DOI:10.2214/AJR.07.2501
AJR 2007; 189:1238-1242
© American Roentgen Ray Society
Quantification and Detectability of In-Stent Stenosis with CT Angiography and MR Angiography in Arterial Stents In Vitro
Melanie B. Blum1,
Maria Schmook,
Rüdiger Schernthaner,
Gundula Edelhauser,
Stefan Puchner,
Johannes Lammer and
Martin A. Funovics
1 All authors: Division of Angiography and Interventional Radiology, Department
of Radiology, Medical University of Vienna, Währinger Gürtel 18-20,
1090 Vienna, Austria.
Received September 25, 2006;
accepted after revision May 19, 2007.
Address correspondence to M. A. Funovics
(martin.funovics{at}meduniwien.ac.at).
Abstract
OBJECTIVE. The purpose of this study was to compare CT angiography
(CTA) and MR angiography (MRA) for the detectability of 75% and 95% stenoses
in phantoms using six different stents.
MATERIALS AND METHODS. Six different stents (Expander, Hemobahn,
SelfX, Smart, Symphony, and Wallstent) were inserted into tubes filled with
contrast agent (ioversol or gadoteric acid). To mimic stenoses of 75% and 95%
of the patent lumen, 8-mm-diameter nylon cylinders were bored in the central
axis (2 mm and 4 mm, respectively) and placed into the stent lumen. Intensity
profiles across stenoses on 2-mm coronal reformatted sections of CTA or MRA
were compared, and the detectability of the residual lumen was assessed using
a subjective score.
RESULTS. CTA showed relative in-stent signal attenuation for the
in-stent stenoses of the tested stents ranging from 75% to 100% of the signal
intensity of the control. SelfX and Symphony showed further shading of the
residual lumen due to beam-hardening artifacts. Overestimation of stenosis was
associated with low-grade stenoses in which the border of the lumen was closer
to the stent struts. MRA showed relative in-stent signal attenuation of the
in-stent stenoses ranging from 30% to 100% of the signal intensity of the
control. Strut thickness tended to correlate with higher attenuation at
CT.
CONCLUSION. CTA may be more suitable for differentiation between 95%
stenosis and occlusion; MRA has higher sensitivity in detecting 75% stenoses.
Strut thickness and mesh size did not prove to be significant predictors for
signal attenuation or overall image quality.
Keywords: CT in-stent restenosis in vitro MRI stenosis
Introduction
With the advent of MDCT and moving-table MRI acquisition, CT angiography
and MR angiography are approaching the diagnostic accuracy of conventional
angiography
[1–5].
CT and MR angiography can be performed on an outpatient basis and are
potentially more cost-effective than invasive methods. A further advantage is
the lower renal burden resulting from the lower nephrotoxicity of MR
angiographic contrast material
[6–8].
These techniques are gradually replacing conventional subtraction angiography
in the diagnostic assessment of vascular status and the planning of
interventional procedures. In the presence of metallic stents, however, the
diagnostic accuracy of CT and MR angiography is impaired
[9,
10]. CT angiography is prone
to beam-hardening effects, and MR angiography suffers from susceptibility
artifacts in the presence of metallic stents
[7].
Assessment after iliofemoral stent placement ideally must address two main
issues. First, hemodynamically relevant stenosis must be accurately detected
and differentiated from low-grade stenosis
[11]. Second, in cases of
severe luminal narrowing, high-grade stenosis must be differentiated from
complete occlusion [12].
Previous studies
[13–17]
have addressed stent artifacts and the detectability of in-stent restenosis on
CT and MR angiography. Few investigations, however, have addressed the
detectability of in-stent restenosis
[7], especially in the
important range of approximately 70–95% stenosis. In addition, some of
the more recently developed and frequently used stent types have not been
evaluated in this context. The purposes of this study were to use CT and MR
angiography to compare frequently used stents to determine on a subjective
scale the diagnosis, despite attenuation and artifact formation in the stent,
of two degrees of standardized stenosis in a phantom and to objectively
quantify the changes inside the stent through the use of signal intensity and
attenuation profiles.
