DOI:10.2214/AJR.07.2951
AJR 2008; 191:26-31
© American Roentgen Ray Society
The Role of ECG-Gated MDCT in the Evaluation of Aortic and Mitral Mechanical Valves: Initial Experience
Eli Konen1,
Orly Goitein1,
Micha S. Feinberg2,
Yael Eshet1,
Ehud Raanani3,
Uri Rimon1 and
Elio Di-Segni1,2
1 Department of Diagnostic Imaging, Sheba Medical Center, Tel Hashomer, Ramat
Gan 52621, Israel.
2 Heart Institute, Sheba Medical Center, Ramat Gan, Israel.
3 Department of Cardiac Surgery, Sheba Medical Center, Ramat Gan, Israel.
Received July 28, 2007;
accepted after revision January 15, 2008.
Address correspondence to E. Konen
(eli.konen{at}sheba.health.gov.il).
Abstract
OBJECTIVE. The objective of our study was to evaluate the role of
ECG-gated MDCT in the functional evaluation of mechanical prosthetic aortic
and mitral valves.
MATERIALS AND METHODS. Twenty sequential patients with 23 mechanical
prosthetic valves were evaluated with an ECG-gated 40- or 64-MDCT scanner.
Multiplanar reformation, maximal-intensity-projection, volume-rendering, and
volume-averaging techniques were used for visualization of valve leaflets in
systole and diastole. The visibility of each mechanical valve was evaluated by
consensus of a radiologist and a cardiologist using a subjective 5-point scale
(0–4). MDCT findings were correlated with fluoroscopic opening and
closing angle measurements and echocardiographic pressure gradient
measurements in 11 and 19 valves, respectively.
RESULTS. The series included 18 bileaflet and five single-leaflet
mechanical valves. The visibility score for the bileaflet mechanical valves
was excellent (score of 4) in all 18 cases, but it was lower for
single-leaflet valves (mean score, 2.8; range, 1–4) (p = 0.04).
Bland-Altman plots showed high agreement between MDCT and fluoroscopy for
measurements of opening and closing angles of bileaflet mechanical valves. In
four patients, a stuck valve was seen on MDCT and was confirmed by
fluoroscopy. Doppler echocardiography showed increased transvalvular pressure
in two of the four patients with a stuck mitral valve and increased
transaortic pressure in four patients with normal prosthetic aortic valve
motion.
CONCLUSION. Our preliminary results suggest that MDCT is a promising
technique for functional evaluation of bileaflet mechanical valves, allowing
reliable measurements of opening and closing leaflet angles. However, the role
of MDCT in the evaluation of single-leaflet valves might be limited.
Keywords: aortic valve cardiac imaging MDCT mitral valve prosthetic valves
Introduction
Although mechanical cardiac valve–related complications are
relatively rare, obstruction of pros thetic valves is a serious and
life-threatening complication caused by thrombus formation, pannus ingrowth,
or both
[1–3].
Frequently, clinical presentation is insidious
[3]. Mechanical prosthetic
valves are currently evaluated using a combination of echocardiography and
cinefluoroscopy
[3–6].
Both methods have their limitations and are, to a certain degree, operator
dependent. Optimal visualization of valve leaflets and of their motion may
occasionally be difficult with these techniques
[5,
7–9].
ECG-gated cardiac CT angiography is a novel and promising noninvasive
technique for coronary artery disease evaluation. We hypothesized that the
same raw data could be further processed, as presented here, for the
evaluation of mechanical prosthetic valves without a need for additional
contrast medium or ionizing radiation. Moreover, prosthetic valves can be
evaluated using the same CT data processing scanning only at valve level and
without contrast injection when coronary visualization is not requested.
Because CT is a 3D technique, it should circumvent the occasional limitations
of the techniques mentioned earlier to allow optimal visualization of
prosthetic valves. The aim of this retrospective study was to report our
initial experience using MDCT for the evaluation of single-leaflet and
bileaflet mechanical prosthetic valves.
Materials and Methods
MDCT data of 20 sequential patients (age range, 32–79 years; mean
age, 60.4 years; 15 men) with 23 mechanical prosthetic valves were
retrospectively studied. The series included 14 aortic and nine mitral
mechanical valves and 18 bileaflet and five single-leaflet valves. The period
between valve replacement and CT ranged from 5 months to 26 years (average
± SD, 5.2 ± 4.4 years). The indications for CT examinations were
coronary artery evaluation in 14 patients (with 15 valves) and valvular
functional evaluation in six patients (with eight valves); in the latter
group, no IV contrast material was used.
Complete body data were available for 15 of the 20 patients: Average height
was 169 cm (range, 153–185 cm); average weight, 78.8 kg (range,
55–112 kg); average body surface area, 1.9 m2 (range,
1.6–2.2 m2); and average body mass index, 27.5
kg/m2 (range, 21.8–37.2 kg/m2).
All patients underwent retrospectively ECG-gated CT using either a 40-MDCT
(seven patients with eight prosthetic valves) or a 64-MDCT (13 patients with
15 prosthetic valves) scanner (Brilliance, Philips Medical Systems). Scanning
was performed with both scanner types at 120 kV using 800–1,000 mAs with
a detector collimation of 0.625 mm and gantry rotation speed of 0.42 second.
The minimal slice thickness was 0.67 mm, and the reconstruction interval was
0.4 mm. Using retrospective ECG gating, reconstructions were obtained in 10
cardiac phases of the R-R interval period. In the 14 patients who were
referred for coronary artery evaluation, 80–100 mL of contrast medium
(iomeprol [Iomeron 400, Bracco Imaging]) was injected IV at a rate of
4–5 mL/s using an automatic injector. The remaining six patients, two of
whom had two mechanical valves each, were referred for a questionable stuck
valve be cause of suboptimal visualization of leaflets during fluoroscopy. In
these cases, the scanning area was limited merely to the valve level and the
exami nation was performed without contrast material. Retrospective review and
analysis of the patients' imaging data were approved by the local ethics
committee.

