Figure 5


Fig. 3B Accuracy of measured percentage stenosis. Values are means; error bars indicate mean ± 0.5 SD; light gray, 3-mm vessel diameter; dark gray, 5-mm vessel diameter. Graph shows effects of attenuation protocols of cardiac CTA on accuracy of detection of 50% stenosis. High-attenuation (500 H) protocol led to significant underestimation of stenosis for smaller vessels but was accurate in depicting larger vessels. Low-attenuation (200 H) protocol led to overestimation of stenosis within clinically acceptable limits, but imaging quality was not as reproducible as in other groups.