Materials and Methods
Phantom and Stents
Six stents of types of materials and structures frequently used in the
management of peripheral arterial occlusive disease were placed individually
in synthetic polyethylene tubes with an inner diameter of 8.5 mm. Standardized
stenoses were made with solid nylon cylinders cut to 2 cm in length and bored
longitudinally to a central axis of 4 or 2 mm. In this study, degree of
stenosis was calculated on the basis of the area of the unobstructed lumen
versus the area of the bore inside the stenosis. Compared with the
unobstructed lumen inside the stents, which measured 8 mm in diameter, the
degree of stenosis was 75% for the 4-mm bore and approximately 95% for the
2-mm bore. The nylon cylinders were placed inside the stents and secured with
glue (Fig. 1). This test
arrangement made it possible to compare stenosis without stent, stent without
stenosis, and stent with stenosis. Tubes without stents but with nylon
cylinders were used as controls. The tubes were plugged on the ends with
silicone, and hypodermic needles were used to fill the tubes with contrast
agent diluted in saline solution.
Two pieces, one with 75% and one with 95% stenosis, of each of the
following stent types were used: nitinol with 1.7-mm2 mesh size and
165-µm strut thickness (Expander, Bolton Medical), nitinol with
3-mm2 mesh size and 170-µm strut thickness (Hemobahn, Prograft
Medical), nitinol with 4.5-mm2 mesh size and 245-µm strut
thickness (SelfX, Abbott Laboratories), nitinol with 2-mm2 mesh
size and 200-µm strut thickness (SMART, Cordis), nitinol with
15-mm2 mesh size and 235-µm strut thickness (Symphony, Boston
Scientific), and cobalt alloy with 1.7-mm2 mesh size and 195-µm
strut thickness (Wallstent, Boston Scientific).
CT Angiography
CT angiography was performed on a 16-MDCT scanner (Somatom Sensation 16,
Siemens Medical Solutions) with the following acquisition parameters: slice
collimation, 16 x 0.75 mm; rotation time, 0.5 second; table feed, 14
mm/rotation; matrix size, 512 x 512. Coronal images (i.e., parallel to
the table plane) were reconstructed with kernel B31 in increments of 1 and 2
mm. The tubes were filled with ioversol (Optiray 320 mg, Mallinckrodt Imaging)
diluted at a ratio of 1:50 to a final concentration of 6.5 mg I/mL. To
determine the concentration of contrast agent, we produced a dilution series
and imaged it with CT and MRI. The concentration that most closely resembled
typical attenuation or signal intensity was selected. Concentrations were
adjusted such that the physiologic concentration during clinical investigation
was closely approximated. Tubes were placed centrally on the gantry parallel
to the longitudinal axis of the scanner with a minimum distance of 5 cm
between the stents in the imaging plane to avoid additional artifacts.
Previous measurements did not show field inhomogeneities even though the
stents were surrounded by air and not embedded in a gelatinous solution.
MR Angiography
MR angiography (Gyroscan T10-NT Powertrak 3000, Philips Medical Systems)
was performed with a 3D spoiled gradient-recalled echo sequence (TR/TE,
4.51.45; flip angle, 35°; matrix size, 160 x 256). A coronal slab
with a total thickness of 96 mm was acquired, and coronal slices measuring 2
mm with 1-mm overlap were reconstructed. The tubes were filled with gadoterate
dimeglumine (Dotarem, Guerbet) diluted at a ratio of 1:200 to a final
concentration of 2.5 µmol/mL. To determine the concentration of contrast
agent, we performed MRI on a dilution series. We compared the in vitro signal
intensities with those of in vivo images and used the concentration that
closely approximated in vivo values. Tubes were placed centrally in the bore
parallel to the longitudinal axis of the magnet with at least 5 cm between
stents to avoid radiating artifacts.