View larger version (80K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 1A —32-year-old man with normal functioning bileaflet aortic
valve (patient 16 in Table 1).
Multiplanar reformation images perpendicular to leaflets' axis enable
measurement of closing (A) and opening (B) angles during
diastole and systole, respectively.
|
|

View larger version (84K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 1B —32-year-old man with normal functioning bileaflet aortic
valve (patient 16 in Table 1).
Multiplanar reformation images perpendicular to leaflets' axis enable
measurement of closing (A) and opening (B) angles during
diastole and systole, respectively.
|
|

View larger version (70K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2A —77-year-old man with normal functioning single-leaflet aortic
mechanical valve (patient 18 in Table
1). Multiplanar reformation images perpendicular to leaflet's axis
enable measurements of closing (A) and opening (B) angles during
diastole and systole, respectively.
|
|

View larger version (82K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 2B —77-year-old man with normal functioning single-leaflet aortic
mechanical valve (patient 18 in Table
1). Multiplanar reformation images perpendicular to leaflet's axis
enable measurements of closing (A) and opening (B) angles during
diastole and systole, respectively.
|
|
Analysis of scans was performed on a dedicated workstation (Extended
Brilliance Workspace, Philips Medical Systems). A variety of 3D postprocessing
techniques, including multiplanar reformation (MPR), maximal intensity
projection, volume rendering, and volume averaging, were used for
visualization of the valve leaflets in systole and diastole. Reformations were
generated perpendicular to the leaflet movement axis, enabling visualization
of 10 frames per R-R interval cine. Measurements of mechanical leaflet opening
and closing angles at end-systole and end-diastole were recorded in all cases
with visible leaflets.
Visibility of the valves during the heart cycle was subjectively estimated
by consensus of an experienced radiologist (5 years of experience with cardiac
CT) and an experienced cardiologist (30 years of experience in cardiac
catheterization) using the following 5-point scale: score of 0, mechanical
leaflet was not visible; 1, leaflet was hardly seen; 2, leaflet was identified
in some of the reconstructed phases but opening and closing angle measurements
could not be obtained reliably; 3, images allowed opening and closing angle
measurements but images were partially blurred; and 4, visualization of
valvular angles was excellent in both end-diastole and end-systole, allowing
reliable measurements. Because of the clinical indications for the MDCT
examinations, the readers were not blinded to fluoroscopy results.
Correlation of MDCT findings with cardiac fluoroscopy was available for 11
valves. Cinefluoroscopy was performed using single-plane cardiac
catheterization equipment with a C-arm unit. The C-arm was rotated to obtain a
tangential view of the prosthetic valve. The examination was considered
successful when a projection was obtained with the x-ray beam parallel to both
the valve ring plane and the tilting axis of the disk or disks, enabling full
visualization of the valve leaflet or disk in profile and allowing calculation
of the valve opening angle.
For both CT and cineradiography, opening and closing valve angles were
defined by frame-by-frame analysis of a single cardiac cycle. For bileaflet
valves, opening and closing angles were measured between the two leaflets in
the fully open and closed positions
[10] (Fig.
1A,
1B). For single-leaflet valves,
measurements of the lesser angle between the disk and the valve strut at full
opening and full closure were recorded (Fig.
2A,
2B). A valve was defined as
stuck when motion of one valve leaflet was absent. Prosthetic valve
obstruction was diagnosed when motion of a leaflet or leaflets was
persistently restricted, with a calculated opening angle of more (for
bileaflet valve) or less (for disk valves) than the values for a normal valve,
as specified by the manufacturer.
Transthoracic Doppler echocardiography was available in 19 patients.
Doppler echocardiography measurements included transvalvular velocity in
meters per second and calculation of the peak and mean pressure gradients
using the modified Bernoulli equation (pressure gradient [mm Hg] =
velocity2 x 4). Valve leaflet motion was qual itatively
assessed.
Leaflet angle measurements obtained with MDCT and those obtained with
fluoroscopy were compared using a two-tailed paired Student's t test
and Bland-Altman analysis. Data were expressed as means and SDs. Visibility
scores for single-leaflet versus bileaflet mechanical valves were compared
using the two-sided Fisher's exact test.
Results
The visibility of bileaflet mechanical valves with MDCT was excellent
(score = 4) in all cases. Conversely, in two of five cases with single-leaflet
valves, the opening and closing angles could not be clearly defined on MDCT,
resulting in a mean visibility score of only 2.8 for single-leaflet mechanical
valves (p = 0.04) (Table
1).

View larger version (106K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5A —66-year-old man with stuck mitral bileaflet mechanical valve
(patient 2 in Table 1).
Multiplanar reformation images (A and B) and volume-rendering
reformations (C and D) during diastole (A and C)
and during systole (B and D) and corresponding fluoroscopic
images (E and F) show stuck leaflet (white arrow,
A). Note soft-tissue attenuation on ventricular border (black
arrow, A) of stuck valve, which is suggestive of pannus.
|
|

View larger version (94K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5B —66-year-old man with stuck mitral bileaflet mechanical valve
(patient 2 in Table 1).
Multiplanar reformation images (A and B) and volume-rendering
reformations (C and D) during diastole (A and C)
and during systole (B and D) and corresponding fluoroscopic
images (E and F) show stuck leaflet (white arrow,
A). Note soft-tissue attenuation on ventricular border (black
arrow, A) of stuck valve, which is suggestive of pannus.
|
|

View larger version (126K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5C —66-year-old man with stuck mitral bileaflet mechanical valve
(patient 2 in Table 1).
Multiplanar reformation images (A and B) and volume-rendering
reformations (C and D) during diastole (A and C)
and during systole (B and D) and corresponding fluoroscopic
images (E and F) show stuck leaflet (white arrow,
A). Note soft-tissue attenuation on ventricular border (black
arrow, A) of stuck valve, which is suggestive of pannus.
|
|

