|
|
||||||||
Original Research |
1 Department of Clinical Radiology, Hiroshima University Hospital, 1-2-3,
Kasumi-cho, Minami-ku, Hiroshima, 734-8551, Japan.
2 CT Lab of Great China, GE Healthcare, Mongkok, Kowloon, Hong Kong.
3 GE Yokogawa Medical Systems, Tokyo, Japan.
4 Department of Radiology, Division of Medical Intelligence and Informatics,
Programs for Applied Biomedicine, Graduate School of Biomedical Sciences,
Hiroshima University, Hiroshima, Japan.
Received March 8, 2007;
accepted after revision August 15, 2007.
Address correspondence to J. Horiguchi
(horiguch{at}hiroshima-u.ac.jp).
Abstract
|
|
|---|
MATERIALS AND METHODS. Coronary artery models (n = 3) with
different plaque densities (
50,
110, and
1,000 H) on a cardiac
phantom were scanned in variable heart rate sequences (n = 14) with
both prospective ECG-triggered and retrospective ECG-gated scanning. Radiation
dose, image quality graded by motion and stairstep artifacts (grade 1,
excellent, to grade 4, poor, with grades 1 and 2 defined as satisfactory),
accuracy of stenosis measurement (area; 18%, 50%, and 82%), and CT
densitometry of plaques (soft,
50; and intermediate,
110 H) were
compared between the two protocols using the Mann-Whitney U test and
repeated measures.
RESULTS. The radiation dose of prospective ECG-triggered CT angiography (CTA) (3.0 mSv) was lower than that of retrospective ECG-gated CTA (11.7–13.0 mSv) when the same tube current (mA) and voltage (kVp) were used in both methods. Prospective ECG-triggered CTA images were assigned a satisfactory quality rating in stable heart rate up to 75 beats per minute (bpm) when using the minimal X-ray exposure time. In this range, there were no significant differences in stenosis measurement (p = 0.17) and CT densitometry (p = 0.93) between the two protocols.
CONCLUSION. Prospective ECG-triggered coronary 64-MDCT has the potential to reduce radiation exposure while maintaining the diagnostic performance of retrospective ECG-gated coronary 64-MDCT.
Keywords: cardiac imaging coronary artery CT angiography densitometry radiation dose stenosis measurement
|
|
|---|
The major drawback of retrospective ECG-gated coronary CTA is its high radiation dose because of overlapping data acquisition, acquiring data in cardiac phases that do not contribute to image reconstruction. The estimated effective dose reported with 16- or 64-MDCT using 120 kV and ECG-correlated tube current modulation (reduction of tube current in systole), which reduces radiation exposure by 30–50% [14, 15], was estimated to be between 6.4 and 14.7 mSv [15–17]. Prospective ECG-triggered coronary CTA has so far been performed using electron-beam CT. Wide coverage and reduced gantry rotation time on 64-MDCT and shortening of the interscan delay enable the acquisition of thin-slice data sets during one breath-hold [18]. The prospective ECG-triggered coronary 64-MDCT software minimizes overlapping X-ray exposure, thus enabling far greater reduction of radiation exposure. The purpose of this study was to compare prospective ECG-triggered and retrospective ECG-gated coronary 64-MDCT as to radiation dose, image quality, accuracy of stenosis measurement, and CT densitometry in a phantom study.
|
|
|---|
|
Coronary Artery Plaque Models
Four coronary artery cylinder models (Coronary Artery Vessel Phantom, Fuyo
Corporation) with a diameter of 4 mm and different plaque densities were
manufactured for this experiment. For plaque materials,
acrylonitrile-butadiene-styrene resin (
50 H), acrylic (
110 H), and
polytetrafluoroethylene (Teflon, DuPont) (
1,000 H) were used. The plaque
CT values of 50, 110, and 1,000 H represented soft, intermediate, and
calcified plaques, respectively. Each coronary artery model had three levels
of plaque size (1, 2, and 3 mm in diameter), resulting in area stenoses of the
coronary arteries of 18%, 50%, and 82%
(Fig. 1). The coronary artery
models were attached to the balloon phantom (mimicking the heart) with the
long axis of the model corresponding to the z-axis and were
surrounded by corn oil (–112 H) simulating epicardial fat
(Fig. 2).