Image Analysis
DICOM images were electronically transferred to a standard PC and analyzed
with freeware (Image J, National Institutes of Health). Mean pixel intensity
profiles were recorded through each stenosis perpendicular to the stent axis.
The dependent variable was the area under the intensity profile curve. We
measured apparent degree of stenosis by comparing the percentage of pixels
above a threshold signal intensity or attenuation in an artificial stenosis
with vessel bore. In addition, a subjective score of 0–4 was given to
each stenosis by two experienced radiologists, who were blinded to degree of
in-stent stenosis and stent used. The radiologists worked in consensus to
allow comparison between subjective and objective data. The subjective score
was an assessment of the detectability of degree of stenosis, that is, absence
of overestimation and underestimation of degree of stenosis. The true degree
of stenosis was known. The measured variables are in-stent signal intensity or
attenuation of unstenosed stent, average signal intensity or attenuation in
the remaining channel inside the stenosed segment, and subjective score of
stenosis detectability.
Results
CT Angiography
Depending on the type of stent, CT angiography showed relative in-stent
attenuation for the in-stent stenoses ranging from 75% to 100% of the
attenuation of the control standard (Table
1). The subjective and absolute scores of stents with 75% and 95%
stenoses are summarized in Figure
2A,
2B,
2C,
2D. Stenoses in stent 3
(nitinol with 4.5-mm2 mesh size and 245-µm strut thickness,
SelfX) and stent 5 (nitinol with 15-mm2 mesh size and 235-µm
strut thickness, Symphony) were the most difficult to detect. These stents
exhibited not only homogeneous attenuation in the lumen inside the stent but
also irregular high-attenuation artifacts due to beam hardening. Depending on
the artifacts induced by the mesh pattern, overestimation of in-stent stenosis
was more likely in the larger-bore cases of 75% stenosis, in which the patent
lumen was necessarily closer to the stent struts. In the presence of strong
artifacts, the relative subjective rating was therefore lower than the
objective score owing to inhomogeneous attenuation in the lumen. The overall
attenuation due to the presence of the stent in unstenosed vessels is shown in
Figure 3A. For both types of
stenoses, strut thickness was more relevant than mesh size in terms of higher
attenuation (stents 3 and 5). In a comparison of the three stents with the
highest attenuation with the three stents with the lowest attenuation, there
was a trend to higher mean strut thickness (220 vs 183 µm) in the stents
with high attenuation on CT angiography but not high signal intensity on MR
angiography. However, neither strut thickness nor mesh size proved to be a
significant predictor of in-stent attenuation or subjective image quality
(p > 0.13). (Figs.
2A and
2B).

View larger version (32K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2A —Vessel phantoms with 75% or 95% stenosis at CT angiography
and MR angiography. Bars represent attenuation of residual lumen
(y-axis) and subjective scores of visibility of residual lumen (light
gray indicates low subjective score; dark gray, high subjective score) inside
respective stents (x-axis) compared with attenuation of stenosis
without stent. Inserts show images with stenosis inside respective stent type.
1 = nitinol with 1.7-mm2 mesh size and 165-µm strut thickness
(Expander, Bolton Medical), 2 = nitinol with 3-mm2 mesh size and
170-µm strut thickness (Hemobahn, Prograft Medical), 3 = nitinol with
4.5-mm2 mesh size and 245-µm strut thickness (SelfX, Abbott
Laboratories), 4 = nitinol with 2-mm2 mesh size and 200-µm strut
thickness (SMART, Cordis), 5 = nitinol with 15-mm2 mesh size and
235-µm strut thickness (Symphony, Boston Scientific), 6 = cobalt alloy with
1.7-mm2 mesh size and 195-µm strut thickness (Wallstent, Boston
Scientific), 7 = standard, SE = standard error. Graph shows results for CT
angiography of 75% stenosis.
|
|

View larger version (28K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2B —Vessel phantoms with 75% or 95% stenosis at CT angiography
and MR angiography. Bars represent attenuation of residual lumen
(y-axis) and subjective scores of visibility of residual lumen (light
gray indicates low subjective score; dark gray, high subjective score) inside
respective stents (x-axis) compared with attenuation of stenosis
without stent. Inserts show images with stenosis inside respective stent type.