View larger version (131K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5D —66-year-old man with stuck mitral bileaflet mechanical valve
(patient 2 in Table 1).
Multiplanar reformation images (A and B) and volume-rendering
reformations (C and D) during diastole (A and C)
and during systole (B and D) and corresponding fluoroscopic
images (E and F) show stuck leaflet (white arrow,
A). Note soft-tissue attenuation on ventricular border (black
arrow, A) of stuck valve, which is suggestive of pannus.
|
|

View larger version (97K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5E —66-year-old man with stuck mitral bileaflet mechanical valve
(patient 2 in Table 1).
Multiplanar reformation images (A and B) and volume-rendering
reformations (C and D) during diastole (A and C)
and during systole (B and D) and corresponding fluoroscopic
images (E and F) show stuck leaflet (white arrow,
A). Note soft-tissue attenuation on ventricular border (black
arrow, A) of stuck valve, which is suggestive of pannus.
|
|

View larger version (95K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 5F —66-year-old man with stuck mitral bileaflet mechanical valve
(patient 2 in Table 1).
Multiplanar reformation images (A and B) and volume-rendering
reformations (C and D) during diastole (A and C)
and during systole (B and D) and corresponding fluoroscopic
images (E and F) show stuck leaflet (white arrow,
A). Note soft-tissue attenuation on ventricular border (black
arrow, A) of stuck valve, which is suggestive of pannus.
|
|

View larger version (92K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 6A —67-year-old man with normal functioning bileaflet aortic
valve and perivalvular leak (patient 8 in
Table 1). Multiplanar
reformation images in three perpendicular planes on systole (A) and
diastole (B) show normal functioning valve with anterior dehiscence
(arrows) causing aortic regurgitation. Ao = aorta, LV = left
ventricle.
|
|

View larger version (94K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 6B —67-year-old man with normal functioning bileaflet aortic
valve and perivalvular leak (patient 8 in
Table 1). Multiplanar
reformation images in three perpendicular planes on systole (A) and
diastole (B) show normal functioning valve with anterior dehiscence
(arrows) causing aortic regurgitation. Ao = aorta, LV = left
ventricle.
|
|

View larger version (114K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 6C —67-year-old man with normal functioning bileaflet aortic
valve and perivalvular leak (patient 8 in
Table 1). Corresponding
superior views of volume-rendering reformation above aortic valve (solid
arrows) show clearly perivalvular leak (open arrows). LA = left
atrium, RA = right atrium, RVOT = right ventricular outflow tract.
|
|

View larger version (120K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 6D —67-year-old man with normal functioning bileaflet aortic
valve and perivalvular leak (patient 8 in
Table 1). Corresponding
superior views of volume-rendering reformation above aortic valve (solid
arrows) show clearly perivalvular leak (open arrows). LA = left
atrium, RA = right atrium, RVOT = right ventricular outflow tract.
|
|
The opening and closing angles are shown in
Table 1. Comparison of opening
and closing angle measurements between MDCT using MPR and fluoroscopy was
available in 11 bileaflet mechanical valves (p = 0.78 and 0.93 for
opening and closing angles, respectively). Bland-Altman plots showed high
agreement between MDCT and fluoroscopy for measurements of opening and closing
angles of bileaflet mechanical valves (Figs.
3 and
4, respectively).

View larger version (6K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 3 —Bland-Altman plot shows agreement between measurements of
bileaflet mechanical valve opening angles obtained by fluoroscopy and MDCT.
Solid line shows mean difference, and dotted lines show mean difference
± 1.96 times SD of differences.
|
|

View larger version (6K):
[in this window]
[in a new window]
[as a PowerPoint slide]
|
Fig. 4 —Bland-Altman plot shows agreement between measurements of
bileaflet mechanical valve closing angles obtained by fluoroscopy and MDCT.
Solid line shows mean difference, and dotted lines show mean difference
± 1.96 times SD of differences.
|
|
A stuck leaflet was shown on MDCT in four patients with a bileaflet mitral
mechanical valve; all were confirmed by fluoroscopy (Fig.
5A,
5B,
5C,
5D,
5E,
5F). In one of those cases, a
soft tissue, presumably a pannus, could be clearly identified at the
ventricular edge of the stuck leaflet (Fig.
5A). In one patient, a stuck valve was initially suspected at
cinefluoroscopy. After visualization of normal motion on MDCT, repeated
fluoroscopy indeed succeeded in showing normal leaflet motion.
The mean opening and closing angles in the remaining patients with
normal-functioning valves were in accordance with the manufacturer's in vitro
data and the clinical literature
[10,
11]
(Table 1).
Echocardiography showed a stuck leaflet in three of the four patients with
a stuck valve with no evidence of a thrombus or pannus in any of the cases. In
the remaining cases with available correlation, valvular morphology and
function were unremarkable. An increased transmitral pressure gradient (mean
> 10 mm Hg) was found on Doppler echocardiography in two of the four
patients with a stuck mitral valve (Table
1). In addition, an increased transaortic pressure gradient (mean
> 30 mm Hg) was noted on Doppler echocardiography in four patients with
normal prosthetic aortic valve motion. In two of these patients, significant
regurgitant flow was identified on Doppler echocardiography: One of those
patients had a single leaflet valve that could hardly be visualized
(visibility score = 1), whereas in the other patient, perivalvular dehiscence
was clearly identified on MDCT and was subsequently confirmed at surgery (Fig.
6A,
6B,
6C,
6D).
Discussion
Prosthetic mechanical valve function is traditionally evaluated with
Doppler echocardiography and fluoroscopy. However, assessment by a single
echocardiographic examination may be limited because of suboptimal
visualization of valve motion and the wide variation among transprosthetic
Doppler-derived pressure gradients. Increased prosthetic valve gradients may
be derived from different prosthetic valve sizes as well as from one of the
many causes of increased flow
[12], such as anemia, fever,
increased heart rate, or regurgitation, as was seen in two patients in our
series. Moreover, in some patients with bileaflet prosthetic valves, a stuck
single leaflet may not increase the transvalvular pressure gradient
significantly if the patient's heart rate is low. Indeed, in the present
series, only two of four patients with a stuck valve showed an increased
prosthetic valve gradient. Although fluoroscopy enables measurements of the
opening and closing angles of the leaflets, it requires very specific patient
positioning so that the x-ray beam parallels both the valve ring and the
tilting axis of the disks; performing this examination might be time-consuming
and obtaining images with the patient in the correct position is frequently
difficult.
ECG-gated MDCT is a novel technique increasingly used for coronary artery
disease evaluation. The same raw data can be used for functional prosthetic
valve assessment. Teshima et al.
[13] showed the usefulness of
8-MDCT in the diagnosis of a pannus in two patients with prosthetic valve
dysfunction. Using the latest 64-MDCT scanners, the total scanning time is
only approximately 10 seconds. Compared with fluoroscopy, MDCT, which
intrinsically is a 3D technique, has the potential to offer optimal