|
|
75 bpm, a two-sector reconstruction
algorithm, offering an improved temporal resolution, was used. A half-scan
reconstruction requires about a two-thirds rotation data acquisition compared
with full rotation (360°) for a full-scan reconstruction. Temporal
resolution in single-sector reconstruction is approximately equal to half the
rotation time. A two-sector reconstruction provided a temporal resolution down
to 87 milliseconds, depending on heart rates. For image reconstruction, a
cardiac phase with the center of the temporal window corresponding to 80%
(diastole of the phantom) of the R-R interval was used.
Prospective ECG-Triggered Coronary 64-MDCT Angiography Protocol
A prospective ECG-triggered software protocol (SnapShot Pulse, GE
Healthcare) in which a 64-detector rows x 0.625 mm-collimation
configuration (40 mm) is used and three or four radiographic exposures per
examination with 5-mm overlapping that can cover 10.5 cm or 14 cm in the
z-axis was available in our institution. This overlapping is used for
minimizing cone-beam artifacts (Fig.
3). The tube current of 650 mA and other scanning parameters were
the same as in the retrospective ECG-gated coronary CTA protocol. In this
software, X-ray exposure time was changeable from 233 to 633 milli-seconds,
maintaining the temporal resolution of 175 milliseconds. We set the exposure
time to the minimum (233 milliseconds) in the current study. The prospective
scanning was performed so that the center of the temporal window corresponded
to 80% of the R-R interval.
|
![]() |
Image Quality Graded by Motion and Stairstep Artifacts
In each of the 14 heart rate sequences, the image quality of three coronary
artery models was assessed separately by three independent observers who had
coronary CTA reading experience of 5, 3, and 3 years. All data sets were
blinded with regard to the protocols, reconstruction technique, and heart
rate, and they were assessed in a random order. Images were displayed with a
fixed window setting (width, 800 H; level, 130 H) and evaluated on a separate
workstation (Advantage Workstation version 4.2, GE Healthcare). We used a
classification system that was modified from Herzog et al.
[22], in which image quality
was graded in a 4-step scale (grade 1, excellent; grade 2, good; grade 3,
moderate; and grade 4, poor). Excellent image quality was attributed to
vessels showing a continuous course, without stairstep artifacts on
volume-rendered or reformation images and appearing on transverse CT scans as
bright circular areas without motion artifacts and surrounded by
low-attenuation fat tissue. Good image quality was assigned in the presence of
discrete blurring of vessel margin, minor motion artifacts seen as a discrete
tail or streak-emitting shadows on transverse images, and minimal stairstep
artifact on volume-rendered or reformation images. Moderate image quality was
assigned to noticeably blurred vessels or plaque margins, distinctly broader
motion artifacts extending less than 5 mm from the vessel center, and
stairstep artifacts of less than 25% of the vascular diameter. Poor image
quality was defined as an inadequate delineation between the vessel and
surrounding tissue, the presence of streak artifacts extending at least 5 mm
from the center of the vessel, and stairstep artifacts of more than 25% of the
vessel diameter. The grading was performed nine times (three coronary artery
models, three readers) per heart rate sequence, and the grades were compared
between prospective ECG-triggered and retrospective ECG-gated images using the
Mann-Whitney U test.
The grade per coronary artery model was also determined by consensus of three readers or by a two- to-one decision. Images from heart rate sequences, in which all grades for coronary artery models (n = 4) on prospective ECG-triggered images were graded 1 or 2, were used for further investigations.
Measurement Error of Stenosis
Stenosis levels of soft, intermediate, and calcified plaques were measured
on both prospective ECG-triggered and retrospective ECG-gated coronary CTA
protocols using the workstation semiautomatic software Vessel Analysis
(Advantage Workstation version 4.2, GE Healthcare). Using this software, one
reader plotted four or five points at the center of the lumen, and then
stenosis in the vessel was automatically calculated. The measurements were
recorded in 12 slices per stenosis level. The measurement error of coronary
artery stenosis was defined as abs (measured area of the stenosis
– known stenosis area)/known stenosis area. The measurement errors of
the three kinds of plaque were compared between prospective ECG-triggered and
retrospective ECG-gated coronary CTA protocols using repeated measures
analysis of variance.