1 = nitinol with 1.7-mm2 mesh size and 165-µm strut thickness
(Expander, Bolton Medical), 2 = nitinol with 3-mm2 mesh size and
170-µm strut thickness (Hemobahn, Prograft Medical), 3 = nitinol with
4.5-mm2 mesh size and 245-µm strut thickness (SelfX, Abbott
Laboratories), 4 = nitinol with 2-mm2 mesh size and 200-µm strut
thickness (SMART, Cordis), 5 = nitinol with 15-mm2 mesh size and
235-µm strut thickness (Symphony, Boston Scientific), 6 = cobalt alloy with
1.7-mm2 mesh size and 195-µm strut thickness (Wallstent, Boston
Scientific), 7 = standard, SE = standard error. Graph shows results for CT
angiography of 95% stenosis.
|
|

View larger version (27K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2C —Vessel phantoms with 75% or 95% stenosis at CT angiography
and MR angiography. Bars represent attenuation of residual lumen
(y-axis) and subjective scores of visibility of residual lumen (light
gray indicates low subjective score; dark gray, high subjective score) inside
respective stents (x-axis) compared with attenuation of stenosis
without stent. Inserts show images with stenosis inside respective stent type.
1 = nitinol with 1.7-mm2 mesh size and 165-µm strut thickness
(Expander, Bolton Medical), 2 = nitinol with 3-mm2 mesh size and
170-µm strut thickness (Hemobahn, Prograft Medical), 3 = nitinol with
4.5-mm2 mesh size and 245-µm strut thickness (SelfX, Abbott
Laboratories), 4 = nitinol with 2-mm2 mesh size and 200-µm strut
thickness (SMART, Cordis), 5 = nitinol with 15-mm2 mesh size and
235-µm strut thickness (Symphony, Boston Scientific), 6 = cobalt alloy with
1.7-mm2 mesh size and 195-µm strut thickness (Wallstent, Boston
Scientific), 7 = standard, SE = standard error. Graph shows results for MR
angiography of 75% stenosis.
|
|

View larger version (21K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2D —Vessel phantoms with 75% or 95% stenosis at CT angiography
and MR angiography. Bars represent attenuation of residual lumen
(y-axis) and subjective scores of visibility of residual lumen (light
gray indicates low subjective score; dark gray, high subjective score) inside
respective stents (x-axis) compared with attenuation of stenosis
without stent. Inserts show images with stenosis inside respective stent type.
1 = nitinol with 1.7-mm2 mesh size and 165-µm strut thickness
(Expander, Bolton Medical), 2 = nitinol with 3-mm2 mesh size and
170-µm strut thickness (Hemobahn, Prograft Medical), 3 = nitinol with
4.5-mm2 mesh size and 245-µm strut thickness (SelfX, Abbott
Laboratories), 4 = nitinol with 2-mm2 mesh size and 200-µm strut
thickness (SMART, Cordis), 5 = nitinol with 15-mm2 mesh size and
235-µm strut thickness (Symphony, Boston Scientific), 6 = cobalt alloy with
1.7-mm2 mesh size and 195-µm strut thickness (Wallstent, Boston
Scientific), 7 = standard, SE = standard error. Graph shows results for MR
angiography of 95% stenosis.
|
|

View larger version (7K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3A —Graphs show stent-related relative attenuation or signal
intensity in vessel phantoms without stenosis. 1 = nitinol with
1.7-mm2 mesh size and 165-µm strut thickness (Expander, Bolton
Medical), 2 = nitinol with 3-mm2 mesh size and 170-µm strut
thickness (Hemobahn, Prograft Medical), 3 = nitinol with 4.5-mm2
mesh size and 245-µm strut thickness (SelfX, Abbott Laboratories), 4 =
nitinol with 2-mm2 mesh size and 200-µm strut thickness (SMART,
Cordis), 5 = nitinol with 15-mm2 mesh size and 235-µm strut
thickness (Symphony, Boston Scientific), 6 = cobalt alloy with
1.7-mm2 mesh size and 195-µm strut thickness (Wallstent, Boston
Scientific). Graph shows results for CT angiography of 95% stenosis.
|
|
MR Angiography
On MR angiography, the relative signal intensity of the in-stent stenoses
ranged from 30% to 100% of the signal intensity of the control
(Table 1, Figs.
2C and
2D). The higher variation in
signal intensity was mainly caused by the high signal attenuation of
ferromagnetic stents (stent 6, Wallstent). Again, the presence of localized
artifacts and resulting inhomogeneous luminal signal intensity caused lower
subjective ratings than the subjective score would suggest. The overall signal
intensity due to the presence of the stent in unstenosed vessels is shown in
Figure 3B. Signal intensity
correlated primarily with stent material, and there was no significant
influence of strut or mesh size.

View larger version (7K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3B —Graphs show stent-related relative attenuation or signal
intensity in vessel phantoms without stenosis. 1 = nitinol with
1.7-mm2 mesh size and 165-µm strut thickness (Expander, Bolton
Medical), 2 = nitinol with 3-mm2 mesh size and 170-µm strut
thickness (Hemobahn, Prograft Medical), 3 = nitinol with 4.5-mm2
mesh size and 245-µm strut thickness (SelfX, Abbott Laboratories), 4 =
nitinol with 2-mm2 mesh size and 200-µm strut thickness (SMART,
Cordis), 5 = nitinol with 15-mm2 mesh size and 235-µm strut
thickness (Symphony, Boston Scientific), 6 = cobalt alloy with
1.7-mm2 mesh size and 195-µm strut thickness (Wallstent, Boston
Scientific). Graph shows results for MR angiography of 95% stenosis.
|
|
Discussion
Metallic stents are commonly used in the management of peripheral arterial
stenosis [18]. In-stent
restenosis, however, is one of the main reasons that long-term patency rates
are low [19]. Although CT
angiography and MR angiography have been successfully implemented for
visualization of the peripheral arteries, visualization of the stent lumen is
impeded by overall signal attenuation inside the stent
[20] and localized signal
inhomogeneities due to artifact formation
[21,
22]. Therefore, intraarterial
subtraction angiography remains the method of choice for detection of in-stent
restenosis due to intimal hyperplasia or thrombus formation. In this study, we
observed that the detectability of instent stenosis on CT angiography was
mainly limited by overall attenuation inside the stent, whereas on MR
angiography, it is dominated by both overall signal attenuation inside the
stent and localized signal inhomogeneity due to artifact formation.
On CT angiography, 95% stenosis was differentiated from total occlusion
inside all evaluated stent types. However, in contrast to the findings on MR
angiography, low-grade stenosis tended to be underestimated owing to the
presence of high-attenuation artifacts, especially in stents 3 and 5. This
effect was caused in part by the relative proximity of the borders of the
larger bore to the stent meshes in the case of 75% stenosis as opposed to the
small and thus more centrally located bore of 95% stenosis.
The results of MR angiography were characterized by a higher variety of
in-stent signal intensities across stent types. Luminal stenosis tended to be
overestimated in general. In 75% stenosis, signal attenuation in the residual
lumen (Fig. 3B) was mainly
responsible for poorer visibility (Fig.
2C) but still allowed accurate delineation of low-grade stenosis
and differentiation from fully patent vessel parts. In 95% stenosis, however,
localized artifacts caused additional inhomogeneity that could have been
misinterpreted as occlusion in stents 2, 4, 5, and 6.
Our experiments were conducted in a static phantom. The effect of blood
flow through the various stents is expected to be negligible on CT
angiography, whereas on MR angiography, flow features are known to cause
changes in the appearance of stent-related artifacts due to intravoxel
dephasing. However, these effects should be minor compared with susceptibility
and eddy currents [23]. The
stents in this experiment were scanned parallel to the z-axis. In
vivo, arteries are often tortuous, and stent orientation influences the degree
of artifact formation and detectability of stenosis
[14]. The most important
causes of in-stent restenosis are thrombus formation within the stented
segment and neointimal growth through smooth-muscle cell proliferation
[24–26].
Use of a central bore in our model may better approximate in vivo conditions,
in which endothelial hyperplasia usually leads to concentric luminal narrowing
[27].
In conclusion, CT angiography and MR angiography of in-stent stenosis
suffer from different degrees of artifact formation depending on stent type
and degree of stenosis. In general, CT angiography seems to be more suitable
for differentiation between high-grade stenosis and occlusion, whereas MR
angiography seems to have higher sensitivity in the early detection of
hemodynamically relevant intermediate-grade stenosis. For stents not made of
nitinol, CT angiography would generally be preferable. Nitinol stents with
small meshes and thin struts tend to be better delineated on CT than on MR
angiography.
References
- Hagspiel KD, Leung DA, Nandalur KR, et al. Contrast-enhanced MR
angiography at 1.5 T after implantation of platinum stents: in vitro and in
vivo comparison with conventional stent designs. AJR2005; 184:288
–294[Abstract/Free Full Text]
- Herborn CU, Goyen M, Quick HH, et al. Whole-body 3D MR angiography
of patients with peripheral arterial occlusive disease.
AJR 2004; 182:1427
–1434[Abstract/Free Full Text]
- Heuschmid M, Krieger A, Beierlein W, et al. Assessment of
peripheral arterial occlusive disease: comparison of multislice-CT angiography
(MS-CTA) and intraarterial digital subtraction angiography (IA-DSA).
Eur J Med Res 2003;8
: 389–396[Medline]
- Loewe C, Schillinger M, Haumer M, et al. MRA versus DSA in the
assessment of occlusive disease in the aortic arch vessels: accuracy in
detecting the severity, number, and length of stenoses. J Endovasc
Ther 2004; 11:152
–160[CrossRef][Medline]
- Zhang HL, Khilnani NM, Prince MR, et al. Diagnostic accuracy of
time-resolved 2D projection MR angiography for symptomatic infrapopliteal
arterial occlusive disease. AJR 2005;184
: 938–947[Abstract/Free Full Text]
- Katzberg RW, Sahler LG, Duda SW, et al. Renal handling and
physiologic effects of the paramagnetic contrast medium Gd-DOTA.
Invest Radiol 1990;25
: 714–719[Medline]
- Maintz D, Tombach B, Juergens KU, Weigel S, Heindel W, Fischbach R.
Revealing in-stent stenoses of the iliac arteries: comparison of multidetector
CT with MR angiography and digital radiographic angiography in a phantom
model. AJR 2002;179
:1319
–1322[Abstract/Free Full Text]
- Prokop M, Debatin JF. MRI contrast media: new developments and
trends. CTA vs. MRA. Eur Radiol 1997;7
[suppl 5]:299
–306[CrossRef][Medline]
- Cejna M, Loewe C, Schoder M, et al. MR angiography vs CT
angiography in the follow-up of nitinol stent grafts in endoluminally treated
aortic aneurysms. Eur Radiol 2002;12
:2443
–2450[Medline]
- Hong C, Chrysant GS, Woodard PK, Bae KT. Coronary artery stent
patency assessed with in-stent contrast enhancement measured at multi-detector
row CT angiography: initial experience. Radiology2004; 233:286
–291[Abstract/Free Full Text]
- Haltmayer M, Mueller T, Horvath W, Luft C, Poelz W, Haidinger D.
Impact of atherosclerotic risk factors on the anatomical distribution of
peripheral arterial disease. Int Angiol2001; 20:200
–207[Medline]
- Funovics MA, Lackner B, Cejna M, et al. Predictors of long-term
results after treatment of iliac artery obliteration by transluminal
angioplasty and stent deployment. Cardiovasc Intervent
Radiol 2002; 25:397
–402[CrossRef][Medline]
- Manke C, Nitz WR, Djavidani B, et al. MR imaging-guided stent
placement in iliac arterial stenoses: a feasibility study.
Radiology 2001;219
: 527–534[Abstract/Free Full Text]
- Maintz D, Kugel H, Schellhammer F, Landwehr P. In vitro evaluation
of intravascular stent artifacts in three-dimensional MR angiography.
Invest Radiol 2001;36
: 218–224[CrossRef][Medline]
- Kramer SC, Wall A, Maintz D, Bachmann R, Kugel H, Heindel W. 3.0
Tesla magnetic resonance angiography of endovascular aortic stent grafts:
phantom measurements in comparison with 1.5 Tesla. Invest
Radiol 2004; 39:413
–417[CrossRef][Medline]
- Hamoir XL, Flohr T, Hamoir V, et al. Coronary arteries: assessment
of image quality and optimal reconstruction window in retrospective ECG-gated
multislice CT at 375-ms gantry rotation time. Eur
Radiol 2005; 15:296
–304[CrossRef][Medline]
- Baum F, Vosshenrich R, Fischer U, Castillo E, Grabbe E. Stent
artifacts in 3D MR angiography: experimental studies [in German].
Rofo 2000; 172:278
–281[Medline]
- Surowiec SM, Davies MG, Eberly SW, et al. Percutaneous angioplasty
and stenting of the superficial femoral artery. J Vasc
Surg 2005; 41:269
–278[CrossRef][Medline]
- Faries P, Morrissey NJ, Teodorescu V, et al. Recent advances in
peripheral angioplasty and stenting. Angiology2002; 53:617
–626[Medline]
- Sardanelli F, Zandrino F, Molinari G, Iozzelli A, Balbi M, Barsotti
A. MR evaluation of coronary stents with navigator echo and breath-hold cine
gradient-echo techniques. Eur Radiol2002; 12:193
–200[CrossRef][Medline]
- Bartels LW, Smits HF, Bakker CJ, Viergever MA. MR imaging of
vascular stents: effects of susceptibility, flow, and radiofrequency eddy
currents. J Vasc Interv Radiol 2001;12
: 365–371[Medline]
- Shetty AN, Bis KG, Kirsch M, Weintraub J, Laub G. Contrast-enhanced
breath-hold three-dimensional magnetic resonance angiography in the evaluation
of renal arteries: optimization of technique and pitfalls. J Magn
Reson Imaging 2000; 12:912
–923[CrossRef][Medline]
- Hilfiker PR, Quick HH, Debatin JF. Plain and covered stent-grafts:
in vitro evaluation of characteristics at three-dimensional MR angiography.
Radiology 1999;211
: 693–697[Abstract/Free Full Text]
- Cejna M, Virmani R, Jones R, et al. Biocompatibility and
performance of the Wallstent and the Wallgraft, Jostent, and Hemobahn
stent-grafts in a sheep model. J Vasc Interv Radiol2002; 13:823
–830[Medline]
- Cejna M, Virmani R, Jones R, et al. Biocompatibility and
performance of the Wallstent and several covered stents in a sheep iliac
artery model. J Vasc Interv Radiol 2001;12
: 351–358[Medline]
- Vorwerk D, Gunther RW, Schurmann K, Wendt G. Aortic and iliac
stenoses: follow-up results of stent placement after insufficient balloon
angioplasty in 118 cases. Radiology 1996;198
: 45–48[Abstract/Free Full Text]
- Yorozuya M, Suzuki H, Iso Y, et al. Comparison of the morphological
changes of restenosis after the implantation of various types of stents in a
swine model. Coron Artery Dis 2002;13
: 305–312[CrossRef][Medline]

CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?