visualization of the valve profiles independent of valve position, the
patient's physical characteristics, limitation of C-arm motion, and the
operator's skills. Our preliminary results suggest that MDCT is a reliable
technique for functional evaluation of bileaflet mechanical valves and might
be helpful in some, but not all, cases with single-leaflet mechanical valves.
Furthermore, MDCT can be performed without contrast material injection and
through a limited span centered only to the level of the prosthetic valve,
thus allowing a shorter scanning time (
5 seconds) and a significant
reduction in the patient's radiation exposure.
The visibility of bileaflet valves on MDCT, presently the most commonly
used mechanical valves, was excellent in all 18 cases in our series when using
the MPR postprocessing technique, and MDCT measurements showed good
correlation with fluoroscopic measurements of opening and closing angles.
However, in two of the five single-leaflet valves, opening and closing angles
could not be identified with MDCT. The difficulty in visualizing
single-leaflet mechanical valves can be explained by the radiolucency of the
disk, which has only a thin metallic rim on its edges. This thin metallic line
can barely be depicted with postprocessing techniques even when using thick
volumes that include the whole contour of the disk. In addition, earlier
single-leaflet valves, such as the Björk-Shiley valve manufactured before
1975, do not contain this rim and are totally radiolucent. In comparison, in
bileaflet mechanical valves, the whole leaflet is metallic to allow easy
delineation of its exact position during each phase of the cardiac cycle using
reformations with thin volumes.
Compared with other techniques, MDCT may have additional value because it
can show perivalvular structures as well as cardiac and mediastinal
abnormalities. In one patient with aortic valve replacement, suspected aortic
regurgitation was suboptimally visualized on echocardiography. MDCT showed a
normal-functioning bileaflet prosthetic aortic valve with an annular
dehiscence causing perivalvular leak, which was confirmed at surgery. In
another patient with a stuck mitral prosthetic valve, MDCT could show also the
cause: A rim of soft tissue was clearly identified along the edge of the stuck
leaflet suggestive of a pannus or thrombus. However, in three additional cases
in our series with a stuck valve, artifacts caused by the dense metallic
valvular rings did not allow adequate visualization of the soft tissues
adjacent to the edges of the leaflet. We extrapolate that for similar reasons
MDCT might be limited in showing questionable vegetations attached to or close
to the metallic leaflets or ring.
One of the main limitations of our study is that both the radiologist and
the cardiologist were not blinded to clinical history and available imaging
results from previous studies when performing measurements on MDCT or
fluoroscopy; this is due to the clinical setup of our study. Although the
unblinded retrospective comparison might lessen the generalizability of the
study, we think that the potential bias when performing measurements based on
MDCT reformations and fluoroscopy is minimal. Another limitation of our study
is that visualization of the leaflets during fluoroscopy was suboptimal in six
patients and fluoroscopy was used as a gold standard. This problem is common
in retrospective clinical research and can be resolved by a prospective study
using both techniques on all patients. In addition, our series included a
relatively small number of patients with mechanical valves, especially those
with a single leaflet. Further larger series are required to evaluate the role
of MDCT in patients with single-leaflet mechanical valves not only with regard
to complications, such as a stuck valve but also with regard to other rare
complications, such as broken struts, that have been encountered in some
patients with the Björk-Shiley single-leaflet valves
[14].
In conclusion, our preliminary experience suggests that ECG-gated MDCT is a
helpful and reliable technique for functional mechanical prosthetic valve
evaluation, mainly in patients with bileaflet valves. It can be performed in
conjunction with preoperative coronary artery evaluation when repeated
valvular replacement is planned or can be used as a stand-alone evaluation in
conjunction with or as an alternative to traditional cinefluoroscopy. Further
prospective studies are needed to confirm these preliminary results.
References
- Rizzoli G, Guglielmi C, Toscano G, et al. Reoperations for acute
prosthetic thrombosis and pannus: an assessment of rates, relationship and
risk. Eur J Cardiothorac Surg 1999;16
: 74–80[Abstract/Free Full Text]
- Teshima H, Hayashida N, Yano H, et al. Obstruction of St. Jude
Medical valves in the aortic position: histology and immunohistochemistry of
pannus. J Thorac Cardiovasc Surg 2003;126
: 401–407[Abstract/Free Full Text]
- Aoyagi S, Nishimi Y, Kawano H, et al. Obstruction of St. Jude
Medical valves in the aortic position: significance of a combination of
cineradiography and echocardiography. J Thorac Cardiovasc
Surg 2000; 120:142
–147[Abstract/Free Full Text]
- Barbetseas J, Nagueh SF, Pitsavos C, Toutouzas PK, Quinones MA,
Zoghbi WA. Differentiating thrombus from pannus formation in obstructed
mechanical prosthetic valves: an evaluation of clinical, transthoracic and
transesophageal echocardiographic parameters. J Am Coll
Cardiol 1998; 32:1410
–1417[Abstract/Free Full Text]
- Montorsi P, Cavoretto D, Alimento M, Muratori M, Pepi M. Prosthetic
mitral valve thrombosis: can fluoroscopy predict the efficacy of thrombolytic
treatment? Circulation 2003;108
[suppl 1]:II79
–II84[Medline]
- Aoyagi S, Arinaga K, Fukunaga S, Tayama E, Kosuga T, Akashi H.
Leaflet movement of the ATS valve in the aortic position: unique behavior
observed in 19-mm valves. Ann Thorac Surg2006; 82:853
–857[Abstract/Free Full Text]
- Faletra F, Constantin C, De Chiara F, et al. Incorrect
echocardiographic diagnosis in patients with mechanical prosthetic valve
dysfunction: correlations with surgical findings. Am J
Med 2000; 108:531
–537[CrossRef][Medline]
- Ronderos RE, Portis M, Stoermann W, Sarmiento C. Are all
echocardiographic findings equally predictive for diagnosis in prosthetic
endocarditis? J Am Soc Echocardiogr 2004;17
: 664–669[CrossRef][Medline]
- Verdel G, Heethaar RM, Jambroes G, van der Werf T. Assessment of
the opening angle of implanted Björk-Shiley prosthetic valves.
Circulation 1983;68
: 355–359[Abstract/Free Full Text]
- Montorsi P, Cavoretto D, Repossini A, Bartorelli AL, Guazzi MD.
Valve design characteristics and cine-fluoroscopic appearance of five
currently available bileaflet prosthetic heart valves. Am J Card
Imaging 1996; 10:29
–41[Medline]
- Prosthetic heart valve information for use. Carbomedics Website.
Available at:
www.carbomedics.com/pdfs/CPHV.pdf.
Accessed March 10, 2008
- Aoyagi S, Nishi Y, Kawara T, Oryoji A, Kosuga K, Oishi K. Doppler
echocardiographic diagnosis of malfunction of a St. Jude Medical mitral valve.
Artif Organs Today 1994;3
: 299–307
- Teshima H, Hayashida N, Fukunaga S, et al. Usefulness of a
multidetector-row computed tomography scanner for detecting pannus formation.
Ann Thorac Surg 2004;77
: 523–526[Abstract/Free Full Text]
- O'Neill WW, Chandler JG, Gordon RE, et al. Radiographic detection
of strut separations in Björk-Shiley convexo-concave mitral valves.
N Engl J Med 1995;333
: 414–419[Abstract/Free Full Text]

CiteULike
Complore
Connotea
Del.icio.us
Digg
Reddit
Technorati What's this?
This article has been cited by other articles:

|
 |

|
 |
 
F. Chenot, P. Montant, C. Goffinet, A. Pasquet, D. Vancraeynest, E. Coche, J.-L. Vanoverschelde, and B. L. Gerber
Evaluation of Anatomic Valve Opening and Leaflet Morphology in Aortic Valve Bioprosthesis by Using Multidetector CT: Comparison with Transthoracic Echocardiography
Radiology,
December 17, 2009;
(2009)
radiol.0000082294v1.
[Abstract]
[Full Text]
|
 |
|

|
 |

|
 |
 
T. M. LaBounty, P. P. Agarwal, A. Chughtai, D. S. Bach, E. Wizauer, and E. A. Kazerooni
Evaluation of Mechanical Heart Valve Size and Function With ECG-Gated 64-MDCT
Am. J. Roentgenol.,
November 1, 2009;
193(5):
W389 - W396.
[Abstract]
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
M. Biteker, S. Gunduz, and M. Ozkan
Role of MDCT in the Evaluation of Prosthetic Heart Valves
Am. J. Roentgenol.,
February 1, 2009;
192(2):
W77 - W77.
[Full Text]
[PDF]
|
 |
|

|
 |

|
 |
 
E. Konen
Reply
Am. J. Roentgenol.,
February 1, 2009;
192(2):
W78 - W78.
[Full Text]
[PDF]
|
 |
|