CT Densitometry of Soft and Intermediate Plaque
CT densities of soft and intermediate plaques were measured on both
prospective ECG-triggered and retrospective ECG-gated coronary CTA protocols.
Regions of interest (ROIs) of 1 mm2 were set at the center of the
50% stenosis plaque (Fig. 4).
CT densitometry measurements were performed in 10 slices per plaque by one
reader. CT densitometry, from the heart rate sequences selected, was compared
between prospective ECG-triggered and retrospective ECG-gated coronary CTA
protocols.
|
|
|
|---|
Radiation Dose
CTDIvol and DLP displayed on the dose report on the CT scanner
were 12.6 mGy and 176 mGy x cm for the prospective ECG-triggered and
49.3–54.5 mGy and 690–763 mGy x cm (depending on the pitch
selected) on the retro-spective ECG-gated coronary CTA. Effective doses,
calculated for typical patient scanning ranges, were 3.0 mSv on prospective
ECG-triggered and 11.7–13.0 mSv on retrospective ECG-gated coronary
CTA.
Image Quality Graded by Motion and Stairstep Artifacts
Image quality grading results of both prospective ECG-triggered and
retrospective ECG-gated coronary CTA on 14 heart rate sequences are shown in
Table 2. Prospective
ECG-triggered CTA on stable heart rate sequences up to 75 bpm had satisfactory
image quality. In this range, the image quality of prospective ECG-triggered
CTA was comparable or superior (at 70 bpm) to that of retro-spective ECG-gated
CTA. In contrast, image quality of prospective ECG-triggered CTA was
unsatisfactory on shift (unstable heart rate) and arrhythmia sequences. In
accordance with coronary artery model analysis, there was a consensus decision
in 52 (62%) and a two-to-one decision in 32 (38%) of 84 assignments. Split
decision between the three readers was not observed. Image quality of
prospective ECG-triggered CTA was graded 1 or 2 (satisfactory) on all images
up to 75 bpm. Therefore, these images were used for further
investigations.
|
|
|
CT Densitometry of Soft and Intermediate Plaque
The results of CT densitometry are shown in
Figure 6. Repeated measures
analysis of variance revealed that there was no statistically significant
difference of CT densitometry between retrospective ECG-gated and prospective
ECG-triggered coronary CTA (p = 0.93); however, there was a
significant difference between soft and intermediate plaques (p <
0.01). The Scheffé test revealed statistical significance between the
plaques (p < 0.01).
|
|
|---|
Furthermore, the decreased tube voltage leads to increased opacification of vascular structures during contrast-enhanced CTA owing to an increase in the photoelectric effect and a decrease in Compton scattering [26]. The radiation dose of the current prospective ECG-triggered coronary CTA is dramatically reduced because overlapping data acquisition is minimized. In addition, the technology principally acquires data only when it is necessary for image reconstruction. Thus the radiation dose of the prospective ECG-triggered CTA in the current study was lower than that of diagnostic coronary angiography, retrospective ECG-gated coronary 64-MDCT, and average annual background radiation in the United States (3.6 mSv) [21]. The low-dose prospective ECG-triggered coronary 64-MDCT CTA has the advantage of following up patients after stent implantation and coronary artery bypass grafts.
To keep the radiation dose to a minimum, we set the X-ray exposure time at 233 milliseconds (two thirds of the gantry rotation speed). The optimal cardiac cycle for reconstruction is known to differ between individuals and coronary arteries. In a study of coronary 16-MDCT with a rotation time of 420 milliseconds, the best image quality was obtained with end-systolic and early diastolic intervals in patients with high heart rates [22]. In a more recent study, however, using 64-MDCT with a rotation time of 330 milliseconds, the best image quality was provided in middiastole in most patients, although it was nondiagnostic at any reconstruction interval in some patients [27]. With the current software, we are able to set the exposure time up to a maximum of 633 milliseconds in an interval of 2 milliseconds. Using the "Dynamic Padding" function, at the expense of increased radiation exposure, other cardiac phases can be retrospectively reconstructed while a temporal resolution of 175 milliseconds is maintained. This function, a trade-off between increased radiation dose—however, far less than that of retrospective-ECG-gated CTA—and reconstruction availability, should be validated in clinical studies.
