum
transvalvular pressure gradients were correlated with the aortic valve area as
measured with CT (r = -0.68, p = 0.0001; and r =
-0.67, p = 0.0001, respectively). Beta-blockers were not given (mean
heart rate, 62.5 ± 10.7 beats per minute).
CONCLUSION. MDCT allows accurate planimetry of the aortic valve area
in patients with aortic stenosis. In patients referred for 64-slice CT
coronary angiography, concomitant aortic stenosis can be identified and
evaluated.
Keywords: aortic stenosis echocardiography 64-slice CT planimetry
Introduction
Coronary angiography with 64-slice CT
[1,
2] is increasingly used in
clinical practice for the exclusion of coronary artery disease. Degenerative
aortic stenosis is frequently present in patients with coronary artery disease
because the pathologic mechanisms are similar
[3]; therefore, the
simultaneous evaluation of aortic valve disease would be desirable in patients
referred for CT coronary angiography (Figs.
1A and
1B).

View larger version (103K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 1A 64-slice CT of coronary arteries in 75-year-old man. MDCT
allows simultaneous evaluation of coronary arteries (A) and aortic
valve (B), shown in 3D by applying volume-rendering technique. Note
that severe calcification (white spots) of both aortic valve and
coronary arteries is frequently seen. LCA = left coronary artery, RCA = right
coronary artery.
|
|

View larger version (141K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 1B 64-slice CT of coronary arteries in 75-year-old man. MDCT
allows simultaneous evaluation of coronary arteries (A) and aortic
valve (B), shown in 3D by applying volume-rendering technique. Note
that severe calcification (white spots) of both aortic valve and
coronary arteries is frequently seen. LCA = left coronary artery, RCA = right
coronary artery.
|
|
Degenerative aortic stenosis is routinely assessed in clinical practice
with transthoracic echocardiography (TTE) on the basis of measurement of the
aortic valve orifice area (aortic valve area) in addition to the transvalvular
pressure gradients [4]. 16-MDCT
has already shown promising results in the planimetric measurement of the
aortic valve area in patients with aortic stenosis
[5,
6]. However, technical
limitations of 16-MDCT have been observed, such as time-consuming
postprocessing [5] and motion
artifacts in patients with high heart rates (> 65 beats per minute [bpm])
caused by limited temporal resolution (
105-250 milliseconds)
[7]. Therefore, premedication
with ß-blockers has been required to achieve images without motion
artifacts. However, the use of ß-blockers is of concern in patients with
aortic stenosis. The relatively new 64-slice CT technology provides higher
temporal resolution (83-166 milliseconds)
[8], which might improve image
quality at higher heart rates.
The aim of our study was to evaluate planimetry of the aortic valve area
with 64-slice CT in patients with aortic stenosis without using
ß-blockers in comparison with measurement of the aortic valve area using
the Doppler continuity equation with TTE and planimetric calculation of the
aortic valve area with transesophageal echocardiography (TEE).
Materials and Methods
Study Population
Thirty-six patients (Table
1) with known aortic valve disease were examined between November
2005 and September 2006. Seventeen patients before aortic valve surgery (AVS)
and 19 patients from the Austrian "TASS Study" population (a
prospective, randomized, double-blind study assessing the effect of statins on
the progression of aortic stenosis) underwent 64-slice CT coronary angiography
for the evaluation of coronary artery disease. Our institutional review board
approved both studies. Written informed consent was obtained from all
patients. Patient exclusion criteria were renal dysfunction (creatinine level
> 1.2 mg/dL), known allergy to iodine contrast material, hyperthyroidism,
left ventricular dysfunction (left ventricular ejection fraction < 35%
severe aortic regurgitation (> grade 2+), previous myocardial infarction,
cardiomyopathy, pregnancy, plasmocytoma, and multiple myeloma.
CT Technique
CT data were acquired using an MDCT scanner (Sensation 64, Siemens Medical
Solutions) with a 32-row detector collimation acquiring 64 x 0.6 mm
slices by applying the z-axis flying-focus technique, a table
translation speed of 3.8 mm per rotation, and a gantry rotation time of 0.33
second. Tube current of 120 kV and 800-900 mAs resulted in an estimated
radiation exposure range of 9.4-14.8 mSv (mean, 11 mSv)
[9]. The scanning direction was
craniocaudal during a single inspiratory breath-hold, and the ECG signal was
recorded simultaneously. A bolus of 90-120 mL of iodixanol, a nonionic iodine
contrast agent, at a concentration of 320 mg I/dL (Visipaque, Amersham) was
injected IV into an antecubital vein using a 20-gauge cannula at a flow rate
of 5 mL/s using an automated injector. Scanning was started automatically by
applying a bolus-tracking technique (ascending aorta; threshold, 100 H) as
previously described for coronary CT angiography
[10].
CT Image Reconstruction and Analysis
Using retrospective ECG gating, a data set containing axial slices
(effective slice width, 1.0 mm; 70% overlap) was reconstructed at 5% steps
during the cardiac cycle. This data set was reviewed in 4D using dedicated
software (4D Inspace, Heart View, Siemens) on an external workstation
(Leonardo, Siemens), and the time point of maximal aortic valve opening was
identified. A series of axial slices was generated at the time point of
maximal aortic valve opening with an effective slice width of 0.75 mm (60%
overlap), a medium-smooth convolution kernel (B 25 f ++), and an image matrix
of 512 x 512 pixels.
This image series was reviewed in 3D by applying multiplanar reformations
(MPRs). The cross-sectional planimetric images of the aortic valve were
reconstructed ranging from the top of the leaflets to the infundibulum. The
smallest measurable aortic valve area was regarded as the effective aortic
valve area. Two independent reviewers circled the effective aortic valve area
with a digital caliper. Image quality was graded on a 3-point scale as 1, good
(no artifacts); 2, acceptable (mild artifacts but acceptable delineation of
the aortic valve area); or 3, inadequate (artifacts without continuous
delineation of the aortic valve area).
Transthoracic Echocardiography
All measurements were performed using a standard sonographic system
(Sequoia 256, Acuson-Siemens Medical Solutions) equipped with a 3.5-1.75-MHz
transducer by an experienced observer. Doppler flow data were acquired from
the left ventricular outflow tract (LVOT) and included LVOT velocity
measurement using pulsed wave Doppler sonography and LVOT diameter. The peak
transvalvular velocity was measured in all patients. In addition, the aortic
valve area was calculated using the continuity equation approach with a
Doppler velocity-time integral as in the study of Dumesnil et al.
[11]. The mean and the peak
transvalvular pressure gradients were calculated. Left ventricular
end-diastolic volume, left ventricular end-systolic volume, and left
ventricular ejection fraction were determined using Simpson's method
[12].

View larger version (120K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2A Planimetry of aortic valve area (AVA) performed after
identifying maximal aortic valve opening using 4D dynamic imaging (see Fig.
S1, cine CT, at
www.ajronline.org).
MDCT images show tricuspid valve in 79-year-old woman with moderate aortic
stenosis (aortic valve area, 1.1 cm2) (A and
C) and bicuspid valve in 53-year-old-man with severe aortic
stenosis (aortic valve area, 0.98 cm2) (B and D).
A and B were reconstructed with multiplanar reformations,
C and D with volume rendering using 1-mm slab.
|
|

View larger version (135K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2B Planimetry of aortic valve area (AVA) performed after
identifying maximal aortic valve opening using 4D dynamic imaging (see Fig.
S1, cine CT, at
www.ajronline.org).
MDCT images show tricuspid valve in 79-year-old woman with moderate aortic
stenosis (aortic valve area, 1.1 cm2) (A and
C) and bicuspid valve in 53-year-old-man with severe aortic
stenosis (aortic valve area, 0.98 cm2) (B and D).
A and B were reconstructed with multiplanar reformations,
C and D with volume rendering using 1-mm slab.
|
|

View larger version (103K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2C Planimetry of aortic valve area (AVA) performed after
identifying maximal aortic valve opening using 4D dynamic imaging (see Fig.
S1, cine CT, at
www.ajronline.org).
MDCT images show tricuspid valve in 79-year-old woman with moderate aortic
stenosis (aortic valve area, 1.1 cm2) (A and
C) and bicuspid valve in 53-year-old-man with severe aortic
stenosis (aortic valve area, 0.98 cm2) (B and D).
A and B were reconstructed with multiplanar reformations,
C and D with volume rendering using 1-mm slab.
|
|

View larger version (95K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2D Planimetry of aortic valve area (AVA) performed after
identifying maximal aortic valve opening using 4D dynamic imaging (see Fig.
S1, cine CT, at
www.ajronline.org).
MDCT images show tricuspid valve in 79-year-old woman with moderate aortic
stenosis (aortic valve area, 1.1 cm2) (A and
C) and bicuspid valve in 53-year-old-man with severe aortic
stenosis (aortic valve area, 0.98 cm2) (B and D).
A and B were reconstructed with multiplanar reformations,
C and D with volume rendering using 1-mm slab.
|
|

View larger version (9K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3A Planimetry of aortic valve area with 64-slice CT versus
transthoracic echocardiography (TTE) using continuity equation for calculation
of aortic valve area (in cm2) with Doppler velocity-time integral
in 32 patients. Linear regression analysis illustrates good correlation
between both imaging techniques.
|
|

View larger version (11K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3B Planimetry of aortic valve area with 64-slice CT versus
transthoracic echocardiography (TTE) using continuity equation for calculation
of aortic valve area (in cm2) with Doppler velocity-time integral
in 32 patients. Bland-Altman plot implies good intertechnique agreement.
|
|

View larger version (9K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3C Planimetry of aortic valve area with 64-slice CT versus
transthoracic echocardiography (TTE) using continuity equation for calculation
of aortic valve area (in cm2) with Doppler velocity-time integral
in 32 patients. Planimetry of aortic valve area with CT versus TEE (n
= 10) shows high concordance on Bland-Altman plot.
|
|
Transesophageal Echocardiography
Planimetry of the aortic valve area was performed by an experienced
reviewer with a multiplane phased-array 2-7-MHz TEE transducer on a sonography
unit (iE33 echocardiography system, Philips Medical Systems) before aortic
valve surgery.
Statistical Analysis
Statistical analysis was performed using SSPS software, version 8.0. The
correlations between the aortic valve area measured by CT, TTE, and TEE and
the transvalvular pressure gradient were determined using linear regression
analysis and the Pearson's correlation coefficient. A two-tailed probability
value of less than 0.05 was considered statistically significant. Bland-Altman
analysis [13] was performed to
evaluate intertechnique agreement by plotting the difference in aortic valve
area between CT and TTE against aortic valve area averages. The mean of the
difference with a bias of ± 1.96 SD denotes the limits of agreement.
Interobserver variability of aortic valve area and left ventricular functional
measurements was computed as a percentage of the mean differences between the
corresponding observations divided by the average of all observations.
Results
Results for planimetry of the aortic valve area by 64-slice CT (Figs.
2A,
2B,
2C, and
2D) versus TTE and TEE are
shown in Figures 3A,
3B, and
3C.
Linear regression analysis showed a good correlation between planimetry of
the aortic valve area by 64-slice CT (1.11 ± 0.42 [SD] cm2)
and the aortic valve area as measured with TTE (1.05 ± 0.42
cm2; r = 0.88; p < 0.001) in 32 patients
(Fig. 3A). The Bland-Altman
plot implies a good intertechnique concordance between CT and TTE, placing 30
of 32 patients between the limits of agreement (0.47-0.35)
(Fig. 3B). The mean pressure
gradient (36.3 ± 17.1 mm Hg) and the maximal pressure gradient (56.8
± 23.7 mm Hg) correlated significantly with the aortic valve area as
quantified with 64-slice CT (r = -0.68, p = 0.0001; and
r = -0.67, p = 0.0001, respectively) (Figs.
4A and
4B). Planimetry of the aortic
valve area with TEE (1.41 ± 1.61 cm2) showed a high
correlation with 64-slice CT (r = 0.99, p < 0.0001) in 10
patients, and the Bland-Altman plot confirmed good concordance
(Fig. 3C).

View larger version (10K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4A Planimetry of aortic valve area with CT versus transthoracic
echocardiography (TTE). Planimetry of aortic valve area on CT versus mean
(A) and maximum (B) transvalvular pressure gradients on TTE show
moderate correlation.
|
|

View larger version (10K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4B Planimetry of aortic valve area with CT versus transthoracic
echocardiography (TTE). Planimetry of aortic valve area on CT versus mean
(A) and maximum (B) transvalvular pressure gradients on TTE show
moderate correlation.
|
|
The mean interobserver variability of the aortic valve area measurement on
CT was 4.8% (absolute variability, 0.51 ± 0.03 cm2). The
mean postprocessing time was 6.3 minutes (range, 4.1-10.8 minutes) for the
physician. Images were reconstructed at 12% of the R-R interval in 34 of 36
patients and at 10% and 16% in one patient each.
The mean heart rate was 62.5 ± 10.7 bpm (range, 42-88 bpm) during
the CT scanning. None of the patients received ß-blockers. Thirty-four
patients were in sinus rhythm, two of whom had multiple premature beats
requiring ECG editing; one additional patient had atrial fibrillation and
another patient had a right ventricular pacemaker. Image quality was good in
29 patients, acceptable in seven (reasons for mild artifacts were atrial
fibrillation, a pacemaker, and premature beats), and insufficient in none of
the patients.
Discussion
Diagnostic imaging of degenerative aortic stenosis based on planimetry of
the aortic valve area using 16-MDCT has been previously reported
[5,
6]. However, postprocessing was
time-consuming and ß-blockers were required
[5] in up to 60% of patients
referred for coronary CT angiography
[14] to achieve appropriate
image quality without motion artifacts in patients with high heart rates (>
65 bpm). In general, ß-blockers are contraindicated in patients with
aortic stenosis because of their negative influence on hemodynamics and
cardiac output.
In contrast to a previous study using 16-MDCT
[5], this study shows that
64-slice CT allows accurate planimetry of the aortic valve area (Figs.
5A,
5B,
5C, and
5D) and good correlation with
both TTE and TEE without using ß-blockers. This might be explained by
improved temporal resolution, which is achieved by faster gantry rotation
times, thus leading to a reduction of motion artifacts at higher heart rates.
In addition, our data suggest that image quality may be improved when compared
with a previously published study using 16-MDCT
[5]. Postprocessing is
facilitated and less time-consuming when using new software that enables 4D
imaging, which permits a dynamic display of valvular motion (cine CT) during
the entire cardiac cycle. This dynamic display is based on stepwise image
reconstruction at arbitrary percentages (e.g., every 5% of the R-R interval)
(Fig. S1, cine CT, available at
www.ajronline.org)
that can be applied for both 16- and 64-slice CT data sets
[15,
16]. When the clips are
reviewed, the time point of the aortic valve opening can be identified; after
the valve is opened, it stays open until the end of systole and allows imaging
of the valve orifice. As the result, the aortic valve area could be accurately
quantified at 12% of the R-R interval in most of our patients.

View larger version (102K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5B Planimetry of aortic valve areas (AVA) in different shapes.
Stenotic triangular aortic valve (area, 0.81 cm2) in 70-year-old
man with severe aortic stenosis and severe calcification (white
spots) of tricuspid valve.
|
|

View larger version (85K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5D Planimetry of aortic valve areas (AVA) in different shapes.
Spotlike aortic valve (area, 0.55 cm2) in 66-year-old woman with
symptomatic, critical severe aortic stenosis of functional bicuspid valve in
whom surgery is indicated.
|
|
The main advantage of 64-slice CT over TTE is that CT allows a measurement
of the "true" anatomic aortic valve orifice area (so-called
planimetry), similar to TEE and cardiac MRI. Therefore, measurement of the
aortic valve area with 64-slice CT is not adversely influenced by hemodynamics
such as left ventricular dysfunction and severe aortic regurgitation, which
are limiting factors in the assessment of aortic stenosis on TTE based on
Doppler flow parameters (such as the continuity equation and the transvalvular
pressure gradients). Therefore, 64-slice CT might be considered a
comprehensive imaging technique under those conditions.
Our study has some limitations. Radiation exposure of cardiac 64-slice CT
scans ranges between 9.4 and 14.8 mSv (mean, 11 mSv)
[8] to a maximum of 21 mSv
[2], which is greater than the
range of cardiac catheterization
[17] but comparable to a
myocardial SPECT scan (
20 mSv)
[18]. ECG dose modulation,
which reduces the radiation exposure by approximately 45-48%
[19], is not recommended in
this specific setting because the tube output is reduced during systole, thus
increasing image noise (e.g., if planimetry of the aortic valve area is
required in patients with limited echocardiographic windows).
In patients undergoing coronary CT angiography, unenhanced CT calcium score
scanning, which has previously often been performed, can be used to determine
the presence and severity of aortic valve calcification
[20,
21]. In patients with aortic
valve calcification, ECG dose modulation should not be applied. Aortic valve
stenosis commonly develops in older patients (> 65 years)
[3]; and available data suggest
that a radiation dose received during CT may not significantly increase the
lifetime risk of cancer in this age group
[22]. However, in younger
patients (< 60 years), the radiation dose from cardiac CT when ECG pulsing
is not used is a matter of concern, and the benefit from cardiac CT should be
carefully considered for each individual.
Iodine contrast agents should not be administered to patients with renal
dysfunction, known allergies, and hyperthyroidism. The maximal heart rate in
our study population was 88 bpm; therefore, the performance of 64-slice CT at
heart rates greater than 88 bpm remains unknown. The low heart rate (mean,
62.5 bpm) in our study population in whom ß-blockers were not
administered might be explained by the unique lifestyle of our population, who
live in an alpine region (Innsbruck, Austrian Alps) characterized by hiking
(the equivalent of endurance training). We acknowledge also that the small
number of investigated patients may lead to a bias.
In conclusion, TEE will remain the primary diagnostic imaging technique for
assessing patients with aortic stenosis in clinical practice because it is
widely available, radiation-free, and reliable. However, planimetry of the
aortic valve area with 64-slice CT is accurate and has been improved;
pretreatment with ß-blockers can be avoided in patients with heart rates
up to 88 bpm. Postprocessing is facilitated using dedicated software that
enables 4D cine imaging of valvular motion. With respect to the emerging use
of 64-slice CT coronary angiography in outpatients for the evaluation of
coronary artery disease, planimetry of the aortic valve area is recommended in
all patients who present with previously unknown aortic valve calcification to
distinguish between nonstenotic aortic valve sclerosis and aortic
stenosis.
References
- Raff GL, Gallagher MJ, O'Neill WW, et al. Diagnostic accuracy of
noninvasive coronary angiography using 64-slice spiral computed tomography.
J Am Coll Cardiol 2005;46
: 552-557[Abstract/Free Full Text]
- Mollet N, Cademartiri F, Mieghem C, et al. High-resolution spiral
computed tomography coronary angiography in patients referred for diagnostic
conventional coronary angiography. Circulation2005; 112:2318
-2323[Abstract/Free Full Text]
- Steward BF, Siscovick D, Lind BK, et al. Clinical factors with
calcific aortic valve disease. J Am Coll Cardiol1997; 29:630
-634[Abstract]
- Bonow R, Carabello B. Guidelines for the management of patients
with valvular heart disease. ACC/AHA guidelines.
Circulation 1998;98
: 1949-1984[Free Full Text]
- Feuchtner GM, Dichtl W, Friedrich GJ, et al. Multislice computed
tomography for detection of patients with aortic valve stenosis and
quantification of severity. J Am Coll Cardiol2006; 47:1410
-1417[Abstract/Free Full Text]
- Alkadhi H, Wildermuth S, Plass A, et al. Aortic stenosis:
comparative evaluation of 16-detector row CT and echocardiography.
Radiology 2006;240
: 47-55[Abstract/Free Full Text]
- Flohr T, Bruder H, Stierstorfer K, et al. New technical
developments in multislice CT. Part 2. Submillimeter 16-slice scanning and
increased gantry rotation speed for cardiac imaging. Rofo Fortschr
Geb Roentgenstr 2002; 174:1022
-1027[CrossRef]
- Flohr T, Stierstorfer K, Raupach R, et al. Performance evaluation
of a 64-MDCT system with z-flying focal spot.
Rofo 2004; 176:1803
-1810[Medline]
- Hausleiter J, Meyer T, Hadamitzky M, et al. Radiation dose
estimates from cardiac multislice computed tomography in daily practice:
impact of different scanning protocols on effective dose estimates.
Circulation 2006;113
: 1305-1310[Abstract/Free Full Text]
- Cademartiri F, Nieman K, van der Lugt A, et al. Intravenous
contrast material administration at 16-detector row helical CT coronary
angiography: test bolus versus bolus-tracking technique.
Radiology 2004;233
: 817-823[Abstract/Free Full Text]
- Dumesnil JG, Honos GN, Lemieux A, et al. Validation and application
of indexed aortic prosthetic valve areas calculated by Doppler
echocardiography. J Am Coll Cardiol 1990;16
: 637-643[Abstract]
- Schiller NB, Shah PM, Crawford M, et al. Recommendations for
quantitation of the left ventricle by two-dimensional echocardiography.
American Society of Echocardiography Committee on Standards, Subcommittee on
Quantitation of Two-Dimensional Echocardiograms. Am Soc
Echocardiogr 1989; 2:358
-367
- Bland JM, Altman DG. Statistical methods for assessing the
agreement between two methods of clinical measurement.
Lancet 1986; 1:307
-310[CrossRef][Medline]
- Mollet NR, Cademartiri F, Nieman K, et al. Multislice spiral
computed tomography coronary angiography in patients with stable angina
pectoris. J Am Coll Cardiol 2004;43
: 2265-2270[Abstract/Free Full Text]
- Muhlenbruch G, Das M, Hohl C, et al. Global left ventricular
function in cardiac CT: evaluation of an automated 3D region-growing
segmentation algorithm. Eur Radiol 2006;16
: 1117-1123[CrossRef][Medline]
- Vogel-Claussen J, Pannu H, Spevak PJ, Fishman EK, Bluemke DA.
Cardiac valve assessment with MR imaging and 64-section multi-detector row CT.
RadioGraphics 2006;26
: 1769-1784[Abstract/Free Full Text]
- Stisova V. Effective dose to patient during cardiac interventional
procedures. Radiat Prot Dosimetry 2004;111
: 271-274[Abstract/Free Full Text]
- Picano E. Economic and biological costs of cardiac imaging.
Cardiovasc Ultrasound 2005;25
: 13
- Jakobs TF, Becker CR, Ohnesorge B, et al. Multislice helical CT of
the heart with retrospective ECG gating: reduction of radiation exposure by
ECG-controlled tube current modulation. Eur Radiol2002; 12:1081
-1086[CrossRef][Medline]
- Morgan-Hughes GJ, Owens PE, Roobottom CA, Marshall AJ. Three
dimensional volume quantification of aortic valve calcification using
multislice computed tomography. Heart2003; 89:1191
-1194[Abstract/Free Full Text]
- Shavelle DM, Budoff MJ, Buljubasic N, et al. Usefulness of aortic
valve calcium scores by electron beam computed tomography as a marker for
aortic stenosis. Am J Cardiol 2003;92
: 349-353[CrossRef][Medline]
- Brenner DJ. Radiation risks potentially associated with low-dose CT
screening of adult smokers for lung cancer. Radiology2004; 231:440
-445[Abstract/Free Full Text]

CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
E. J. Halpern, R. Mallya, M. Sewell, M. Shulman, and D. R. Zwas
Differences in Aortic Valve Area Measured with CT Planimetry and Echocardiography (Continuity Equation) Are Related to Divergent Estimates of Left Ventricular Outflow Tract Area
Am. J. Roentgenol.,
June 1, 2009;
192(6):
1668 - 1673.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Akhtar, E. M. Tuzcu, S. R. Kapadia, L. G. Svensson, R. K. Greenberg, E. E. Roselli, S. Halliburton, V. Kurra, P. Schoenhagen, and S. Sola
Aortic root morphology in patients undergoing percutaneous aortic valve replacement: Evidence of aortic root remodeling
J. Thorac. Cardiovasc. Surg.,
April 1, 2009;
137(4):
950 - 956.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. J. Cawley, J. H. Maki, and C. M. Otto
Cardiovascular Magnetic Resonance Imaging for Valvular Heart Disease: Technique and Validation
Circulation,
January 27, 2009;
119(3):
468 - 478.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
P. J. de Feyter, S. Achenbach, and K. Nieman
CHAPTER 6 Cardiovascular Computed Tomography
ESC Textbook of Cardiovascular Medicine,
January 1, 2009;
2(1):
med-9780199566990-chapter - med-9780199566990-chapter.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
T. M. LaBounty, B. Sundaram, P. Agarwal, W. A. Armstrong, E. A. Kazerooni, and E. Yamada
Aortic Valve Area on 64-MDCT Correlates with Transesophageal Echocardiography in Aortic Stenosis
Am. J. Roentgenol.,
December 1, 2008;
191(6):
1652 - 1658.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
S. Schroeder, S. Achenbach, F. Bengel, C. Burgstahler, F. Cademartiri, P. de Feyter, R. George, P. Kaufmann, A. F. Kopp, J. Knuuti, et al.
Cardiac computed tomography: indications, applications, limitations, and training requirements: Report of a Writing Group deployed by the Working Group Nuclear Cardiology and Cardiac CT of the European Society of Cardiology and the European Council of Nuclear Cardiology
Eur. Heart J.,
February 2, 2008;
29(4):
531 - 556.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
L. F. Tops, S. C. Krishnan, J. D. Schuijf, M. J. Schalij, and J. J. Bax
Noncoronary applications of cardiac multidetector row computed tomography.
J. Am. Coll. Cardiol. Img.,
January 1, 2008;
1(1):
94 - 106.
[Abstract]
[Full Text]
[PDF]
|
 |
|