Limited temporal and spatial resolutions compared with catheter angiography and calcium deposition are major causes of reduced diagnostic performance of coronary CTA. Bulky calcification causes blooming effect and volume averaging. Therefore, the accuracy for detection of coronary stenosis is reported to be lower in the presence of severe calcification [27–29]. The results of measurement errors of stenosis on soft and intermediate plaques are acceptable in consideration of the spatial resolution of the CT scanner and the plaque sizes. We, however, must regretfully acknowledge that the stenosis measurement for calcified plaque is difficult on coronary CTA.
With the introduction of the previously mentioned software, we believe that our strategy for single-source coronary 64-MDCT will be as follows: If the patient's heart rate is up to 75 bpm and stable, prospective ECG-triggered CTA with minimal X-ray exposure is recommended. The robustness of prospective ECG-triggered CTA with the padding function against heart rates up to 75 bpm with a small variation needs investigation. If the patient's heart rate is higher than 75 bpm and stable, the use of retrospective ECG-gated CTA with ECG-correlated tube current modulation and multisector reconstruction is recommended. If the patient's heart rate is mildly irregular, retrospective ECG-gated CTA without ECG-correlated tube current modulation followed by ECG editing [30]—that is, by arbitrarily modifying the position of the temporal windows within the cardiac cycle to correct and compensate for part or all of the artifacts produced by mild heart rhythm irregularities—will be necessary.
This study has several limitations. We reconstructed only one cardiac phase (the center of the temporal window corresponding to 80% of the R-R cycle) on the prospective ECG-triggered CTA. Therefore, we did not validate the Dynamic Padding function. We did not investigate low-voltage scanning as another method of decreasing radiation exposure. We think that these issues should be verified in clinical studies.
In conclusion, SnapShot Pulse software has the potential to reduce the radiation dose of 64-MDCT. When using the minimal X-ray exposure time, this software seems to be comparable to retrospective ECG-gated coronary 64-MDCT on stable heart rates up to 75 bpm.
|
|
|---|
This article has been cited by other articles:
![]() |
B. Huang, M. W.-M. Law, H. K.-F. Mak, S. P.-F. Kwok, and P.-L. Khong Pediatric 64-MDCT Coronary Angiography With ECG-Modulated Tube Current: Radiation Dose and Cancer Risk Am. J. Roentgenol., August 1, 2009; 193(2): 539 - 544. [Abstract] [Full Text] [PDF] |
||||
![]() |
C. Fujioka, J. Horiguchi, M. Kiguchi, H. Yamamoto, T. Kitagawa, and K. Ito Survey of Aorta and Coronary Arteries With Prospective ECG-Triggered 100-kV 64-MDCT Angiography Am. J. Roentgenol., July 1, 2009; 193(1): 227 - 233. [Abstract] [Full Text] [PDF] |
||||
![]() |
K. M. Takakuwa, E. J. Halpern, E. L. Gingold, D. C. Levin, and F. S. Shofer Radiation Dose in a "Triple Rule-Out" Coronary CT Angiography Protocol of Emergency Department Patients Using 64-MDCT: The Impact of ECG-Based Tube Current Modulation on Age, Sex, and Body Mass Index Am. J. Roentgenol., April 1, 2009; 192(4): 866 - 872. [Abstract] [Full Text] [PDF] |
||||
![]() |
P. Stolzmann, H. Scheffel, S. Leschka, A. Plass, S. Baumuller, B. Marincek, and H. Alkadhi Influence of Calcifications on Diagnostic Accuracy of Coronary CT Angiography Using Prospective ECG Triggering Am. J. Roentgenol., December 1, 2008; 191(6): 1684 - 1689. [Abstract] [Full Text] [PDF] |
||||
![]() |
E J HALL and D J BRENNER Cancer risks from diagnostic radiology Br. J. Radiol., May 1, 2008; 81(965): 362 - 378. [Abstract] [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |